Osteocompatibility of Si
3N
4-coated carbon fiber-reinforced polyetheretherketone (CFRP) and hydroxyapatite-coated CFRP with antibiotics and antithrombotic drugs.
J Artif Organs 2022:10.1007/s10047-022-01340-5. [PMID:
35729412 DOI:
10.1007/s10047-022-01340-5]
[Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/17/2022] [Accepted: 05/17/2022] [Indexed: 10/18/2022]
Abstract
This study used a rabbit model to investigate the osteocompatibility of Si3N4-coated carbon fiber-reinforced polyetheretherketone (CFRP) and hydroxyapatite (HA)-coated CFRP with antibiotics (vancomycin [VCM]) and antithrombotic drugs (polyvinylpyrrolidone [PVP]). HA-coated cylindrical CFRP implants were used as the controls (HA), and HA-coated implants treated with VCM and PVP were prepared (HA-VP) as the test groups; a cylindrical CFRP coated with Si3N4 was also prepared (SiN). Ten implants from each group were randomly inserted into the femoral diaphysis of rabbits. The pull-out test, radiological analysis using micro-computed tomography (µ-CT), and histological analysis were performed. The pull-out strength of the SiN group was lower than that of the HA group. µ-CT analysis revealed that the amount of bone formation around the implant in the SiN group was inferior to that in the HA group. Conversely, the HA-VP group had equivalent pull-out strength and bone formation as analyzed by µ-CT compared to the HA group. In conclusion, the additional surface treatment of the HA-coated CFRP with VCM and PVP provided sufficient bone fixation and formation.
Collapse