Gonzales KD, Douglas G, Pichakron KO, Kwiat DA, Gallardo SG, Encinas JL, Hirose S, Harrison MR. Magnamosis III: delivery of a magnetic compression anastomosis device using minimally invasive endoscopic techniques.
J Pediatr Surg 2012;
47:1291-5. [PMID:
22703808 DOI:
10.1016/j.jpedsurg.2012.03.042]
[Citation(s) in RCA: 42] [Impact Index Per Article: 3.2] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 03/05/2012] [Accepted: 03/06/2012] [Indexed: 12/11/2022]
Abstract
PURPOSE
Magnamosis creates a secure sutureless anastomosis through magnetic compression. In this study, we further develop and test delivery devices capable of creating a secure duodeno-colonic anastomosis using available minimally invasive endoscopic techniques.
METHODS
Eight pigs underwent general anesthesia. Colonoscopy was used to deliver 1 magnetic ring to the hepatic flexure. Simultaneously, upper endoscopy delivered the other magnetic ring into the duodenum using a variety of techniques. The 2 magnetic rings were brought into magnetic proximity under laparoscopic guidance. The pigs were recovered and examined daily followed by sacrificing at 1, 2, 4, and 6 weeks.
RESULTS
The device designed to deliver and release each magnetic ring evolved from using a guide wire and balloon devices to redesigning the magnetic ring casing with a groove to accommodate an endoscopic snare. Laparoscopic visualization assured safe magnet mating of intestinal segments. The duodeno-colonic anastomoses created with the snare yielded widely patent anastomoses. In vitro testing revealed excellent burst pressure. Histology revealed complete healing as early as 1 week.
CONCLUSION
We redesigned the magnamosis device to facilitate delivery by endoscopic techniques. The snare technique allows endoscopic positioning and controlled release of the magnetic rings for a secure side-to-side duodeno-colonic anastomosis.