[Citation(s) in RCA: 13] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Subscribe] [Scholar Register] [Received: 04/23/2014] [Revised: 06/06/2014] [Accepted: 06/07/2014] [Indexed: 05/21/2023]
Abstract
Ultrashort echo time (UTE) imaging is a well-known technique used in medical MRI, however, the implementation of the sequence remains non-trivial. This paper introduces UTE for non-medical applications and outlines a method for the implementation of UTE to enable accurate slice selection and short acquisition times. Slice selection in UTE requires fast, accurate switching of the gradient and r.f. pulses. Here a gradient "pre-equalization" technique is used to optimize the gradient switching and achieve an effective echo time of 10μs. In order to minimize the echo time, k-space is sampled radially. A compressed sensing approach is used to minimize the total acquisition time. Using the corrections for slice selection and acquisition along with novel image reconstruction techniques, UTE is shown to be a viable method to study samples of cork and rubber with a shorter signal lifetime than can typically be measured. Further, the compressed sensing image reconstruction algorithm is shown to provide accurate images of the samples with as little as 12.5% of the full k-space data set, potentially permitting real time imaging of short T2(*) materials.