Bandorski D, Gehron J, Höltgen R. Interference between pacemakers/implantable cardioverter defibrillators and video capsule endoscopy.
World J Gastrointest Endosc 2013;
5:201-202. [PMID:
23596547 PMCID:
PMC3627847 DOI:
10.4253/wjge.v5.i4.201]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 07/15/2012] [Revised: 10/03/2012] [Accepted: 01/21/2013] [Indexed: 02/05/2023] Open
Abstract
Our Letter to the Editor, related to the article “Small bowel capsule endoscopy in patients with cardiac pacemakers and implantable cardioverter defibrillators: Outcome analysis using telemetry” by Cuschieri et al, comments on some small errors, that slipped into the authors discussions. The given informations concerning the pacemaker- and implantable cardioverter defibrillators modes were inaccurate and differ between the text and the table. Moreover, as 8 of 20 patient’s pacemakers were programmed to VOO or DOO (“interference mode”) and one patient was not monitored by telemetry during capsule endoscopy, 9 of 20 patients (45%) lack the informations of possible interference between capsule endoscopy their implanted device. Another objection refers to the interpretation of an electrocardiogram (figure 1, trace B) presented: in contrast to the author’s opinion the marked spike should be interpreted as an artefact and not as ”undersensing of a fibrillatory wave”. Finally, three comments to cited reviews were not complete respectively not quoted correctly.