Abstract
BACKGROUND
Despite the rapid growth of fluorescence imaging, accurate sampling of tissue sections remains challenging. Development of novel technologies to improve intraoperative assessment of tissue is needed.
METHODS
A novel contact probe-based fluorescence dosimeter device, optimized for IRDye800CW quantification, was developed. After evaluation of the device in a phantom setup, its clinical value was defined ex vivo in patients with head and neck squamous cell carcinoma who received panitumumab-IRDye800CW.
RESULTS
Ten patients were enrolled with a total of 216 data points obtained. Final histopathology showed tumor in 119 spots and normal tissue in 97 spots. Fluorescence-to-excitation ratios in tumor tissue were more than three times higher than those in normal tissue. The area under the curve was 0.86 (95% CI: 0.81-0.91) for tumor detection.
CONCLUSIONS
Fluorescence-guided tissue preselection using a fluorescence dosimeter could have substantial impact on tissue sampling for frozen section analysis and potentially reduce sampling errors.