Monson KL, Wall JS, Hainfeld JF. Visibility and stability of a 12-tungsten atom complex in the scanning transmission electron microscope.
Ultramicroscopy 1987;
21:147-56. [PMID:
3564170 DOI:
10.1016/0304-3991(87)90081-7]
[Citation(s) in RCA: 9] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/06/2023]
Abstract
A complex consisting of 12 tungsten atoms has been studied in terms of signal-to-noise (S/N) and dose response in the scanning transmission electron microscope (STEM), to evaluate its suitability for use as a approximately 1 nm resolution biological label. Molecular weight of the complex was measured as a function of radius of integration, and results were in agreement with the calculated formula weight. S/N was highest at the lowest radius of integration (0.25 nm), and decreased monotonically with increasing radius. The complex was clearly visible at a dose of 4 X 10(3) e/nm2, and exhibited negligible mass loss (approximately 8%) after an accumulated dose of 1.28 X 10(5) e/nm2. Beam-induced motion was small, 0.46 nm rms after 4 X 10(4) e/nm2. Some intensity fluctuations were observed between successive scans of the same clusters, for which a diffraction-based explanation is advanced. Upon suitable functionalization, the tungsten complex is expected to complement the undecagold cluster already in use for site-specific labeling.
Collapse