Neumann HJ, Meisberger A, Mathieu B. Improvement of the safety systems in cell separators. The new safety concept for the cell separator AS 104 (Fresenius).
Infusionsther Klin Ernahr 1987;
14 Suppl 4:43-51. [PMID:
3316023 DOI:
10.1159/000226204]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 01/05/2023]
Abstract
The new, extended safety requirements for cell separators call for a new generation of machines to address current safety needs. In order to ensure safe treatment, the complete safety concept is of decisive importance - this includes closed seal-less systems, continuous operation, fractions collected externally to the centrifuge, and pump tubings, connections and clamps which are color and direction coded. 'Human failure' as a cause of fault should be excluded for these highly technical appliances - for example, through the use of 'fail-safe' circuits, drip-controlled ACD flow by a separate adjustable pump, detection of damages to blood components by hemolysis control.