Costantino A, Piagnani A, Caccia R, Sorge A, Maggioni M, Perbellini R, Donato F, D'Ambrosio R, Sed NPO, Valenti L, Prati D, Vecchi M, Lampertico P, Fraquelli M. Reproducibility and accuracy of a pocket-size ultrasound device in assessing liver steatosis.
Dig Liver Dis 2024;
56:1032-1038. [PMID:
38016894 DOI:
10.1016/j.dld.2023.11.014]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 08/25/2023] [Revised: 11/07/2023] [Accepted: 11/09/2023] [Indexed: 11/30/2023]
Abstract
BACKGROUND
This diagnostic prospective study compared the feasibility and diagnostic accuracy of Pocket-size Ultrasound Devices (PUDs) against standard ultrasound (US) in detecting liver steatosis using the controlled attenuation parameter (CAP) and liver biopsy as reference standards.
MATERIALS AND METHODS
Consecutive patients with chronic liver diseases were assessed for the presence of steatosis using PUD and US. A CAP cut-off value >275 dB/m was applied to establish ≥S1. A 26-patient subgroup underwent liver biopsy. PUD reproducibility was evaluated using Cohen's k statistic. Diagnostic accuracy of PUD and US was given as Sensibility (Sn), Specificity (Sp), Positive and Negative Predictive Values (PPV, NPV), positive and negative Likelihood Ratio (LR+, LR-).
RESULTS
81 consecutive patients (69% males) with multiple etiologies were enroled. PUD inter-observer agreement was good (k 0.77, 95%CI 0.62-0.93). PUD and US identified ≥S1 according to CAP values respectively with Sn 0.87, Sp 0.61, PPV 0.49, NPV 0.91, LR+ 2.04, LR- 0.07, AUROC 0.74 and Sn 0.96, Sp 0.54, PPV 0.47, NPV 0.97, LR+ 2.10, LR- 0.07, AUROC 0.75.
CONCLUSIONS
PUD shows good reproducibility and diagnostic accuracy in ruling liver steatosis out, representing a useful point-of-care tool to avail of hepatologists interested in excluding NAFLD, but with basic US skills.