[Determination of methylmercury in urine by direct mercury analyzer].
ZHONGHUA LAO DONG WEI SHENG ZHI YE BING ZA ZHI = ZHONGHUA LAODONG WEISHENG ZHIYEBING ZAZHI = CHINESE JOURNAL OF INDUSTRIAL HYGIENE AND OCCUPATIONAL DISEASES 2023;
41:304-306. [PMID:
37248187 DOI:
10.3760/cma.j.cn121094-20211110-00556]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Subscribe] [Scholar Register] [Indexed: 05/31/2023]
Abstract
Objective: To develop asolvent extraction-direct mercury analyzer method for determination of methylmercury in urine. Methods: After the urinehydrolyzesd by hydrobromic acid, methylmercury was extracted by tolueneand reverse-extracted from L-cysteine solution, it was then detectedbydirect mercuryanalyzer. Results: The linear range was 0.2-50.0 μg/L, and the related coefficient was 0.9999. The relative standard deviations (RSD) within the group were 5.04%-6.64%, and the RSD between the group were 5.65%-8.11 %. The average recovery efficiencies were 85.4%-95.5%. The detection limitation was 0.0482 μg/L and the quantification concentrations was 0.1607 μg/L. Conclusion: The method, which has low detection limit, high sensitivity, easy to operate, is stability for the determination of methylmercury in urine.
Collapse