Guo T, Ma Y, Jerban S, Jang H, Zhao W, Chang EY, Chen M, Bydder GM, Du J. T
1 measurement of bound water in cortical bone using 3D adiabatic inversion recovery ultrashort echo time (3D IR-UTE) Cones imaging.
Magn Reson Med 2019;
84:634-645. [PMID:
31863519 DOI:
10.1002/mrm.28140]
[Citation(s) in RCA: 7] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/14/2019] [Revised: 11/04/2019] [Accepted: 11/30/2019] [Indexed: 11/11/2022]
Abstract
PURPOSE
We describe the measurement of bound water T1 ( T 1 BW ) of cortical bone in vitro and in vivo with a 3D adiabatic inversion recovery ultrashort echo time (IR-UTE) Cones sequence using a clinical 3T scanner.
METHODS
A series IR-UTE data from 6 repetition times (TRs) with 5 inversion times (TIs) at each TR were acquired from 12 human tibial bone specimens, and data from 4 TRs with 5 TIs at each TR were acquired from the tibial midshafts of 8 healthy volunteers. The pore water nulling point was calculated from exponential fitting of the inversion recovery curve at each TR. Bone specimens and volunteers were then scanned again with the calculated nulling point at each TR. T 1 BW was derived through exponential fitting of data from IR-UTE images acquired at different TRs using the calculated pore water nulling point for each TR.
RESULTS
In vitro pore water nulling TIs were 141.3 ± 11.6, 123.4 ± 8.9, 101.3 ± 6.2, 88.9 ± 5.3, 74.8 ± 4.2, and 59.2 ± 3.9 ms for the 6 TRs of 500, 400, 300, 250, 200, and 150 ms, respectively. In vivo pore water nulling TIs were 132.8 ± 12.8, 110.3 ± 10.0, 80.0 ± 7.2, and 63.9 ± 5.4 ms for the 4 TRs of 400, 300, 200, and 150 ms, respectively. Excellent exponential fitting was achieved for IR-UTE imaging of bound water with pore water nulled at each TR. The mean T 1 BW was 106.9 ± 6.3 ms in vitro and 112.3 ± 16.4 ms in vivo.
CONCLUSION
Using the 3D IR-UTE Cones with a variable TR/TI approach, T 1 BW of cortical bone was calculated after complete nulling of pore water signals.
Collapse