Jung BK, Cho CH, Moon KC, Sung Hur D, Yoon JA, Yoon SY. Detection of platelet-monocyte aggregates by the ADAM(®) image cytometer.
Int J Med Sci 2014;
11:1228-33. [PMID:
25317068 PMCID:
PMC4196123 DOI:
10.7150/ijms.10008]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 06/30/2014] [Accepted: 08/31/2014] [Indexed: 01/28/2023] Open
Abstract
BACKGROUND
Inappropriate platelet activation is known to be associated with various thrombotic disorders. Platelet-monocyte aggregates (PMAs), whose formation is mediated by platelet surface P-selectin (CD62P), can be used as a reliable marker to detect platelet activation. Previous studies have generally detected PMAs through flow cytometry-based approaches. Recently, the ADAM(®) image cytometer (Nanoentek Inc., Seoul, Korea) was developed for image-based cellular analysis. In this study, we detected PMAs with the ADAM(®) cytometer, evaluated the reproducibility of the measurements made by the ADAM(®) cytometer, and compared the abilities of the ADAM(®) cytometer and a flow cytometric assay to detect PMAs.
METHODS
Whole blood samples were collected from patients. Within 5 minutes of collection, anticoagulated whole blood samples were fixed in 10% paraformaldehyde and 5% glyoxal. Nineteen clinical specimens were collected; each was analyzed three times with the ADAM(®) cytometer in order to assess the reproducibility of its measurements. To compare the ability of the ADAM(®) cytometer with that of a flow cytometer to detect PMAs, each cytometer was used for 23 clinical samples and the correlation of the measurements was determined.
RESULTS
The PMA measurements made by the ADAM(®) cytometer showed good reproducibility (CV < 10% for all specimens). Moreover, the PMA measurements made by the ADAM(®) cytometer exhibited a high correlation with those made by a flow cytometric assay (R = 0.944).
CONCLUSIONS
The ADAM(®) cytometer is a suitable alternative method to the flow cytometry-based assays. Since the ADAM cytometer does not need specialized instrument knowledge or software proficiency (unlike flow cytometry), the ADAM(®) cytometer can be used as a rapid and reliable POCT device to measure platelet activation in peripheral blood. This, in turn, will provide valuable information regarding patient propensities to thrombotic diseases.