Abstract
To explore the possibility of performing percutaneous lung biopsy safely in patients mechanically ventilated with positive-end expiratory pressure (PEEP), transthoracic needle biopsy was performed in 16 anesthetized mongrel dogs mechanically ventilated with 10 cm H2O of PEEP. To obtain the biopsy sample, a 25-gauge "skinny needle" was passed through a 20-gauge sheath and placed up to 6.25 cm deep. After satisfactory samples were obtained, both needles were withdrawn in the control group, but in the treated group, the outer sheath was used to inject 0.5 ml of isobutyl 2-cyanoacrylate to seal the needle track. Pneumothorax was present in 7 (87.5%) of 8 dogs in the control group and in 2 (25%) of 8 dogs in the treated group (p = .0147). Lung tissue exhibited an apparent foreign-body granulomatous inflammatory reaction. This technique could extend the benefits of transthoracic needle biopsy to mechanically ventilated patients, but further studies to prove the safety of isobutyl 2-cyanoacrylate are necessary.
Collapse