Orland HJ, Jones D. Cardiac pacemaker induced ventricular fibrillation during surgical diathermy.
Anaesth Intensive Care 1975;
3:321-6. [PMID:
1211612 DOI:
10.1177/0310057x7500300407]
[Citation(s) in RCA: 11] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/26/2022]
Abstract
Two demand pacemakers were observed to repeatedly fibrillate the heart during clinical surgery employing diathermy. Marked hypotension associated with a rapid, irregular venticular action was observed in another patient who had a fixed rate pacemaker implanted. Pacemakers were tested with function generators over a range of frequencies and some increased their rate to between 200-600 impulse/min. Animal experiments confirmed that, contrary to manufacturers advice, surgical diathermy produces myocardial fibrillation in the presence of some pacemakers. Mechanisms of these effects and their clinical significance are discussed.