Entering into 2.0 cystinuric management with a medical digital tool to monitor urine pH: a prospective, randomized study.
Actas Urol Esp 2023;
47:560-565. [PMID:
37086842 DOI:
10.1016/j.acuroe.2023.04.004]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/26/2023] [Revised: 02/17/2023] [Accepted: 02/20/2023] [Indexed: 04/24/2023]
Abstract
BACKGROUND AND OBJECTIVES
Individuals with cystinuria can experiment recurrent lithiasis events due to the relative insolubility of cystine at physiological urine pH, resulting in renal function decline. The Lit-Control® pH Meter is a medical device that accurately allows urine pH self-monitoring. The main objective of this study was to compare the usability of the Lit-Control® pH Meter with the reactive strips for self-monitoring of urinary pH at home by patients with cystinuria, and their overall satisfaction with each tool.
PATIENTS AND METHODS
We included 28 patients (9 females and 19 males, age 19-76 years), who were randomly assigned to monitor their urine pH with reactive strips (n = 17) or the Lit-Control® pH-meter (n = 11).
RESULTS
After six months of use, the satisfaction with the two methods was similarly high, but the patients rated (0-10 scale) the pH meter better in terms of ease of learning (mean ± SD, 8.11 ± 0.60 vs. 7.06 ± 1.18; P = 0.038), ease to prepare (8.22 ± 0.67 vs. 7.25 ± 1.18; P = 0.034), and ease of use (8.22 ± 0.67 vs. 7.25 ± 1.39; P = 0.062). Overall, patients did not reach the alkalinization goals (pH between 7.0 and 8.0).
CONCLUSIONS
The Lit-Control® pH Meter demonstrated to be an easy-to-use device that can facilitate urinary pH control by cystinuric patients. A prospective study is warranted to assess the correlation between urine pH monitoring, a treat to target approach, and the recurrence of cystine stones.
Collapse