Cai X, Epstein FH. Free-breathing cine DENSE MRI using phase cycling with matchmaking and stimulated-echo image-based navigators.
Magn Reson Med 2018;
80:1907-1921. [PMID:
29607538 PMCID:
PMC6107388 DOI:
10.1002/mrm.27199]
[Citation(s) in RCA: 6] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/14/2017] [Revised: 02/19/2018] [Accepted: 03/06/2018] [Indexed: 01/17/2023]
Abstract
PURPOSE
This study aimed to develop a self-navigated method for free-breathing spiral cine displacement encoding with stimulated echoes (DENSE), a myocardial strain imaging technique that uses phase-cycling for artifact suppression. The method needed to address 2 consequences of motion for DENSE: striping artifacts from incomplete suppression of the T1 -relaxation echo and blurring.
METHODS
The method identifies phase-cycled spiral interleaves at matched respiratory phases by minimizing the residual signal due to T1 relaxation after phase-cycling subtraction. Next, the method reconstructs image-based navigators from matched phase-cycled interleaves that are comprised of the stimulated echo (ste-iNAVs). Ste-iNAVs are used for motion estimation and compensation of k-space data. The method was demonstrated in phantoms and compared to diaphragm-based navigator (dNAV) and conventional iNAV (c-iNAV) methods for the reconstruction of free-breathing volunteer data sets (N = 10).
RESULTS
Phantom experiments demonstrated that the proposed method removes striping artifacts and blurring due to motion. Volunteer results showed that respiratory motion measured by ste-iNAVs was better correlated than c-iNAVs to dNAV data (R2 = 0.82 ± 0.03 vs. 0.70 ± 0.05, P < 0.05). Match-making reconstructions of free-breathing data sets achieved lower residual T1 -relaxation echo energy (1.04 ± 0.01 vs. 1.18 ± 0.04 for dNAV and 1.18 ± 0.03 for c-iNAV, P < 0.05), higher apparent SNR (11.93 ± 1.05 vs. 10.68 ± 1.06 for dNAV and 10.66 ± 0.99 for c-iNAV, P < 0.05), and better phase quality (0.147 ± 0.012 vs. 0.166 ± 0.017 for dNAV, P = 0.06, and 0.168 ± 0.015 for c-iNAV, P < 0.05) than dNAV and c-iNAV methods.
CONCLUSION
For free-breathing cine DENSE, the proposed method addresses both types of breathing-induced artifacts and provides better quality images than conventional dNAV and iNAV methods.
Collapse