[Performance of matrix assisted laser desorption ionization time of flight biotyper system in clinical bacteria identification].
ZHONGHUA YI XUE ZA ZHI 2018;
98:3607-3612. [PMID:
30486579 DOI:
10.3760/cma.j.issn.0376-2491.2018.44.012]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Subscribe] [Scholar Register] [Indexed: 06/09/2023]
Abstract
Objective: To evaluate the performance of MALDI Biotyper system in identification of clinically isolated pathogens so as to provide a new rapid identification method. Methods: Total 21 270 pathogens strains, isolated from the First Affiliated Hospital of Fujian Medical Universityduring Nov. 2015 to Dec. 2016, were identified by VITEK-Ⅱ, API and MALDI Biotyper system, respectively.The isolated strains were confirmed by DNA sequencing. Results: The identification of common bacteria with MALDI Biotyper and phenotypic system is highly consistent (>95% and >90%). Among 43 strains of anaerobic bacteria, MALDI Biotyper could identify 90.7% bacteria to species level and 97.7% bacteria to genus level with the statistical significance(χ(2)=6.76, P<0.01), while phenotypic system only identified 65.1% bacteria to species and 69.8% bacteria to genus. Also, no statistical significance was shown for Trichosporon and Candida(P>0.05). MALDI Biotyper could identify 76% filamentous fungi and all of Actinomycetes, Nocardia, Mycobacterium and Legionella to genus level. Conclusions: MALDI Biotyper is an easy-performed, sensitive method for the identification of clinically isolated pathogens. Additionally, the pretreatment and reference database has the effect on identification.