Left and Right Ventricle Leads Switch as a Solution for TWave Oversensing - How a Good Idea Turned Out Bad.
Indian Pacing Electrophysiol J 2014;
14:157-60. [PMID:
24920871 PMCID:
PMC4032783 DOI:
10.1016/s0972-6292(16)30757-4]
[Citation(s) in RCA: 4] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/16/2022] Open
Abstract
A 50-year-old male with a CRT defibrillator received inappropriate ICD shocks due to T-wave oversensing. Decreasing the sensitivity to avoid T wave oversensing was not an option due to a suboptimal R-wave sensing amplitude. We decided to re-plug the LV lead in the RV port and the RV lead in the LV port. This however led to intermittent phrenic nerve stimulation due to mandatory bipolar (tip-ring) or unipolar (tip-can) pacing on the LV-lead from the RV port. Re-intervention was necessary with the implantation of an additional pacing/sensing RV lead. A software programmable choice to switch sensing and tachycardia detection from RV to LV lead could be a valuable feature in future CRT devices.