[Sonographic detection of microcalcifications - potential of new method].
ULTRASCHALL IN DER MEDIZIN (STUTTGART, GERMANY : 1980) 2012;
33:357-365. [PMID:
22322544 DOI:
10.1055/s-0031-1299128]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/31/2023]
Abstract
PURPOSE
Does the easier microcalcification detection (EMD) method enable sonographic visualization of microcalcifications in breast core biopsy specimens compared with mammography?
MATERIALS AND METHODS
In a prospective randomized study, 105 core biopsy specimens obtained with stereotactic guidance were examined by mammography and ultrasound. EMD is integrated in a high-end ultrasound system and uses three level settings (0 - 2 blue, 3 - 5 violet, and 6 - 8 black-and-white; 14 MHz). Detection of microcalcifications per core specimen was determined for ultrasound and mammography. EMD image quality was rated on a scale of 1 - 9. ANOVA and Sidak post-hoc testing, Pearson regression analysis (r), and Spearman rank correlation (rho) were performed. The intraclass correlation coefficient (ICC) was calculated, and an ROC analysis was conducted.
RESULTS
The blue color map 1 was assigned the highest mean score of 1.5 ± 0.7 (p< 0.05 compared with black-and-white and violet). There was good correlation between the two modalities (r= 0.708 and rho= 0.694) with detection of 3.5 ± 3.1 microcalcifications per specimen by ultrasound versus 4.3 ± 4.8 by mammography (p> 0.05). The ICC of 0.773 indicates little disagreement between the two modalities. ROC analysis showed mammography to be superior to ultrasound compared with histological detection of microcalcifications (AUC= 0.837 vs. AUC= 0.728).
CONCLUSION
Sonographic detection of microcalcifications in stereotactic biopsy specimens using the EMD method correlates well with digital mammography. Mammography is slightly superior.