Approximation of Resting Energy Expenditure in Intensive Care Unit Patients Using the SenseWear Bracelet: A Comparison With Indirect Calorimetry.
JPEN J Parenter Enteral Nutr 2016;
41:976-980. [PMID:
26979281 DOI:
10.1177/0148607116633808]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/15/2022]
Abstract
BACKGROUND AND AIMS
Indirect calorimetry (IC) is the gold standard for determining energy expenditure in patients requiring mechanical ventilation. Metabolic armbands using data derived from dermal measurements have been proposed as an alternative to IC in healthy subjects, but their utility during critical illness is unclear. The aim of this study was to determine the level of agreement between the SenseWear armband and the Deltatrac Metabolic Monitor in mechanically ventilated intensive care unit (ICU) patients.
METHODS
Adult ICU patients requiring invasive ventilator therapy were eligible for inclusion. Simultaneous measurements were performed with the SenseWear Armband and Deltatrac under stable conditions. Resting energy expenditure (REE) values were registered for both instruments and compared with Bland-Altman plots.
RESULTS
Forty-two measurements were performed in 30 patients. The SenseWear Armband measured significantly higher REE values as compared with IC (mean bias, 85 kcal/24 h; P = .027). Less variability was noted between individual SenseWear measurements and REE as predicted by the Harris-Benedict equation (2 SD, ±327 kcal/24 h) than when IC was compared with SenseWear and Harris-Benedict (2 SD, ±473 and ±543 kcal/24 h, respectively).
CONCLUSIONS
The systematic bias and large variability of the SenseWear armband when compared with gas exchange measurements confer limited benefits over the Harris Benedict equation in determining caloric requirements of ICU patients.
Collapse