Ning R, Gao R, Piñero DP, Zhang J, Gao Q, Jin Y, Wang Y, Wang C, Huang J. Repeatability and reproducibility of corneal higher-order aberrations measurements after small incision lenticule extraction using the Scheimpflug-Placido topographer.
EYE AND VISION 2022;
9:1. [PMID:
34983674 PMCID:
PMC8728949 DOI:
10.1186/s40662-021-00274-y]
[Citation(s) in RCA: 2] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Subscribe] [Scholar Register] [Received: 05/13/2021] [Accepted: 11/24/2021] [Indexed: 11/18/2022]
Abstract
Background
To evaluate the precision of corneal higher-order aberrations measurements after small incision lenticule extraction (SMILE) using the Sirius Scheimpflug-Placido topographer (CSO, Italy).
Methods
Seventy-five eyes from 75 postoperative subjects were included in this prospective study. Three consecutive corneal aberrometric measurements were obtained with the Scheimpflug-Placido topographer by two experienced operators to assess intra- and inter-observer reproducibility. The within-subject standard deviation (Sw), test-retest repeatability (TRT) and the intraclass correlation coefficient (ICC) were calculated.
Results
For intraobserver repeatability of anterior and total corneal aberrations, all ICCs were more than 0.922, except for trefoil (0.722 to 0.768). The ICCs of total root mean square (RMS), coma Z (3, ± 1), and spherical aberration Z (4, 0) were over 0.810 while higher-order RMS, trefoil Z (3, ± 3), and astigmatism II Z (4, ± 2) were below 0.634 for posterior corneal surface aberrations. All Sw values for all types of aberrations were equal to or below 0.07 μm. Regarding interobserver reproducibility, all TRT values were no more than 0.12 μm, 0.05 μm, and 0.11 μm for anterior, posterior, and total corneal aberrations, respectively. The ICC values ranged from 0.875 to 0.989, from 0.686 to 0.976 and over 0.834 for anterior, posterior, and total corneal aberrations, respectively.
Conclusions
The repeatability of measurements of anterior and total corneal aberrations with the Sirius system in corneas after SMILE surgery was high, except for trefoil. There was some variability in posterior corneal aberrometric measurements. High reproducibility of corneal aberrometric measurements was observed between measurements of both examiners, except for trefoil, with poor to moderate reproducibility.
Collapse