O'Brien SM, Winter C, Burden CA, Boulvain M, Draycott TJ, Crofts JF. Pressure and traction on a model fetal head and neck associated with the use of forceps, Kiwi™ ventouse and the BD Odon Device™ in operative vaginal birth: a simulation study.
BJOG 2019;
124 Suppl 4:19-25. [PMID:
28940875 PMCID:
PMC7198111 DOI:
10.1111/1471-0528.14760]
[Citation(s) in RCA: 16] [Impact Index Per Article: 2.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Journal Information] [Subscribe] [Scholar Register] [Accepted: 07/05/2017] [Indexed: 11/30/2022]
Abstract
Objective
To determine the pressure and traction forces exerted on a model fetal head by the BD
Odon Device, forceps and Kiwi ventouse during simulated births.
Design
Simulation study.
Setting
Simulated operative vaginal birth.
Population or Sample
84 simulated operative vaginal births.
Methods
A bespoke fetal mannequin with pressure sensors around the head and strain gauge across
the neck was used to investigate pressure applied over the head, and traction across the
neck during 84 simulated births using the BD Odon Device, non-rotational forceps and
Kiwi ventouse.
Main Outcome Measures
Peak pressure on the fetal face and lateral aspects of the head during correct use of
the BD Odon Device and forceps. Peak pressure on orbits and neck during misplacement of
the BD Odon Device and forceps. Peak traction force generated until instrument failure
using the BD Odon Device, forceps and Kiwi ventouse.
Results
When correctly sited and using 80kPa inflation pressure on the cuff, the BD Odon Device
generated a lower peak pressure on the fetal head than forceps (83kPa vs 146kPa). When
instruments were purposefully misplaced over the orbits the BD Odon Device generated a
lower peak pressure on the orbits than forceps (70kPa vs 123kPa). When purposefully
misplaced over the neck the BD Odon Device, compared to forceps, generated a greater
peak pressure on the anterio-lateral aspect of the neck (56kPa vs 17kPa) and a lower
peak pressure on the posterior aspect of the neck (76kPa vs 93kPa) than forceps. In
cases of true cephalic disproportion the BD Odon Device ‘popped-off’ at a
lower traction force than forceps (208N vs 270N).
Conclusions
In simulated assisted vaginal birth with correctly placed instruments the peak pressure
exerted on the fetal head by a BD Odon Device is lower than pressure exerted by
non-rotational forceps. In cases in which delivery of the fetal head is not possible due
to cephalo-pelvic disproportion lower traction forces could be applied using the BD Odon
Device than with forceps before the procedure was abandoned due to device failure.