Islam S, Gulati RK, Domic M, Pal A, Kant K, Kim A. Performance Evaluation of Magnetic Resonance Coupling Method for Intra-Body Network (IBNet).
IEEE Trans Biomed Eng 2021;
69:1901-1908. [PMID:
34818185 DOI:
10.1109/tbme.2021.3130408]
[Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022]
Abstract
Effective management of emerging medical devices can lead to new insights in healthcare. Thus, a human body communication (HBC) is becoming increasingly important. In this paper, we present magnetic resonance (MR) coupling as a promising method for intra-body network (IBNet). The study reveals that MR coupling can effectively send or receive signals in biological tissue, with a maximum path loss of PL 33 dB (i.e. at 13.56 MHz), which is lower than other methods (e.g., galvanic, capacitive, or RF) for the same distance. The angular orientation of the transmitter and receiver coils at short and long distances also show a minor variation of the path loss (0.19 PL 0.62 dB), but more dependency on the distance (0.0547 dB/cm). Additionally, different postures during the MR coupling essentially do not affect path loss (PL 0.21 dB). In the multi-nodal transmission scenario, the MR coupling demonstrates that two nodes can simultaneously receive signals with -16.77 dBm loss at 60 cm and 100 cm distances, respectively. Such multi-node MR transmission can be utilized for communication, sensing, and powering wearable and implantable devices.
Collapse