A six-month evaluation of the VivaSight™ video double-lumen endotracheal tube after introduction into thoracic anaesthetic practice at a single institution.
Anaesth Intensive Care 2017;
45:189-195. [PMID:
28267940 DOI:
10.1177/0310057x1704500208]
[Citation(s) in RCA: 8] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/27/2022]
Abstract
For a six-month period, all airway options used for non-emergent patients undergoing thoracic surgery requiring one-lung ventilation at a single institution were assessed after introduction of the VivaSight™ double-lumen endotracheal tube (VivaSight-DL), a novel double-lumen tube with an integrated camera. This device displays a continuous view of the position of the tube relative to the carina. A total of 72 patients had lung separation with the VivaSight-DL. Lung separation was achieved on first attempt without additional manipulation in 85% of cases. In only three cases (4%) was a fibreoptic bronchoscope required, in each instance to reposition the tube after intraoperative dislodgement. The VivaSight-DL represents a novel method of one-lung ventilation allowing rapid identification of intraoperative airway problems and reducing the need for fibreoptic bronchoscopy.
Collapse