Brayman AA, Strickler PL, Luan H, Barned SL, Raeman CH, Cox C, Miller MW. Hemolysis of 40% hematocrit, Albunex-supplemented human erythrocytes by pulsed ultrasound: frequency, acoustic pressure and pulse length dependence.
Ultrasound Med Biol 1997;
23:1237-1250. [PMID:
9372572 DOI:
10.1016/s0301-5629(97)00126-9]
[Citation(s) in RCA: 19] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/22/2023]
Abstract
The dependence of hemolysis produced by pulsed ultrasound on ultrasound frequency, acoustic pressure and pulse length was explored. Human erythrocytes (40% hematocrit; in Albunex-supplemented autologous plasma) were exposed (60 s) to 20 or 200 microns pulses of ultrasound at frequencies of 1.02, 2.24 or 3.46 MHz and at peak negative pressures [P-] ranging from 0.0 to approximately 3.0 MPa in 0.5 MPa increments. The duty factor was 0.01. At each frequency, hemolysis increased with increasing acoustic pressure and depended weakly on pulse duration. At relatively high acoustic pressures, hemolysis depended strongly on ultrasound frequency; at lower pressures, the frequency dependence was weaker. The potential clinical significance of ultrasonic hemolysis is discussed.