Detection of oesophageal intubations using cuff pressures in a pig trachea–oesophagus model.
Br J Anaesth 2007;
99:740-3. [PMID:
17715140 DOI:
10.1093/bja/aem244]
[Citation(s) in RCA: 9] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022] Open
Abstract
BACKGROUND
The cuff pressures may be different in oesophageal and tracheal intubations. We conducted a study to evaluate if cuff pressures of endotracheal tubes (ETTs) could provide information to distinguish tracheal or oesophageal intubations in a pig trachea-oesophagus model.
METHODS
In each preparation of pig trachea-oesophagus model, the trachea and the oesophagus were intubated separately with a cuffed ETT, and the cuff pressures were measured after each 1 ml increment of air (1-10 ml) during inflation. The cuff pressures and the pressure-volume relationships in both intubations were compared.
RESULTS
The cuff pressures of oesophageal intubations were significantly higher than those of tracheal intubations in all comparisons from 1 to 10 ml of cuff volumes (P < 0.05). The cuff pressure-volume curve was steeper in the oesophageal intubation group, and the difference between the two curves was the largest when the cuff volume was 4-5 ml.
CONCLUSIONS
We conclude that the cuff pressures may be useful in detecting oesophageal intubations. This method is faster than other confirmation measures as it can detect inadvertent oesophageal intubations at the time of inflating the cuffs.
Collapse