Abstract
BACKGROUND
Aortic regurgitation is mainly evaluated by trans-thoracic echocardiography using multi-parametric qualitative and semi quantitative tools. All those parameters can fail to meet expectations, resulting in an imperfect diagnostic reliability and assessment of aortic regurgitation severity can be challenging.
OBJECTIVES
We sought to evaluate feasibility and intra- and inter-observer reproducibility of aortic regurgitant orifice area measured by planimetry with tridimensional trans-esophageal echocardiography on patients with at least grade 2/4 aortic regurgitation.
PATIENTS AND METHODS
Consecutive patients with at least grade 2/4 aortic regurgitation measured by trans-thoracic echocardiography and referred for trans-esophageal echocardiography for any reason were included. Planimetric reconstructions of regurgitant orifice area were studied and reproducibility indexes between senior and junior observers were calculated.
RESULTS
Twenty-three patients were included in this study. Intra- and inter-observer reproducibility were excellent with an ICC of 0.95 [0.88-0.98], P<0.0001 and 0.91 [0.79-0.96], P<0.0001, respectively. Mean length of the measurement was 6.6±0.9min [CI95% 6.23-7.01].
CONCLUSION
Planimetric measurement of the aortic regurgitant orifice using tridimensional trans-esophageal echocardiography seems to be feasible and has great intra- and inter-observer reproducibility. Reconstruction durations were compatible with a daily use. There is a need now to investigate the reliability of this measurement as compared with the reference technique.