Bakhtiari S, Alvandi A, Pajavand H, Navabi J, Najafi F, Abiri R. Development and Diagnostic Evaluation of Loop-Mediated Isothermal Amplification Using a New Gene Target for Rapid Detection of Helicobacter pylori.
Jundishapur J Microbiol 2016;
9:e28831. [PMID:
27540449 PMCID:
PMC4976074 DOI:
10.5812/jjm.28831]
[Citation(s) in RCA: 16] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/30/2015] [Revised: 11/18/2015] [Accepted: 11/27/2015] [Indexed: 01/05/2023] Open
Abstract
BACKGROUND
Helicobacter pylori cause chronic gastritis and subsequent diseases like gastric and duodenal ulcers and gastric adenocarcinoma. Current methods for detecting H. pylori have several disadvantages and it is of utmost importance to develop a simple, quick, accurate, and cost-effective diagnostic test.
OBJECTIVES
The aim of this study was to set up and evaluate a diagnostic value of loop- mediated isothermal amplification (LAMP) for detecting H. pylori.
PATIENTS AND METHODS
The analytical sensitivity values (limit of detection) of LAMP and polymerase chain reaction (PCR) were determined using serial dilutions of H. pylori DNA. Analytical specificity of the methods using new designed primers targeted ureC gene was also determined.
RESULTS
The detection limits of the LAMP and PCR assay were similar and were 10 fg of pure DNA of H. pylori, which is equal to 6 copy numbers of H. pylori genome. Analytical specificity of the tests was 100% because the tests were positive only with H. pylori DNA.
CONCLUSIONS
The analytical sensitivity of LAMP and PCR methods, using the designed primers, was 8 times more than any other reported methods. The designed methods are specific and sensitive for detection of H. pylori in different clinical and environmental samples.
Collapse