Noroozifar M, Khorasani-Motlagh M, Akbari R. Application of pneumatic flow injection-tandem spectrometer system for chromium speciation.
JOURNAL OF AUTOMATED METHODS & MANAGEMENT IN CHEMISTRY 2007;
2007:34373. [PMID:
18273388 PMCID:
PMC2216071 DOI:
10.1155/2007/34373]
[Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Received: 03/12/2007] [Accepted: 06/29/2007] [Indexed: 05/25/2023]
Abstract
In this approach, a pneumatic flow injection-tandem spectrometer system, without a delivery pump, has been developed to study chromium speciation. In this system, suction force of pneumatic nebulizer of a flame atomic absorption spectrometer has been used for solution delivery through the manifold. Cr(VI) and total Cr concentrations were determined using UV-Vis and FAAS spectrometers, respectively. The Cr(III) was determined by difference. The calibration curves were linear up to 10 mug mL-1 and 20 mug mL-1 for Cr(VI) and total Cr with detection limit of 0.12 mug mL-1 and 0.07 mug mL-1 for Cr(VI) and Cr(III), respectively. The midrange precision and accuracy are less than 1.98% and +/- 2.50% for two species, respectively, at a sampling rate of 100 h-1. This system was applied for the determination of the chromium species in spiked and natural waters as well as industrial waters.