Calcaterra D, Jazayeri MA, Turek JW, Parekh KR, Bashir M, Karam K, Farivar RS. Aortic Root Reconstruction with a New Dacron Graft Featuring Prefabricated Coronary Side Branches: Lessons Learned from the Cabrol Procedure.
AORTA : OFFICIAL JOURNAL OF THE AORTIC INSTITUTE AT YALE-NEW HAVEN HOSPITAL 2017;
5:1-10. [PMID:
28868309 DOI:
10.12945/j.aorta.2017.16.023]
[Citation(s) in RCA: 3] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Subscribe] [Scholar Register] [Received: 05/03/2016] [Accepted: 10/24/2016] [Indexed: 11/18/2022]
Abstract
BACKGROUND
Coronary button reimplantation can represent a technical challenge of aortic root reconstruction that can be associated with significant morbidity and mortality. With the goal of simplifying coronary reimplantation and reducing the incidence of related complications, we designed a new Dacron graft with prefabricated coronary branches to minimize coronary artery mobilization and prevent the potential mechanical complications of reattachment to the body of the graft.
METHODS
Between June 2010 and May 2012, we implanted the graft in eight patients (six males, two females) ranging in age from 42-68 years (mean, 54 years). Six procedures were modified Bentall reconstructions, and two procedures were valve-sparing root replacements using the reimplantation technique.
RESULTS
There were no complications and no morbidity or mortality related to coronary reattachment. All patients were alive and doing well at a mean follow-up of 26 months (range, 17-38 months). At an extended mean follow-up of 42 months (range, 25-56 months), one patient died of stroke-related complications. No radiologic or clinical evidence of impairment of coronary perfusion was identified in any patient.
CONCLUSIONS
The use of this new graft model may simplify the technique of root reconstruction and potentially lower the incidence of mechanical complications related to coronary button reimplantation.