Abstract
OBJECTIVES
To assess the relevance of current monitoring alarms as a warning system in the adult ICU.
DESIGN
Prospective, observational study.
SETTINGS
Two university hospital, and three general hospital, ICUs.
PATIENTS
Hundred thirty-one patients, ventilated at admission, from different shifts (morning, evening, night) combined with different stages of stay, early (0-3 days), intermediate (4-6 days) and late (> 6 days).
INTERVENTIONS
Experienced nurses were asked to record the patient's characteristics and, for each alarm event, the reason, type and consequence.
MEASUREMENTS AND MAIN RESULTS
The mean age of the patients included was 59.8 +/- 16.4 and SAPS1 was 15.9 +/- 7.4. We recorded 1971 h of care. The shift distribution was 78 mornings, 85 evenings and 83 nights; the stage distribution was 88 early, 78 intermediate and 80 late. There were 3188 alarms, an average of one alarm every 37 min: 23.7% were due to staff manipulation, 17.5% to technical problems and 58.8% to the patients. Alarms originated from ventilators (37.8%), cardiovascular monitors (32.7%), pulse oximeters (14.9%) and capnography (13.5%). Of the alarms, 25.8% had a consequence such as sensor repositioning, suction, modification of the therapy (drug or ventilation). Only 5.9% of the alarms led to a physician's being called. The positive predictive value of an alarm was 27% and its negative predictive value was 99%. The sensitivity was 97% and the specificity 58%.
CONCLUSIONS
The study confirms that the level of monitoring in ICUs generates a great number of false-positive alarms.
Collapse