Background: After encouraging results with the Edwards Sapien and XT valves, this study aimed to review procedural data and early outcomes for the Sapien 3 valves for transcatheter pulmonary valve replacement (TPVR). Methods: We performed a multicenter, retrospective analysis of cases who underwent a Sapien 3 TPVR between 2015 and 2017 in 7 centers in Germany with a follow-up of up to 2 years. Results: 56 patients could be enrolled (weight 58,5 ± 25,0 kg; 53% Tetralogy of Fallot, 45% native RVOT). Most procedures were two-stage procedures (82,1%) with 100% prestenting. Valve sizes were 20 mm (n = 1), 23 mm (n = 15), 26 mm (n = 27), 29 mm (n = 13). Procedural success rate was 96.4%. Two patients underwent surgical valve implantation after balloon rupture during TPVR. Follow-up data were available up to 24-month post TPVR. The rate of patients with ? moderate and severe pulmonary regurgitation decreased to 0% after TPVR, peak systolic gradient decreased from 24,2 (SD±20,9) mmHg to 7,1 mmHg (SD±5,0). There were no endocarditis, severe tricuspid valve impairment or stent fractures. Conclusions: With the Edwards Sapien 3 valve, the patient pool for TPVR can be substantially extended. Continued data collection is necessary to verify long-term results.