Biocompatibility evaluation of a thermoplastic rubber for wireless telemetric intracranial pressure sensor coating.
J Mech Behav Biomed Mater 2015;
45:83-9. [PMID:
25688030 DOI:
10.1016/j.jmbbm.2015.01.018]
[Citation(s) in RCA: 7] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/31/2014] [Revised: 01/21/2015] [Accepted: 01/26/2015] [Indexed: 12/13/2022]
Abstract
This study investigated the biocompatibility of the experimental thermoplastic rubber Arbomatrix(™) that will be used as the protective coating on a novel intracranial pressure (ICP) sensor silicon chip. Arbomatrix(™) was benchmarked against biocompatible commercial silicone rubber shunt tubing in the brain via a rat model with 60-day implant duration. A bare silicon chip was also implanted. The results showed similar cellular distribution in the brain-implant boundary and surrounding tissues. Quantitative analysis of neuron and glia density did not show significant difference between implants. Through histological and immunohistochemical evaluation we conclude that Arbomatrix(™) is well tolerated by the brain. Due to its exceptional barrier properties Arbomatrix(™) has already been shown to be an excellent protective coating for new ICP monitoring chip.