Effect of Disposable Elevator Cap Duodenoscopes on Persistent Microbial Contamination and Technical Performance of Endoscopic Retrograde Cholangiopancreatography: The ICECAP Randomized Clinical Trial.
JAMA Intern Med 2023;
183:191-200. [PMID:
36689215 PMCID:
PMC9871945 DOI:
10.1001/jamainternmed.2022.6394]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 10/10/2022] [Accepted: 11/23/2022] [Indexed: 01/24/2023]
Abstract
Importance
Infection transmission following endoscopic retrograde cholangiopancreatography (ERCP) can occur due to persistent contamination of duodenoscopes despite high-level disinfection to completely eliminate microorganisms on the instrument.
Objective
To determine (1) contamination rates after high-level disinfection and (2) technical performance of duodenoscopes with disposable elevator caps compared with those with standard designs.
Design, Setting, and Participants
In this parallel-arm multicenter randomized clinical trial at 2 tertiary ERCP centers in Canada, all patients 18 years and older and undergoing ERCP for any indication were eligible.
Intervention
The intervention was use of duodenoscopes with disposable elevator caps compared with duodenoscopes with a standard design.
Main Outcomes and Measures
Coprimary outcomes were persistent microbial contamination of the duodenoscope elevator or channel, defined as growth of at least 10 colony-forming units of any organism or any growth of gram-negative bacteria following high-level disinfection (superiority outcome), and technical success of ERCP according to a priori criteria (noninferiority outcome with an a priori noninferiority margin of 7%), assessed by blinded reviewers.
Results
From December 2019 to February 2022, 518 patients were enrolled (259 disposable elevator cap duodenoscopes, 259 standard duodenoscopes). Patients had a mean (SD) age of 60.7 (17.0) years and 258 (49.8%) were female. No significant differences were observed between study groups, including in ERCP difficulty. Persistent microbial contamination was detected in 11.2% (24 of 214) of standard duodenoscopes and 3.8% (8 of 208) of disposable elevator cap duodenoscopes (P = .004), corresponding to a relative risk of 0.34 (95% CI, 0.16-0.75) and number needed to treat of 13.6 (95% CI, 8.1-42.7) to avoid persistent contamination. Technical success using the disposable cap scope was noninferior to that of the standard scope (94.6% vs 90.7%, P = .13). There were no differences between study groups in adverse events and other secondary outcomes.
Conclusions and Relevance
In this randomized clinical trial, disposable elevator cap duodenoscopes exhibited reduced contamination following high-level disinfection compared with standard scope designs, without affecting the technical performance and safety of ERCP.
Trial Registration
ClinicalTrials.gov Identifier: NCT04040504.