Pauli L, Nam J, Pasquali M, Behr M. Transient stress-based and strain-based hemolysis estimation in a simplified blood pump.
INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN BIOMEDICAL ENGINEERING 2013;
29:1148-60. [PMID:
23922311 PMCID:
PMC4166438 DOI:
10.1002/cnm.2576]
[Citation(s) in RCA: 18] [Impact Index Per Article: 1.6] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Received: 01/05/2013] [Revised: 04/30/2013] [Accepted: 06/06/2013] [Indexed: 05/12/2023]
Abstract
We compare two approaches to numerical estimation of mechanical hemolysis in a simplified blood pump model. The stress-based model relies on the instantaneous shear stress in the blood flow, whereas the strain-based model uses an additional tensor equation to relate distortion of red blood cells to a shear stress measure. We use the newly proposed least-squares finite element method (LSFEM) to prevent negative concentration fields and show a stable and volume preserving LSFEM for the tensor equation. Application of both models to a simplified centrifugal blood pump at three different operating conditions shows that the stress-based model overestimates the rate of hemolysis. The strain-based model is found to deliver lower hemolysis rates because it incorporates a more detailed description of biophysical phenomena into the simulation process.
Collapse