Roland JT, Gantz BJ, Waltzman SB, Parkinson AJ. United States multicenter clinical trial of the cochlear nucleus hybrid implant system.
Laryngoscope 2015;
126:175-81. [PMID:
26152811 PMCID:
PMC4704985 DOI:
10.1002/lary.25451]
[Citation(s) in RCA: 111] [Impact Index Per Article: 11.1] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/08/2015] [Revised: 04/30/2015] [Accepted: 05/28/2015] [Indexed: 12/02/2022]
Abstract
Objectives/Hypothesis
To evaluate the safety and efficacy of acoustic and electric sound processing for individuals with significant residual low‐frequency hearing and severe‐to‐profound high‐frequency sensorineural hearing loss.
Study Design
Prospective, single‐arm repeated measures, single‐subject design.
Methods
Fifty individuals, ≥ 18 years old, with low‐frequency hearing and severe high‐frequency loss were implanted with the Cochlear Nucleus Hybrid L24 implant at 10 investigational sites. Preoperatively, subjects demonstrated consonant‐nucleus‐consonant word scores of 10% through 60% in the ear to be implanted. Subjects were assessed prospectively, preoperatively, and postoperatively on coprimary endpoints of consonant‐nucleus‐consonant words, AzBio sentences in noise, and self‐assessment measures.
Results
Significant mean improvements were observed for coprimary endpoints: consonant‐nucleus‐consonant words (35.8 percentage points) and AzBio sentences in noise (32.0 percentage points), both at P < 0.001. Ninety‐six percent of subjects performed equal or better on speech in quiet and 90% in noise. Eighty‐two percent of subjects showed improved performance on speech in quiet and 74% in noise. Self‐assessments were positive, corroborating speech perception results.
Conclusion
The Nucleus Hybrid System provides significant improvements in speech intelligibility in quiet and noise for individuals with severe high‐frequency loss and some low‐frequency hearing. This device expands indications to hearing‐impaired individuals who perform poorly with amplification due to bilateral high‐frequency hearing loss and who previously were not implant candidates.
Level of Evidence
2b. Laryngoscope, 126:175–181, 2016
Collapse