Worsening tricuspid regurgitation after ICD implantation is rather due to transvenous lead than natural progression.
Int J Cardiol 2023;
376:76-80. [PMID:
36758860 DOI:
10.1016/j.ijcard.2023.02.007]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 11/16/2022] [Revised: 01/17/2023] [Accepted: 02/06/2023] [Indexed: 02/11/2023]
Abstract
BACKGROUND
Transvenous implantable cardioverter-defibrillators (TV-ICDs) are associated with greater tricuspid regurgitation (TR) severity, which leads to increased mortality. The pathophysiology is assumed to be lead-related, hence, treatment includes lead extraction. However, TR may also naturally occur in the high-risk ICD population, or may be caused by right ventricular pacing. We sought to evaluate the effect of ICD type (with or without lead) and pacing percentage on post-implantation TR severity.
METHODS
In this retrospective cohort study, consecutive patients were included with a primary S-ICD or TV-ICD implantation between 2009 and 2019 and echocardiography studies ≤3 months before and ≤ 3 years post-implantation. The effect of ICD type on TR severity at follow-up was estimated adjusting for ventricular pacing percentage and potential confounders. The effect of ventricular pacing percentage on TR severity at follow-up was adjusted for potential confounders.
RESULTS
118 patients were included (mean age 52 ± 21): 31 (26%) with an S-ICD and 87 (74%) with a TV-ICD. Median 20 months post-implantation, worsening TR was found in 11/31 (34%) S-ICD patients and 45/87 (52%) TV-ICD patients (p = 0.15). Adjusted for age, atrial fibrillation, baseline TR and mitral regurgitation, ventricular pacing percentage, ICD dwelling time, BMI, hypertension and left ventricular ejection fraction, TV-ICDs were significantly associated with greater TR severity (OR 9.90, p = 0.002). Ventricular pacing percentage was very low, and not significantly associated with greater TR severity (OR 0.95, p = 0.066).
CONCLUSIONS
Our results suggest that greater TR severity in ICD patients is mainly caused by the transvenous lead, rather than natural progression in the ICD population.
Collapse