The Potential of Utilizing Mid-Energy X-Rays for In-Line Phase Sensitive Breast Cancer Imaging.
BIOMEDICAL SPECTROSCOPY AND IMAGING 2020;
9:89-102. [PMID:
34141562 PMCID:
PMC8208526 DOI:
10.3233/bsi-200204]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/12/2023]
Abstract
OBJECTIVE
The objective of this study is to demonstrate the potential of utilizing mid-energy x-rays for in-line phase-sensitive breast cancer imaging by phantom studies.
METHODS
The midenergy (50-80kV) in-line phase sensitive imaging prototype was used to acquire images of the contrast-detail mammography (CDMAM) phantom, an ACR accreditation phantom, and an acrylic edge phantom. The low-dose mid-energy phase-sensitive images were acquired at 60 kV with a radiation dose of 0.9 mGy, while the high-energy phase-sensitive images were acquired at 90 kV with a radiation dose of 1.2 mGy. The Phase-Attenuation Duality (PAD) principle for soft tissue was used for the phase retrieval. A blind observer study was conducted and paired-sample T-test were performed to compare the mean differences in the two imaging systems.
RESULTS
The correct detection ratio for the CDMAM phantom for phase-contrast images acquired by the low-dose mid-energy system was 56.91%, whereas images acquired by the high-energy system correctly revealed only 40.97% of discs. The correct detection ratios were 57.88% and 43.41% for phase-retrieved images acquired by the low-dose mid-energy and high-energy imaging systems, respectively. The reading scores for all three groups of objects in the ACR phantom were higher for the mid energy imaging system as compared to the high-energy system for both phase-contrast and phase- retrieved images. The calculated edge enhancement index (EEI) from the acrylic edge phantom image for the mid-energy system was higher than that calculated for the high-energy imaging system. The quantitative analyses showed a higher Contrast to Noise Ratio (CNR) as well as a higher Figure of Merit (FOM) in images acquired by the low-dose mid-energy imaging system.
CONCLUSION
The PAD based retrieval method can be applied in mid-energy system without remarkably affecting the image quality, and in fact, it improves the lesion detectability with a patient dose saving of 25%.