Vandenberk B, Murray K, Rizkallah J. Risk Mitigation of Pacemaker Pocket Erosion in the Thin – A Case Report.
CJC Open 2022;
4:585-587. [PMID:
35734518 PMCID:
PMC9207779 DOI:
10.1016/j.cjco.2022.03.002]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/06/2022] [Accepted: 03/04/2022] [Indexed: 11/19/2022] Open
Abstract
The aging population, particularly the thin and frail, has an increased risk of long-term cardiac implantable electronic device complications. This case is that of an elderly, thin-skinned patient who presented with a pacemaker pocket erosion 4 years after elective generator change, potentiated by a small pocket size with a superficial suture fixating the generator in the subcutaneous pocket. The risk for device erosion may have been mitigated during the generator change by increasing the size of the pocket, using a submuscular pocket, and potentially an absorbable antibacterial envelope. Fixation of the generator is considered optional.