6-Month Evaluation of a Transcatheter Aortic Valve (Myval) in a Novel Ovine, Supra-Aortic Banding Model.
JACC Basic Transl Sci 2022;
7:486-495. [PMID:
35663633 PMCID:
PMC9156430 DOI:
10.1016/j.jacbts.2022.01.008]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 10/18/2021] [Revised: 01/19/2022] [Accepted: 01/19/2022] [Indexed: 10/26/2022]
Abstract
The aim of the study was to evaluate a balloon expandable transcatheter heart valve (THV) system (Myval) at 6-month follow-up in ovine banding model. Eleven THV systems were implanted via carotid approach. There were 2 procedure-related deaths and 2 premature deaths. At 6 months all valves that completed follow-up (n = 7) were functional, with no significant regurgitation, calcification, thrombi, or vegetation. Mean pressure gradient was 21.9 ± 11 mm Hg, maximum velocity = 3.3 ± 1 m/s, and ejection fraction was 53.3 ± 6%. Myval THV showed optimal hemodynamic performance and biocompatibility.