101
|
A simple method to shorten the unfolding time of prehydrated hydrophobic intraocular lens. Can J Ophthalmol 2014; 49:382-7. [DOI: 10.1016/j.jcjo.2014.06.002] [Citation(s) in RCA: 10] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/01/2014] [Revised: 04/02/2014] [Accepted: 06/18/2014] [Indexed: 11/22/2022]
|
102
|
Dai Y, Huang Y, Liu T, Xie L. Laboratory analyses of two explanted hydrophobic acrylic intraocular lenses. Indian J Ophthalmol 2014; 62:737-9. [PMID: 25005210 PMCID: PMC4131335 DOI: 10.4103/0301-4738.136262] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/15/2022] Open
Abstract
Two three-piece hydrophobic acrylic intraocular lenses (IOLs) were explanted from two patients at 7 and 9 years, respectively, after implantation, because of poor fundus visualisation and/or a clinically significant decrease in visual acuity related to their opacified IOLs. In addition to light microscopy, scanning electron microscopy and energy dispersive X-ray spectroscopy, confocal laser scanning microscopy was used for the first time to observe the explanted IOLs. The clinical aspect seemed to correspond to the phenomenon of surface light scattering, while laboratory analyses showed dense glistenings in the central layer of the IOL optic, which had no change next to the surface. Further studies on these phenomena are needed.
Collapse
Affiliation(s)
| | | | | | - Lixin Xie
- Shandong Provincial Key Laboratory of Ophthalmology, Shandong Eye Institute, Shandong Academy of Medical Sciences, Qingdao, China
| |
Collapse
|
103
|
Beheregaray S, Yamamoto T, Hiraoka T, Oshika T. Influence on visual function of forward light scattering associated with subsurface nanoglistenings in intraocular lenses. J Cataract Refract Surg 2014; 40:1147-54. [DOI: 10.1016/j.jcrs.2013.10.047] [Citation(s) in RCA: 17] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/14/2013] [Revised: 10/15/2013] [Accepted: 10/30/2013] [Indexed: 10/25/2022]
|
104
|
Evaluation of the impact of light scatter from glistenings in pseudophakic eyes. J Cataract Refract Surg 2014; 40:95-103. [PMID: 24355722 DOI: 10.1016/j.jcrs.2013.10.018] [Citation(s) in RCA: 25] [Impact Index Per Article: 2.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/17/2013] [Revised: 05/28/2013] [Accepted: 06/05/2013] [Indexed: 11/22/2022]
Abstract
PURPOSE To study the impact of light scatter from glistenings in pseudophakic eyes using ray tracing in a model eye SETTING Department of Research, Advanced Vision Science, Inc., Goleta, California, USA. DESIGN Mathematical modeling and simulation. METHODS A pseudophakic eye model was constructed in Zemax using the Arizona eye model as the basis. The Mie scattering theory was used to describe the intensity and direction of light as it scatters for a spherical particle immersed in a given media (intraocular lens [IOL]). The modeling and evaluation of scatter and modulation transfer function (MTF) were performed for several biomaterials with various size and density of glistenings under scotopic, mesopic, and photopic conditions. RESULTS As predicted by the Mie theory, the amount of scatter was a function of the relative difference in refractive index between the media and the scatterer, the size of the scatterer, and the volume fraction of the scatterer. The simulation demonstrated that an increase in density of glistenings can lead to a significant drop in the MTF of the IOL and the pseudophakic eye. This effect was more pronounced in IOLs with smaller cavitations, and the observation was consistent for all tested biomaterials. CONCLUSIONS Mathematical modeling demonstrated that glistenings in IOLs will lead to reduction in the MTF of the IOL and the pseudophakic eye. The loss in MTF was more pronounced at high densities and small cavitation sizes across all biomaterials. Inconsistent and poor clinical quantification of glistenings in IOLs may explain some inconsistencies in the literature.
Collapse
|
105
|
Morris C, Werner L, Barra D, Liu E, Stallings S, Floyd A. Light scattering and light transmittance of cadaver eye-explanted intraocular lenses of different materials. J Cataract Refract Surg 2014; 40:129-37. [PMID: 24355724 DOI: 10.1016/j.jcrs.2013.10.016] [Citation(s) in RCA: 10] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/07/2013] [Revised: 07/01/2013] [Accepted: 07/02/2013] [Indexed: 11/17/2022]
Abstract
PURPOSE To evaluate light scattering and light transmittance in cadaver eye-explanted intraocular lenses (IOLs) manufactured from different materials. SETTING John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA. DESIGN Experimental study. METHODS Forty-nine pseudophakic cadaver eyes were selected according to IOL material/type and implantation duration, and the IOLs were explanted. Hydrophobic acrylic, hydrophilic acrylic, poly(methyl methacrylate) (PMMA), and silicone IOLs were included. Gross and light microscopy was performed for all IOLs. Light scattering was measured with an EAS 1000 Scheimpflug camera, and light transmittance was assessed using a Lambda 35 UV/Vis spectrophotometer (single-beam configuration with an RSA PE-20 integrating sphere). Analyses were performed at room temperature in the hydrated state and compared with analyses of controls. RESULTS The highest levels of surface light scattering were measured for 3-piece hydrophobic acrylic, which was also the IOL type with the longest implantation duration among the Acrysof hydrophobic acrylic IOLs. Hydrophilic acrylic, PMMA, and silicone IOLs exhibited relatively low light-scattering levels. The lowest light-scattering levels were observed with PMMA IOLs (1-piece looped and 3-piece) and plate silicone IOLs, which represent the IOL types with the longest implantation duration in this series. Light transmittance values measured for all IOL types appeared to be similar to the values of the corresponding control IOLs. CONCLUSIONS The phenomenon of surface light scattering (nanoglistenings) is more particularly related to hydrophobic acrylic IOLs and increases with implantation time. No significant effect of surface light scattering on IOL light transmittance was found.
Collapse
Affiliation(s)
- Caleb Morris
- From the Department of Ophthalmology and Visual Sciences (Morris, Werner, Liu, Stallings, Floyd), John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA; and Visão Institutos Oftalmológicos Associados (Barra), Brasília, DF, Brazil
| | - Liliana Werner
- From the Department of Ophthalmology and Visual Sciences (Morris, Werner, Liu, Stallings, Floyd), John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA; and Visão Institutos Oftalmológicos Associados (Barra), Brasília, DF, Brazil.
| | - Daniel Barra
- From the Department of Ophthalmology and Visual Sciences (Morris, Werner, Liu, Stallings, Floyd), John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA; and Visão Institutos Oftalmológicos Associados (Barra), Brasília, DF, Brazil
| | - Erica Liu
- From the Department of Ophthalmology and Visual Sciences (Morris, Werner, Liu, Stallings, Floyd), John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA; and Visão Institutos Oftalmológicos Associados (Barra), Brasília, DF, Brazil
| | - Shannon Stallings
- From the Department of Ophthalmology and Visual Sciences (Morris, Werner, Liu, Stallings, Floyd), John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA; and Visão Institutos Oftalmológicos Associados (Barra), Brasília, DF, Brazil
| | - Anne Floyd
- From the Department of Ophthalmology and Visual Sciences (Morris, Werner, Liu, Stallings, Floyd), John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA; and Visão Institutos Oftalmológicos Associados (Barra), Brasília, DF, Brazil
| |
Collapse
|
106
|
Xi L, Liu Y, Zhao F, Chen C, Cheng B. Analysis of glistenings in hydrophobic acrylic intraocular lenses on visual performance. Int J Ophthalmol 2014; 7:446-51. [PMID: 24967189 DOI: 10.3980/j.issn.2222-3959.2014.03.11] [Citation(s) in RCA: 15] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [Key Words] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/04/2014] [Accepted: 04/05/2014] [Indexed: 11/02/2022] Open
Abstract
AIM To assess patients' visual performance with glistenings in one piece soft hydrophobic acrylic intraocular lenses (IOLs) (Alcon) 2 years postoperatively. METHODS This cross section trial included 120 eyes with one piece IOL at 2 years postoperatively. Glistening was classified in 4 groups, ranging from 0 (none) to 3 (most evident) according to their severity in IOLs optics observed under a slit lamp. All eyes underwent a uncorrected and best-corrected visual acuity evaluation (UCVA and BCVA, LogMAR scale), a complete clinical examination, a contrast sensitivity (CS) evaluation by F.A.C.T chart, and a visual field test by Humphrey Field Analyzer ‖ (HFA). One-way ANOVA was used for quantitative data, while Pearson χ (2) test was used for qualitative data to analyze the visual function of 4 glistening groups. RESULTS Totally 120 eyes were enrolled with 30 eyes in each glistening group. There was no statistical correlation between glistening grades and patients' age, IOLs power, postoperative UCVA and BCVA (P>0.05). Quantificationally, CS values among each group were not statistically different. However, qualitative analysis showed there were more eyes in grade 3 group than in grade 0 group having abnormally declined CS at high spatial frequency (10% vs 36.7% at 18 cpd, P=0.029; 6.7% vs 26.7% at 12 cpd, P=0.013). Mean deviation (MD) of the visual field test was -2.14±2.31, -1.97±2.23, -3.02±3.17, -4.12±3.38 in group 0 to 3 respectively. There was a significant decrease in the most serious glistenings group (P =0.018). CONCLUSION Glistenings may potentially have an impact on contrast sensitivity at high spatial frequency and MD in visual field test.
Collapse
Affiliation(s)
- Lei Xi
- State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou 510060, Guangdong Province, China
| | - Yi Liu
- State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou 510060, Guangdong Province, China
| | - Feng Zhao
- State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou 510060, Guangdong Province, China
| | - Chuan Chen
- State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou 510060, Guangdong Province, China
| | - Bing Cheng
- State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou 510060, Guangdong Province, China
| |
Collapse
|
107
|
Post-DSAEK optical changes: a comprehensive prospective analysis on the role of ocular wavefront aberrations, haze, and corneal thickness. Cornea 2014; 32:1567-77. [PMID: 24162748 DOI: 10.1097/ico.0b013e3182a9b182] [Citation(s) in RCA: 35] [Impact Index Per Article: 3.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/25/2022]
Abstract
PURPOSE The aim was to assess the visual impact of ocular wavefront aberrations, corneal thickness, and corneal light scatter prospectively after performing a Descemet stripping automated endothelial keratoplasty (DSAEK) in humans. METHODS Data were obtained prospectively from 20 eyes preoperatively and at 1, 3, 6, and 12 months post-DSAEK. At each visit, the best spectacle-corrected visual acuity and visual acuity with glare (brightness acuity testing) were recorded, and ocular wavefront measurements and corneal optical coherence tomography (OCT) were performed. The magnitude and the sign of individual Zernike terms [higher-order aberrations (HOAs)] were determined. Epithelial, host stromal, donor stromal, and total corneal thicknesses were quantified. The brightness and intensity profiles of OCT images were generated to quantify light scatter in the whole cornea, subepithelial region, anterior and posterior host stroma, interface, and donor stroma. RESULTS The mean best spectacle-corrected visual acuity and glare disability at low light levels improved from 1 to 12 months post-DSAEK. All corneal thicknesses and ocular lower-order aberrations and HOAs were found to be stable from 1 to 12 months, whereas total corneal, host stromal, and interface brightness intensities decreased significantly over the same period. A repeated measures analysis of variance performed across the follow-up period revealed that the change in scatter, but not the change in the HOAs, could account for the variability occurring in the acuity from 1 to 12 months post-DSAEK. CONCLUSIONS Although ocular HOAs and scatter are both elevated over normal values post-DSAEK, our results demonstrate that the improvements in visual performance occurring over the first year post-DSAEK are associated with decreasing light scatter. In contrast, there were no significant changes in the ocular HOAs during this time. Because corneal light scatter decreased between 1 and 12 months despite there being stable corneal thicknesses over the same period, we conclude that factors that induced light scatter, other than tissue thickness or swelling (corneal edema), significantly impacted the visual improvements that occurred over time post-DSAEK. A better understanding of the cellular and extracellular matrix changes of the subepithelial region and interface, incurred by the surgical creation of a lamellar host-graft interface, and the subsequent healing of these tissues, is warranted.
Collapse
|
108
|
Comparison of the Retinal Straylight in Pseudophakic Eyes with PMMA, Hydrophobic Acrylic, and Hydrophilic Acrylic Spherical Intraocular Lens. J Ophthalmol 2014; 2014:340759. [PMID: 24724015 PMCID: PMC3960563 DOI: 10.1155/2014/340759] [Citation(s) in RCA: 7] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/16/2013] [Accepted: 01/28/2014] [Indexed: 11/21/2022] Open
Abstract
Purpose. To investigate the intraocular straylight value after cataract surgery.
Methods. In this study, 76 eyes from 62 patients were subdivided into three groups. A hydrophobic acrylic, a hydrophilic acrylic, and a PMMA IOL were respectively, implanted in 24 eyes, 28 eyes, and 24 eyes. Straylight was measured using C-Quant at 1 week and 1 month postoperatively in natural and dilated pupils. Results. The hydrophilic acrylic IOLs showed significantly lower straylight values than those of the hydrophobic acrylic IOLs in dilated pupils at 1 week and 1 month after surgery (P < 0.05). However, the straylight values of the hydrophilic acrylic IOLs were the lowest among the three IOL groups. No significant difference was observed in straylight between 1 week and 1 month postoperatively in each group with natural and dilated pupils (P > 0.05). Moreover, no significant difference was found in straylight between natural and dilated pupils in each group at 1 week and 1 month postoperatively (P > 0.05). Conclusions. Although the hydrophobic acrylic IOL induced more intraocular straylight, straylight differences among the 3 IOLs were minimal. Pupil size showed no effect on intraocular straylight; the intraocular straylight was stable 1 week after surgery.
Collapse
|
109
|
Minami K, Maruyama Y, Honbo M, Matsunaga J, Miyata K. In vitro examination of surface light scattering in hydrophobic acrylic intraocular lenses. J Cataract Refract Surg 2014; 40:652-6. [DOI: 10.1016/j.jcrs.2013.10.034] [Citation(s) in RCA: 9] [Impact Index Per Article: 0.9] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/17/2013] [Revised: 09/28/2013] [Accepted: 10/20/2013] [Indexed: 11/27/2022]
|
110
|
Heiner P, Ligabue E, Fan A, Lam D. Safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (enVista®) - results of a European and Asian-Pacific study. Clin Ophthalmol 2014; 8:629-35. [PMID: 24729678 PMCID: PMC3974695 DOI: 10.2147/opth.s56135] [Citation(s) in RCA: 15] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/23/2022] Open
Abstract
Purpose To evaluate the safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (IOL) (enVista® MX60; Bausch and Lomb Incorporated, Rochester, NY, USA) following implantation to correct aphakia subsequent to extracapsular cataract extraction in adults. Subjects and methods This was an open-label, non-interventional, observational study conducted in 19 university and private-practice settings in Europe and the Asia-Pacific region to investigate clinical outcomes of the MX60 IOL in standard practice. Eligible subjects were at least 18 years of age and had undergone standard phacoemulsification and extracapsular cataract extraction with implantation of the MX60 IOL. The primary safety endpoint was the occurrence of adverse events, and the primary effectiveness endpoints included visual and refractive outcomes and stability, with data collected up to 2 years post-procedure. Results In this multicenter study, pooled data of 255 eyes were collected and analyzed. Excellent visual and refractive outcomes and stability were demonstrated. At postoperative visit 4 (61–180 days postoperative), 62.2% of subjects achieved a Snellen best-corrected distance visual acuity (CDVA) of 20/20 (decimal 1.00), and 97.8% of subjects achieved a CDVA of 20/40 (decimal 0.50) or better. One eye (1.0%) underwent neodymium:yttrium aluminum garnet capsulotomy at 12 months post-procedure. No glistenings of any grade were reported for any subject at any visit. Adverse events were infrequent and were consistent with incidences generally reported with cataract surgery. Conclusion This study, which enrolled all comers, provided evidence of the excellent safety and effectiveness of the MX60 IOL in standard practice. Favorable clinical outcomes included outstanding visual and refractive outcomes and stability. No glistenings were reported at any postoperative visit.
Collapse
Affiliation(s)
| | | | - Alex Fan
- Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Shatin, New Territories, Hong Kong
| | - Dennis Lam
- Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Shatin, New Territories, Hong Kong
| |
Collapse
|
111
|
Packer M, Rajan M, Ligabue E, Heiner P. Clinical properties of a novel, glistening-free, single-piece, hydrophobic acrylic IOL. Clin Ophthalmol 2014; 8:421-7. [PMID: 24591811 PMCID: PMC3937253 DOI: 10.2147/opth.s57114] [Citation(s) in RCA: 26] [Impact Index Per Article: 2.6] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/05/2022] Open
Abstract
A new, single-piece, hydrophobic acrylic lens – the first constructed with a lens optic and haptics comprised of a hydroxyethyl methacrylate-polyethylene glycol phenyl ether acrylate–styrene copolymer, cross-linked with ethylene glycol dimethacrylate, and labeled as “glistening-free” – was recently introduced globally. Glistenings have been a significant source of clinical concern with previous hydrophobic lens designs. This new monofocal lens provides enhanced, clear optics for lens-based surgery. The superior optical clarity of this lens is achieved through the elimination of glistenings, enhanced surface durability, high refractive index, a high Abbe number, and an aspheric design. Additionally, the lens design reduces the risk of developing posterior capsule opacification.
Collapse
Affiliation(s)
- Mark Packer
- Oregon Health and Science University, Portland, OR, USA
| | - Mohan Rajan
- Rajan Eye Care Hospital Pvt Ltd, Chennai, India
| | - Edoardo Ligabue
- Ophthalmology Department, Centro Diagnostico Italiano, Milan, Italy
| | - Peter Heiner
- Bond University Medical School, Robina, QLD, Australia
| |
Collapse
|
112
|
Packer M. Multifocal intraocular lens technology: biomaterial, optical design and review of clinical outcomes. EXPERT REVIEW OF OPHTHALMOLOGY 2014. [DOI: 10.1586/eop.11.40] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/08/2022]
|
113
|
Light transmittance of 1-piece hydrophobic acrylic intraocular lenses with surface light scattering removed from cadaver eyes. J Cataract Refract Surg 2014; 40:114-20. [DOI: 10.1016/j.jcrs.2013.05.050] [Citation(s) in RCA: 14] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/30/2013] [Revised: 04/29/2013] [Accepted: 05/23/2013] [Indexed: 11/19/2022]
|
114
|
Brockmann T, Brockmann C, Nietzsche S, Bertelmann E, Strobel J, Dawczynski J. Scanning electron microscopic characteristics of commercially available 1- and 3-piece intraocular lenses. J Cataract Refract Surg 2013; 39:1893-9. [DOI: 10.1016/j.jcrs.2013.04.046] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/26/2022]
|
115
|
Bissen-Miyajima H, Minami K, Yoshino M, Taira Y. Surface light scattering and visual function of diffractive multifocal hydrophobic acrylic intraocular lenses 6 years after implantation. J Cataract Refract Surg 2013; 39:1729-33. [DOI: 10.1016/j.jcrs.2013.05.041] [Citation(s) in RCA: 8] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/22/2013] [Revised: 05/17/2013] [Accepted: 05/21/2013] [Indexed: 10/26/2022]
|
116
|
Packer M, Fry L, Lavery KT, Lehmann R, McDonald J, Nichamin L, Bearie B, Hayashida J, Altmann GE, Khodai O. Safety and effectiveness of a glistening-free single-piece hydrophobic acrylic intraocular lens (enVista). Clin Ophthalmol 2013; 7:1905-12. [PMID: 24109169 PMCID: PMC3792945 DOI: 10.2147/opth.s50499] [Citation(s) in RCA: 27] [Impact Index Per Article: 2.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/02/2022] Open
Abstract
Purpose To evaluate the safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (IOL; enVista model MX60; Bausch & Lomb, Rochester, NY, USA) when used to correct aphakia following cataract extraction in adults. Methods This was a prospective case series (NCT01230060) conducted in private practices in the US. Eligible subjects were adult patients with age-related cataract amenable to treatment with standard phacoemulsification/extracapsular cataract extraction. With follow-up of 6 months, primary safety and effectiveness end points included the rates of US Food and Drug Administration (FDA)-defined cumulative and persistent adverse events and the percentage of subjects who achieved best-corrected visual acuity (BCVA) of 20/40 or better at final visit. To evaluate rotational stability, subjects were randomized (1:1:1:1) to have the lens implanted in one of four axis positions in 45° increments. Results A total of 122 subjects were enrolled. The rate of cumulative and persistent adverse events did not significantly exceed historical controls, as per FDA draft guidance. At the final postoperative visit, all subjects (100%) achieved a BCVA of 20/40 compared with the FDA historical control of 96.7%. Rotation of the IOL between the two final follow-up visits was ≤5° for 100% of eyes, and refractive stability was demonstrated. A low evaluation of posterior capsule opacification score was demonstrated, and no glistenings of any grade were reported for any subject at any visit. Conclusion This study demonstrated the safety and effectiveness of the MX60 IOL. Favorable clinical outcomes included preserved BCVA, excellent rotational and refractive stability, no glistenings, and a low evaluation of posterior capsule opacification score.
Collapse
Affiliation(s)
- Mark Packer
- Department of Ophthalmology, Oregon Health and Science University, Eugene, OR, USA
| | | | | | | | | | | | | | | | | | | |
Collapse
|
117
|
Hennig A, Schroeder B, Gilbert C. Bilateral pediatric cataract surgery: outcomes of 390 children from Nepal and Northern India. J Pediatr Ophthalmol Strabismus 2013; 50:312-9. [PMID: 23565713 DOI: 10.3928/01913913-20130402-01] [Citation(s) in RCA: 12] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 09/14/2012] [Accepted: 03/04/2013] [Indexed: 11/20/2022]
Abstract
PURPOSE To report the outcomes of bilateral pediatric cataract surgery from eastern Nepal and northern India. METHODS Preoperative, intraoperative, and postoperative data of 390 children who underwent surgery bilaterally between 2007 and 2009 were analyzed. RESULTS Forty-two (10.8%) children came from Nepal and 348 (89.2%) from India (mainly Bihar State). Intraocular lens (IOL) implantation with posterior capsule opening and anterior vitrectomy were achieved in 386 (99.0%) children bilaterally. Median age at surgery was 7 years and 69.2% were male. At first presentation, 243 (62.3%) of the children were blind (< 3/60 in the better eye). After more than 1 year, 53.5% had a normal visual status (range: 6/6 to 6/18), 5.6% of children were still blind, and mean refractive error spherical equivalent was +1.0 ± 2.4 diopters. Astigmatism changed from suture-induced with the rule at discharge to against the rule within 3 weeks of surgery. Mean long-term astigmatic error was 1.0 ± 0.9 diopters after 1 year. Glaucoma was rare. CONCLUSIONS Even in a setting with limited resources, successful, cost-effective, high-volume surgery for pediatric cataract is possible. Despite late presentation and limited follow-up, more than half achieved good outcomes after more than 1 year. Only 5.6% remained blind due to amblyopia or eye anomalies. Bilateral surgery during one hospital stay, IOL implantation with undercorrection according to age, aggressive surgery to prevent secondary cataract, intensive anti-inflammatory therapy, and provision of durable, high-quality spectacles to take home all proved beneficial because many children cannot attend for regular follow-up.
Collapse
|
118
|
Bozukova D, Pagnoulle C, Jérôme C. Biomechanical and optical properties of 2 new hydrophobic platforms for intraocular lenses. J Cataract Refract Surg 2013; 39:1404-14. [DOI: 10.1016/j.jcrs.2013.01.050] [Citation(s) in RCA: 34] [Impact Index Per Article: 3.1] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/06/2012] [Revised: 01/09/2013] [Accepted: 01/27/2013] [Indexed: 10/26/2022]
|
119
|
Leydolt C, Schriefl S, Stifter E, Haszcz A, Menapace R. Posterior capsule opacification with the iMics1 NY-60 and AcrySof SN60WF 1-piece hydrophobic acrylic intraocular lenses: 3-year results of a randomized trial. Am J Ophthalmol 2013; 156:375-381.e2. [PMID: 23677137 DOI: 10.1016/j.ajo.2013.04.007] [Citation(s) in RCA: 30] [Impact Index Per Article: 2.7] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/23/2012] [Revised: 04/04/2013] [Accepted: 04/05/2013] [Indexed: 11/25/2022]
Abstract
PURPOSE To compare the intensity of posterior capsule opacification (PCO) 3 years after implantation of 2 different 1-piece foldable hydrophobic acrylic intraocular lenses (IOLs). DESIGN Randomized, prospective, patient- and examiner-masked clinical trial with intraindividual comparison. METHODS One hundred patients with bilateral age-related cataract (200 eyes) had standard cataract surgery with implantation of an iMics1 NY-60 IOL (Hoya Corp) in one eye and an AcrySof SN60WF IOL (Alcon Laboratories) in the other eye. Follow-up examinations were performed at 1 week and 3 years. Digital retroillumination images were obtained of each eye. The main outcome measure was PCO score (scale, 0 to 10) assessed subjectively at the slit lamp and objectively using automated image analysis software (Automated Quantification of After-Cataract) 3 years after surgery. RESULTS The objective PCO score (mean ± standard deviation) was 3.0 ± 2.0 for the iMics1 NY-60 IOL and 1.9 ± 1.4 for the AcrySof SN60WF IOL (P < .001). Three years after surgery, 35.6% of patients underwent a neodymium:yttrium-aluminum-garnet capsulotomy in the iMics1 NY-60 eye and 13.7% underwent a capsulotomy in the AcrySof SN60WF eye (P = .001). There was no statistically significant difference in best-corrected visual acuity, rhexis-IOL overlap, capsular folds, or anterior capsule opacification. Glistening formations were found in no iMics1 NY-60 IOLs, but in 97% of the AcrySof SN60WF IOLs. CONCLUSIONS Comparison of 2 sharp-edged single-piece IOLs of similar design and hydrophobic acrylic material indicated a statistically significant difference in PCO and neodymium:yttrium-aluminum-garnet capsulotomy rate 3 years after surgery.
Collapse
|
120
|
Thomes BE, Callaghan TA. Evaluation of in vitro glistening formation in hydrophobic acrylic intraocular lenses. Clin Ophthalmol 2013; 7:1529-34. [PMID: 23926419 PMCID: PMC3732199 DOI: 10.2147/opth.s44208] [Citation(s) in RCA: 35] [Impact Index Per Article: 3.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/25/2022] Open
Abstract
Purpose To evaluate the impact of continuous manufacturing process improvements on glistening formation in AcrySof hydrophobic acrylic intraocular lenses (IOLs) by comparing lenses manufactured in 2003 with lenses manufactured in 2012. Methods Glistenings were experimentally created as aqueous-filled microvacuoles utilizing an accelerated laboratory method by immersing the lens in water at 45°C for 24 hours and then reducing the temperature to 37°C for 2.5 hours. To determine the impact of continuous improvements on glistening formation, AcrySof IOLs (Alcon Laboratories, Inc) Model SB30 AL manufactured in 2003 were compared with AcrySof IOLs Model SN60 WF manufactured in 2012. Results Glistenings were present in all lenses after the accelerated microvacuole test method. The mean microvacuole density for IOLs manufactured in 2003 was 315.7 microvacuoles per square millimeter (MVs/mm2) with a glistening severity less than 100 MVs/mm2 in 1.0% of the IOLs. The mean microvacuole density for IOLs manufactured in 2012 was 39.9 MVs/mm2 with a glistening severity less than 100 MVs/mm2 in 95.2% of the IOLs. Laboratory-induced microvacuole density was significantly lower in IOLs manufactured in 2012 (mean 39.9 MVs/mm2) relative to IOLs manufactured in 2003 (mean 315.7 MVs/mm2) as indicated by the Wilcoxon test of significance (P < 0.0005). Conclusion AcrySof IOLs recently manufactured in 2012 demonstrated a significant reduction in glistening density (87% reduction in mean density) as a result of continuous manufacturing process improvements compared with IOLs manufactured in 2003.
Collapse
Affiliation(s)
- Brett E Thomes
- Alcon Research Ltd (A Novartis Company), Fort Worth, TX, USA
| | | |
Collapse
|
121
|
van der Mooren M, Franssen L, Piers P. Effects of glistenings in intraocular lenses. BIOMEDICAL OPTICS EXPRESS 2013; 4:1294-1304. [PMID: 24009993 PMCID: PMC3756585 DOI: 10.1364/boe.4.001294] [Citation(s) in RCA: 21] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Subscribe] [Scholar Register] [Received: 05/21/2013] [Revised: 06/27/2013] [Accepted: 07/08/2013] [Indexed: 06/02/2023]
Abstract
Glistenings consist of multiple microvacuoles in intraocular lenses (IOLs) that cause retinal stray light and may affect quality of vision. For four IOL types, the microvacuole particle size distribution and particle volume density was measured using confocal light microscopy and dark field microscopy, and the corresponding extinction coefficient γ was determined. The light scatter contribution induced by microvacuoles was measured as function of both angle and extinction, and was verified by calculations using Mie theory. Two IOL types possessed significant glistenings having stray light levels higher than that of a healthy 20 year old crystalline lens corresponding to γ ≥ 0.08 mm(-1).
Collapse
|
122
|
Bodnar ZM, Rozot P, Leishman L, Ollerton A, Michelson J, Plasse-Fauque S, Werner L. Calcification of a hydrophilic acrylic intraocular lens: case report with laboratory analysis. J Fr Ophtalmol 2013; 36:e113-7. [PMID: 23688611 DOI: 10.1016/j.jfo.2012.11.012] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/26/2012] [Revised: 11/05/2012] [Accepted: 11/07/2012] [Indexed: 11/26/2022]
Abstract
We analyzed a single-piece plate-type hydrophilic acrylic posterior chamber intraocular lens (IOL) that was explanted due to a progressive loss of vision, which occurred 6 years after uncomplicated phacoemulsification. Gross and light microscopy, as well as anterior segment optical coherence tomography (OCT) revealed granular deposits below the IOL surface. Light scattering, as measured with Scheimpflug photography and densitometry analyses was found to be increased; spectrophotometry demonstrated a decrease in the light transmittance of the explanted lens. The granular deposits within the IOL material were found to be composed of calcium by histochemical methods (alizarin red and Von Kossa stains). To our knowledge this is the only report of calcification of this IOL design.
Collapse
Affiliation(s)
- Z M Bodnar
- John A.-Moran Eye Center, University of Utah, 65, Mario-Capecchi Drive, Salt Lake City, 84132 Utah, USA
| | | | | | | | | | | | | |
Collapse
|
123
|
Rønbeck M, Behndig A, Taube M, Koivula A, Kugelberg M. Comparison of glistenings in intraocular lenses with three different materials: 12-year follow-up. Acta Ophthalmol 2013; 91:66-70. [PMID: 22035345 DOI: 10.1111/j.1755-3768.2011.02277.x] [Citation(s) in RCA: 27] [Impact Index Per Article: 2.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/26/2022]
Abstract
PURPOSE To compare the degree of lens glistenings associated with three intraocular lenses (IOLs) of different materials and examine the relationship between the dioptric power of the optics and lens glistenings in a long-term study. SETTING St. Erik Eye Hospital, Stockholm, Sweden. METHODS Forty-six eyes of 46 patients underwent standard phacoemulsification and implantation with a heparin-surface-modified (HSM) polymethylmethacrylate (PMMA) IOL, a silicone IOL or a hydrophobic acrylic IOL. Evaluations of the patients and the glistenings were conducted 11.3-13.4 years postoperatively. The glistenings were examined using Scheimpflug imaging and subsequently analysed using an image analysis program. RESULTS The median follow-up time was 12.2 years (range, 11.3-13.4). The hydrophobic acrylic IOL had significantly more lens glistenings than the silicone (p = 0.003) and the PMMA (p = 0.000) IOLs. The silicone IOL had significantly more lens glistenings than the PMMA lens (p = 0.048). The IOL power did not affect the degree of lens glistenings in the hydrophobic acrylic IOL group (p = 0.64). The other groups had too little lens glistenings to evaluate the relationship. CONCLUSION In this long-term follow-up study, the hydrophobic acrylic IOL had a significantly higher degree of lens glistenings compared to the silicone and PMMA IOLs. The PMMA IOL had almost no lens glistenings. The IOL dioptric power was not significantly correlated with the degree of lens glistenings associated with the hydrophobic acrylic IOL.
Collapse
|
124
|
|
125
|
Ollerton A, Werner L, Fuller SR, Kavoussi SC, McIntyre SJ, Mamalis N. Evaluation of a new single-piece 4% water content hydrophobic acrylic intraocular lens in the rabbit model. J Cataract Refract Surg 2012; 38:1827-32. [DOI: 10.1016/j.jcrs.2012.05.039] [Citation(s) in RCA: 16] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/13/2012] [Revised: 04/02/2012] [Accepted: 05/02/2012] [Indexed: 10/28/2022]
|
126
|
Colin J, Praud D, Touboul D, Schweitzer C. Incidence of glistenings with the latest generation of yellow-tinted hydrophobic acrylic intraocular lenses. J Cataract Refract Surg 2012; 38:1140-6. [PMID: 22727284 DOI: 10.1016/j.jcrs.2012.01.031] [Citation(s) in RCA: 40] [Impact Index Per Article: 3.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/03/2011] [Revised: 01/30/2012] [Accepted: 01/31/2012] [Indexed: 10/28/2022]
Abstract
PURPOSE To evaluate the incidence of glistenings in a large series of consecutive eyes with a blue light-filtering hydrophobic acrylic intraocular lens (IOL) and the relationship with parameters such as age, sex, follow-up length, and corrected distance visual acuity (CDVA). SETTING Service d'Ophtalmologie, Centre Hospitalier et Universitaire Bordeaux, Bordeaux, France. DESIGN Cross-sectional study. METHODS This study included 111 eyes of 74 patients (age range 33 to 86 years). All cases had cataract surgery with Acrysof SN60WF monofocal IOL implantation and had a routine postoperative examination between March 2011 and June 2011. The incidence and severity of glistenings were evaluated with the slitlamp. Also evaluated was their relationship with age, sex, follow-up length, IOL power, and CDVA. Glistenings were subjectively graded (0 = absent; 1 = moderate; 2 = dense). RESULTS Glistenings occurred in 96 eyes (86.5%). Glistenings were of grade 1 severity in 45 eyes (40.5%) and of grade 2 severity in 51 eyes (45.9%). The follow-up was significantly longer in eyes with grade 2 glistenings (P ≤ .01). A limited, but significant, correlation was found between glistening severity and length of follow-up (r = 0.32, P<.01). Although there was a trend toward decreased visual acuities at higher glistening grades (r = 0.22, P = .01), there were no significant differences in CDVA between the glistening severity groups (P = .14). CONCLUSIONS Glistenings were common in eyes with the blue light-filtering hydrophobic acrylic IOL and increased over time. FINANCIAL DISCLOSURE No author has a financial or proprietary interest in any material or method mentioned.
Collapse
Affiliation(s)
- Joseph Colin
- Service d'Ophtalmologie, CHU Pellegrin, Bordeaux, France
| | | | | | | |
Collapse
|
127
|
Werner L, Michelson J, Ollerton A, Leishman L, Bodnar Z. Anterior segment optical coherence tomography in the assessment of postoperative intraocular lens optic changes. J Cataract Refract Surg 2012; 38:1077-85. [PMID: 22624909 DOI: 10.1016/j.jcrs.2012.01.027] [Citation(s) in RCA: 17] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/14/2011] [Revised: 01/10/2012] [Accepted: 01/10/2012] [Indexed: 11/19/2022]
Abstract
PURPOSE To evaluate the use of anterior segment optical coherence tomography (AS-OCT) to assess postoperative intraocular lens (IOL) optic changes. SETTING John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA. DESIGN Experimental study. METHODS Intraocular lenses explanted because of various complications were used, including poly(methyl methacrylate) (PMMA) IOLs with snowflake degeneration, hydrophilic acrylic IOLs with calcification, a silicone IOL with calcification from an eye with asteroid hyalosis, and hydrophobic acrylic IOLs explanted because of decentration, subluxation, or uveitis-glaucoma-hyphema syndrome. After gross and light microscopy, the IOLs were examined in the dry and hydrated states using AS-OCT. Selected hydrophilic acrylic IOLs were stained for calcium. In-the-bag IOLs in pseudophakic cadaver eyes were also evaluated by AS-OCT before and after explantation to confirm correspondence with the clinical situation. RESULTS Intraoptic changes, such as snowflake lesions in PMMA IOLs, calcification in hydrophilic acrylic IOLs, and glistenings in hydrophobic acrylic IOLs, could be imaged by AS-OCT. The method was also helpful in analyzing the location and density. However, in cases of more superficial changes, unless the lesions/deposits were present on the optic surface with an extension to the optic substance of at least 0.1 mm, they could not be clearly differentiated from the overall outline of the IOL surface. CONCLUSIONS Anterior segment OCT may be helpful in assessing the presence, location, and density of intraoptic changes, avoiding a misdiagnosis of IOL opacification and the performance of unnecessary procedures, such as posterior capsulotomy or vitrectomy.
Collapse
Affiliation(s)
- Liliana Werner
- Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, Utah 84132, USA.
| | | | | | | | | |
Collapse
|
128
|
Michelson J, Werner L, Ollerton A, Leishman L, Bodnar Z. Light scattering and light transmittance in intraocular lenses explanted because of optic opacification. J Cataract Refract Surg 2012; 38:1476-85. [DOI: 10.1016/j.jcrs.2012.03.038] [Citation(s) in RCA: 29] [Impact Index Per Article: 2.4] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/31/2012] [Revised: 03/13/2012] [Accepted: 03/14/2012] [Indexed: 10/28/2022]
|
129
|
Hayashi K, Hirata A, Yoshida M, Yoshimura K, Hayashi H. Long-term effect of surface light scattering and glistenings of intraocular lenses on visual function. Am J Ophthalmol 2012; 154:240-251.e2. [PMID: 22633349 DOI: 10.1016/j.ajo.2012.03.011] [Citation(s) in RCA: 52] [Impact Index Per Article: 4.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/18/2012] [Revised: 03/08/2012] [Accepted: 03/08/2012] [Indexed: 10/28/2022]
Abstract
PURPOSE To investigate the long-term effect of surface light scattering and glistenings of various intraocular lenses (IOLs) on visual function and optical aberrations after cataract surgery. DESIGN Case-control study. METHODS Thirty-five eyes that underwent implantation of a hydrophobic acrylic, silicone, or polymethyl methacrylate (PMMA) IOL more than 10 years ago were recruited. The scattering light intensity of the surface and internal matrix of the optic was measured using Scheimpflug photography. Visual acuity (VA) was measured using VA charts, and contrast VA and that with glare (glare VA) were examined using a contrast sensitivity tester. Ocular higher-order aberrations (HOAs) were measured using a Hartmann-Shack aberrometer. RESULTS Mean scattering light intensity of the surface and internal matrix of the optic was significantly higher in the acrylic group than in the silicone and PMMA groups (P < .0001). Mean uncorrected VA, photopic and mesopic contrast VA and glare VA, and HOAs did not differ significantly among groups, although mean corrected VA in the acrylic group was significantly better than that in the other groups (P = .0023). Scattering light intensity of the surface and internal matrix did not correlate with VA, contrast VA, or glare VA, and did not correlate with ocular and internal optic HOAs in the acrylic group. CONCLUSIONS At more than 10 years postoperatively, visual function, including contrast sensitivity, and ocular HOAs were comparable among eyes that received acrylic, silicone, and PMMA IOLs. Surface scattering and glistenings with the acrylic IOLs were not significantly correlated with visual function and optical aberrations.
Collapse
|
130
|
Pagnoulle C, Bozukova D, Gobin L, Bertrand V, Gillet-De Pauw MC. Assessment of new-generation glistening-free hydrophobic acrylic intraocular lens material. J Cataract Refract Surg 2012; 38:1271-7. [DOI: 10.1016/j.jcrs.2012.02.041] [Citation(s) in RCA: 23] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/21/2011] [Revised: 02/03/2012] [Accepted: 02/03/2012] [Indexed: 11/28/2022]
|
131
|
|
132
|
Dhital A, Spalton DJ, Goyal S, Werner L. Calcification in hydrophilic intraocular lenses associated with injection of intraocular gas. Am J Ophthalmol 2012; 153:1154-60.e1. [PMID: 22325304 DOI: 10.1016/j.ajo.2011.11.017] [Citation(s) in RCA: 54] [Impact Index Per Article: 4.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/01/2011] [Revised: 11/08/2011] [Accepted: 11/09/2011] [Indexed: 10/14/2022]
Abstract
PURPOSE To report a distinct type of calcification in hydrophilic intraocular lenses in complicated, traumatized eyes with a history of intraocular gas use. DESIGN Observational case series. METHODS Three cases of hydrophilic intraocular lens (IOL) opacification confined to the pupillary area are reported from clinical practice in London, UK. Clinical details and analysis of the explanted intraocular lenses are provided with environmental scanning electron microscopy images and x-ray energy-dispersive spectroscopy results. RESULTS All cases were associated with use of intraocular gas in complicated traumatized eyes, and had central areas of IOL opacification over the pupillary zone, confined to the anterior surface of the IOL. Analysis of the lenses showed the opacified areas to be composed of calcium and phosphate. CONCLUSION The areas of opacification in all 3 hydrophilic IOLs were attributable to calcification. We postulate that intraocular gas use and the altered blood-aqueous barrier of these complicated traumatized eyes co-act to trigger secondary IOL calcification. Further experimental testing is needed to confirm this clinical association.
Collapse
|
133
|
Lundström M, Barry P, Henry Y, Rosen P, Stenevi U. Evidence-based guidelines for cataract surgery: Guidelines based on data in the European Registry of Quality Outcomes for Cataract and Refractive Surgery database. J Cataract Refract Surg 2012; 38:1086-93. [DOI: 10.1016/j.jcrs.2012.03.006] [Citation(s) in RCA: 131] [Impact Index Per Article: 10.9] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/28/2011] [Accepted: 12/20/2011] [Indexed: 10/28/2022]
|
134
|
Tanaka K, Kakisu K, Okabe T, Kobayakawa S, Tochikubo T. Calcification and Membrane Formation on the Surface of Intraocular Lenses in a Rabbit Model. Curr Eye Res 2012; 37:471-8. [DOI: 10.3109/02713683.2011.648254] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/13/2022]
|
135
|
Miyata K, Honbo M, Otani S, Nejima R, Minami K. Effect on visual acuity of increased surface light scattering in intraocular lenses. J Cataract Refract Surg 2012; 38:221-6. [PMID: 22322162 DOI: 10.1016/j.jcrs.2011.08.042] [Citation(s) in RCA: 34] [Impact Index Per Article: 2.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/05/2011] [Revised: 08/02/2011] [Accepted: 08/18/2011] [Indexed: 10/14/2022]
Abstract
PURPOSE To assess the long-term increase in surface light scattering of foldable hydrophobic acrylic intraocular lenses (IOLs) and to evaluate its effect on visual acuity. SETTING Miyata Eye Hospital, Miyazaki, Japan. DESIGN Cross-sectional study. METHODS The clinical records of patients who received an Acrysof IOL (study group) or a Sensar AR40 or AR40e IOL (control group) were reviewed retrospectively. Surface light scattering of IOLs was measured 1 year postoperatively or later. The corrected distance visual acuity (CDVA) at the surface-light-scattering examination was compared with the CDVA 1 month after IOL implantation. The changes in CDVA over time were compared with the increase in surface light scattering. RESULTS The study evaluated 466 eyes of 337 patients. Surface light scattering in the study group continued to increase up to 15 years postoperatively. The light scattering was higher on the anterior IOL surface after 4.5 years. Increased surface light scattering had no significant impact on CDVA; however, there were more cases with decreased CDVA when the surface light scattering exceeded 50 computer-compatible tape steps. CONCLUSION Surface light scattering with the implanted foldable hydrophobic acrylic IOL continuously increased during the postoperative years and was a risk for decreased visual acuity.
Collapse
|
136
|
Khoramnia R, Salgado J, Auffarth G, Schmidt S, Wegner A, Kobuch K, Winkler von Mohrenfels C. Eintrübung einer hydrophilen Intraokularlinse 4 Jahre nach Kataraktoperation. Ophthalmologe 2012; 109:483-6. [DOI: 10.1007/s00347-011-2487-6] [Citation(s) in RCA: 16] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/29/2022]
|
137
|
Kim DH, James RH, Landry RJ, Calogero D, Anderson J, Ilev IK. Quantification of glistenings in intraocular lenses using a ballistic-photon removing integrating-sphere method. APPLIED OPTICS 2011; 50:6461-7. [PMID: 22193122 PMCID: PMC10961905 DOI: 10.1364/ao.50.006461] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/31/2023]
Abstract
An alternative method for quantification of glistenings in intraocular lenses (IOLs) using an integrating sphere with an adjustable back aperture to remove ballistic photons is presented. Glistenings in soft IOLs have been known for more than a decade; however, their severity and visual impact are still under investigation. A number of studies have been made to quantitatively describe glistenings in IOLs. Quantization and precise grading of IOLs will provide needed information to evaluate the severity and visual impact of glistenings in patients. We investigated the use of a simple modification of an integrating-sphere method to eliminate ballistic photons to quantitatively measure scattered light from glistenings in IOLs. The method described in this paper provides a simple and effective way to quantitatively characterize glistenings in vitro. It may be especially useful to quantify scattering associated with low-grade glistenings where the density of the scattering centers is low. Finally, the modified integrating-sphere method may also be generally applicable to quantitatively characterize scattering from other optical media.
Collapse
Affiliation(s)
- Do-Hyun Kim
- Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration, 10903 New Hampshire Avenue, Silver Spring, Maryland 20993, USA. do‐
| | | | | | | | | | | |
Collapse
|