1
|
Bancheri J, Seuntjens J. A semi-analytical procedure to determine the ion recombination correction factor in high dose-per-pulse beams. Med Phys 2024; 51:4458-4471. [PMID: 38446555 DOI: 10.1002/mp.17005] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/05/2023] [Revised: 02/08/2024] [Accepted: 02/13/2024] [Indexed: 03/08/2024] Open
Abstract
BACKGROUND The conventional theories and methods of determining the ion recombination correction factor, such as Boag theory and the related two voltage method and Jaffé plot extrapolation, do not seem to yield accurate results in FLASH /high dose per pulse (DPP) beams ( > $>$ 10 mGy DPP). This is due to the presence of a large free electron fraction that distorts the electric field inside the chamber sensitive volume. To understand the influence of these effects on the ion recombination correction factor and to develop new expressions for it, it is necessary to re-visit the underlying physics. PURPOSE To present a mathematical procedure to develop an analytical expression for the ion recombination correction factor. The expression is the basis for an extrapolation method so the correction factor can be determined in a clinical setting. METHODS A semi-analytical solution method, the homotopy perturbation method (HPM), is used to solve the partial differential equations (PDEs) describing the charge carrier physics, including space charge and free electrons. The electron velocity and attachment rate are modeled as functions of the electric field strength. An expression for the charge collection efficiency and ion recombination correction factor are developed. A fit procedure based on this expression is used to compare it to measured data from previously published articles. Another fit procedure using a general equation is also proposed and compared to the data. RESULTS The series obtained for the charge collection efficiency and the ion recombination correction factor are determined to be asymptotic series and the optimal truncation established. The ion recombination correction factor exhibits a1 / V 2 $1/V^2$ dependency due to the free electron presence. The fit using this expression agrees well with measured data as long as (1) the DPP is below 1 Gy for chambers with a 1 mm plate separation and (2) when the DPP is below 3 Gy for chambers with a 0.5 mm plate separation. In these DPP ranges, the deviation between measured and fit value did not exceed 6%. In both chamber cases the voltage range where the fit applies decreases as DPP increases. The general equation yielded comparable results. CONCLUSIONS The HPM was shown to be applicable to a complex system of PDEs and generate meaningful and novel solutions, as they include both space charge and free electrons. The HPM also lends itself to other chamber geometries. The fit procedure was also shown to yield accurate results for the ion recombination correction up to the 1 Gy DPP level.
Collapse
Affiliation(s)
- Julien Bancheri
- Department of Physics & Medical Physics Unit, McGill University, Montreal, Quebec, Canada
| | - Jan Seuntjens
- Princess Margaret Cancer Centre, Radiation Medicine Program, University Health Network, Department of Medical Biophysics, University of Toronto, Toronto, Canada
| |
Collapse
|
2
|
Mastella E, Szilagyi KE, De Guglielmo E, Fabbri S, Calderoni F, Stefanelli A, Di Domenico G, Turra A. Dosimetric characterization of a mobile accelerator dedicated for intraoperative radiation therapy: Monte Carlo simulations and experimental validation. Phys Med 2022; 104:167-173. [PMID: 36463581 DOI: 10.1016/j.ejmp.2022.11.006] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 08/31/2022] [Revised: 10/12/2022] [Accepted: 11/25/2022] [Indexed: 12/03/2022] Open
Abstract
PURPOSE This Technical Note validates previously published data about the dosimetry of the electron beams produced by a mobile accelerator dedicated for intraoperative radiation therapy (IORT). The evaluation of the directional response of a PTW microDiamond detector is presented together with a detailed analysis of the output factors (OFs) for bevelled applicators. METHODS The OFs of the 6, 8, 10 and 12 MeV electron beams produced by a light intraoperative accelerator (LIAC, SIT, Italy) were measured in a commercial water phantom using the microDiamond. A set of flat and bevelled applicators with sizes ranging from 4 to 10 cm was characterized. For bevelled applicators, a correction for the angular dependence of the microDiamond was calculated using a home-made spherical phantom. Correction factors were obtained through measurements performed rotating the accelerator treatment head at 0°, 15°, 30° and 45°. RESULTS For flat applicators, the average deviation between measured and simulated OFs was (-1.1 ± 0.7)%. The microDiamond showed a higher angular dependence for the 6 MeV beam (∼8% for angles up to 45°, range 92 % ÷ 100 %), while the variations for 8, 10 and 12 MeV beams were ∼ 4 % (range 97 % ÷ 101 %). Correcting for this dependence, the average deviation of the OFs for bevelled applicators was (-0.9 ± 1.6)%. CONCLUSIONS The presented results were in very good agreement with those reported in literature. Very similar deviations were found between flat and bevelled applicators confirming the suitability of our method to determine the angular dependence correction factors of the microDiamond detector.
Collapse
Affiliation(s)
- Edoardo Mastella
- Struttura Complessa di Fisica Medica, Azienda Ospedaliero-Universitaria di Ferrara, via A. Moro 8, I-44124 Cona (Ferrara), Italy.
| | - Klarisa E Szilagyi
- Dipartimento di Fisica e Scienze della Terra, Università degli Studi di Ferrara, via Saragat 1, I-44122 Ferrara, Italy
| | - Eleonora De Guglielmo
- Struttura Complessa di Fisica Medica, Azienda Ospedaliero-Universitaria di Ferrara, via A. Moro 8, I-44124 Cona (Ferrara), Italy
| | - Sara Fabbri
- Struttura Complessa di Fisica Medica, Azienda Ospedaliero-Universitaria di Ferrara, via A. Moro 8, I-44124 Cona (Ferrara), Italy
| | - Francesca Calderoni
- Struttura Complessa di Fisica Medica, Azienda Ospedaliero-Universitaria di Ferrara, via A. Moro 8, I-44124 Cona (Ferrara), Italy
| | - Antonio Stefanelli
- Struttura Complessa di Radioterapia Oncologica, Azienda Ospedaliero-Universitaria di Ferrara, via A. Moro 8, I-44124 Cona (Ferrara), Italy
| | - Giovanni Di Domenico
- Dipartimento di Fisica e Scienze della Terra, Università degli Studi di Ferrara, via Saragat 1, I-44122 Ferrara, Italy
| | - Alessandro Turra
- Struttura Complessa di Fisica Medica, Azienda Ospedaliero-Universitaria di Ferrara, via A. Moro 8, I-44124 Cona (Ferrara), Italy
| |
Collapse
|
3
|
Kranzer R, Poppinga D, Weidner J, Schüller A, Hackel T, Looe HK, Poppe B. Ion collection efficiency of ionization chambers in ultra-high dose-per-pulse electron beams. Med Phys 2021; 48:819-830. [PMID: 33251606 DOI: 10.1002/mp.14620] [Citation(s) in RCA: 30] [Impact Index Per Article: 10.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/19/2020] [Revised: 09/30/2020] [Accepted: 11/18/2020] [Indexed: 02/03/2023] Open
Abstract
PURPOSE The ion collection efficiency of vented ionization chambers has been investigated in an ultra-high dose-per-pulse (DPP) electron beam. The role of the chamber design and the electric field strength in the sensitive air volume have been evaluated. METHODS An advanced Markus chamber and three specially designed parallel plate air-filled ionization chambers (EWC: End Window Chamber) with varying electrode distance of 0.5, 1, and 2 mm have been investigated. Their ion collection efficiencies were determined experimentally using two methods: extrapolation of Jaffé plots and comparison against a DPP-independent reference detector. The latter was achieved by calibrating a current transformer against alanine dosimeters. All measurements were performed in a 24 MeV electron beam with DPP values between 0.01 and 3 Gy. Additionally, the numerical approach introduced by Gotz et al. was implemented taking into account space charge effects at these ultra-high DPPs. The method has been extended to obtain time-resolved and position-dependent electric field distortions within the air cavity. RESULTS The ion collection efficiency of the investigated ionization chambers drops significantly in the ultra-high DPP range. The extent of this drop is dependent on the electrode distance, the applied chamber voltage and thus the field strength in the sensitive air volume. For the Advanced Markus chamber, a good agreement between the experimental, numerical and the results of Petersson et al. could be shown. Using the three EWCs with different electrode spacing, an improvement of the ion collection efficiency and a reduction of the polarity effect with decreasing electrode distance could be demonstrated. Furthermore, the results revealed that the determination of the ion collection efficiency from the Jaffé plots and therefore also from two-voltage method typically underestimate the ion collection efficiency in the region of high dose-per-pulse (3 to 130 mGy) and overestimate the ion collection efficiency at ultra-high dose-per-pulse (>1 Gy per pulse). CONCLUSIONS In this work, the ion collection efficiency determined with different methods and ionization chambers have been compared and discussed. As expected, an increase of the electric field in the ionization chamber, either by applying a higher bias voltage or a reduction of the electrode distance, improves the ion collection efficiency and also reduces the polarity effect. For the Advanced Markus chamber, the experimental results obtained by comparison against a reference agree well with the numerical solution. Based on these results, it seems possible to keep the recombination loss less than or equal to 5% up to a dose-per-pulse of 3 Gy with an appropriately designed ionization chamber, which corresponds to the level accepted in conventional radiotherapy dosimetry protocols.
Collapse
Affiliation(s)
- Rafael Kranzer
- PTW-Freiburg, Freiburg, 79115, Germany.,University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, 26121, Germany
| | | | | | - Andreas Schüller
- Physikalisch-Technische Bundesanstalt, Braunschweig, 38116, Germany
| | - Thomas Hackel
- Physikalisch-Technische Bundesanstalt, Braunschweig, 38116, Germany
| | - Hui Khee Looe
- University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, 26121, Germany
| | - Björn Poppe
- University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, 26121, Germany
| |
Collapse
|
4
|
Esplen N, Mendonca MS, Bazalova-Carter M. Physics and biology of ultrahigh dose-rate (FLASH) radiotherapy: a topical review. Phys Med Biol 2020; 65:23TR03. [PMID: 32721941 DOI: 10.1088/1361-6560/abaa28] [Citation(s) in RCA: 114] [Impact Index Per Article: 28.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/19/2022]
Abstract
Ultrahigh dose-rate radiotherapy (RT), or 'FLASH' therapy, has gained significant momentum following various in vivo studies published since 2014 which have demonstrated a reduction in normal tissue toxicity and similar tumor control for FLASH-RT when compared with conventional dose-rate RT. Subsequent studies have sought to investigate the potential for FLASH normal tissue protection and the literature has been since been inundated with publications on FLASH therapies. Today, FLASH-RT is considered by some as having the potential to 'revolutionize radiotherapy'. FLASH-RT is considered by some as having the potential to 'revolutionize radiotherapy'. The goal of this review article is to present the current state of this intriguing RT technique and to review existing publications on FLASH-RT in terms of its physical and biological aspects. In the physics section, the current landscape of ultrahigh dose-rate radiation delivery and dosimetry is presented. Specifically, electron, photon and proton radiation sources capable of delivering ultrahigh dose-rates along with their beam delivery parameters are thoroughly discussed. Additionally, the benefits and drawbacks of radiation detectors suitable for dosimetry in FLASH-RT are presented. The biology section comprises a summary of pioneering in vitro ultrahigh dose-rate studies performed in the 1960s and early 1970s and continues with a summary of the recent literature investigating normal and tumor tissue responses in electron, photon and proton beams. The section is concluded with possible mechanistic explanations of the FLASH normal-tissue protection effect (FLASH effect). Finally, challenges associated with clinical translation of FLASH-RT and its future prospects are critically discussed; specifically, proposed treatment machines and publications on treatment planning for FLASH-RT are reviewed.
Collapse
Affiliation(s)
- Nolan Esplen
- Department of Physics and Astronomy, University of Victoria, Victoria, BC, Canada
| | | | | |
Collapse
|
5
|
D'Arienzo M, Andreoli S, Pimpinella M. Evaluation of the uncertainty associated with the ion recombination correction in high dose-per-pulse electron beam dosimetry: an MC approach. Phys Med Biol 2020; 65:09NT01. [PMID: 32097891 DOI: 10.1088/1361-6560/ab79c2] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
Abstract
The high dose and dose-per-pulse rates (up to 130 mGy/pulse) produced by some intraoperative radiation therapy (IORT) accelerators pose specific dosimetric problems due to the high density of electric charge per pulse produced in the ionization chamber cavity. In particular, the correction factor for ion recombination, ks , calculated with the traditional two-voltage method is significantly overestimated and three alternative models have been proposed in the literature allowing for the presence of a free-electron component. However, at present there is no general consensus on the best model to assess the ion recombination correction and controversy remains on the uncertainty associated with k s . In the present work we adopted a Monte Carlo (MC) approach to assess the uncertainty associated with the ion recombination correction in plane-parallel chambers used in high dose-per-pulse electron beam dosimetry. The uncertainty associated with k s was calculated for the following plane-parallel ionization chambers: Scanditronix/Wellhofer Parallel Plate Chamber PPC05 and PPC40, PTW Advanced Markus Model 34 045 and PTW Roos Model 34 001. Input variables for MC calculations were derived from experimental data at 28 and 73 mGy/pulse. Taken together, the results of this study indicate that ks values calculated according to the three ion recombination models do not overlap within their standard uncertainties, suggesting that an additional type-B uncertainty component would be necessary to take into account possible differences between the models. Our results indicate that the combined relative standard uncertainty in k s should be calculated as the sum in quadrature of a (type-A) MC-based uncertainty component and a (type-B) uncertainty contribution evaluated assuming a uniform distribution between k s values obtained from the two extreme models.
Collapse
Affiliation(s)
- Marco D'Arienzo
- ENEA, National Institute of Ionizing Radiation Metrology, Rome, Italy
| | | | | |
Collapse
|
6
|
Characterization of Fricke-gelatin dosimeters for intraoperative Radiation Therapy dosimetry. Radiat Phys Chem Oxf Engl 1993 2020. [DOI: 10.1016/j.radphyschem.2019.108629] [Citation(s) in RCA: 5] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/24/2022]
|
7
|
Avanzo M, Dassie A, Chandra Acharya P, Chiovati P, Pirrone G, Avigo C, Barresi L, Dang Quoc S, Fiagbedzi E, Navarria F, Palazzari E, Bertola G, De Paoli A, Stancanello J, Sartor G. Electron radiotherapy (IOERT) for applications outside of the breast: Dosimetry and influence of tissue inhomogeneities. Phys Med 2020; 69:82-89. [PMID: 31841774 DOI: 10.1016/j.ejmp.2019.12.003] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 08/09/2019] [Revised: 11/17/2019] [Accepted: 12/04/2019] [Indexed: 12/12/2022] Open
Abstract
PURPOSE The purpose of study is to investigate the dosimetry of electron intraoperative radiotherapy (IOERT) of the Intraop Mobetron 2000 mobile LINAC in treatments outside of the breast. After commissioning and external validation of dosimetry, we report in vivo results of measurements for treatments outside the breast in a large patient cohort, and investigate if the presence of inhomogeneities can affect in vivo measurements. METHODS AND MATERIALS Applicator factors and profile curves were measured with a stereotactic diode. The applicators factors of the 6 cm flat and beveled applicators were also confirmed with radiochromic films, parallel-plate ion chamber and by an external audit performed with ThermoLuminescent Dosimeters (TLDs). The influence of bone on dose was investigated by using radiochromic films attached to an insert equivalent to cortical bone, immersed in the water phantom. In vivo dosimetry was performed on 126 patients treated with IOERT using metal oxide-silicon semiconductor field effect transistors (MOSFETs) placed on the tumor bed. RESULTS Relatively small differences were found among different detectors for measurements of applicator factors. In the external audit, the agreement with the TLD was mostly within ±0.2%. The largest increase of dose due to the presence of cortical bone insert was +6.0% with energy 12 MeV and 3 cm applicator. On average, in vivo dose was significantly (+3.1%) larger than prescribed dose. CONCLUSION IOERT in applications outside the breast results in low discrepancies between in vivo and prescribed doses, which can be also explained with the presence of tissue inhomogeneity.
Collapse
Affiliation(s)
- Michele Avanzo
- Department of Medical Physics, Centro di Riferimento Oncologico di Aviano (CRO) IRCCS, 33081 Aviano, Italy.
| | - Andrea Dassie
- Department of Medical Physics, Centro di Riferimento Oncologico di Aviano (CRO) IRCCS, 33081 Aviano, Italy
| | | | - Paola Chiovati
- Department of Medical Physics, Centro di Riferimento Oncologico di Aviano (CRO) IRCCS, 33081 Aviano, Italy
| | - Giovanni Pirrone
- Department of Medical Physics, Centro di Riferimento Oncologico di Aviano (CRO) IRCCS, 33081 Aviano, Italy
| | - Cinzia Avigo
- Medical Physics, ULSS 1 Dolomiti- S. Martino Hospital, Medical Physics Department, Belluno, Italy
| | - Loredana Barresi
- Department of Medical Physics, Centro di Riferimento Oncologico di Aviano (CRO) IRCCS, 33081 Aviano, Italy
| | - Soai Dang Quoc
- Medicinal Supplies, Medical Physics Division, Hanoi Oncology Hospital, Hanoi, Vietnam
| | | | - Federico Navarria
- Radiation Oncology, Centro di Riferimento Oncologico di Aviano (CRO) IRCCS, 33081 Aviano, Italy
| | - Elisa Palazzari
- Radiation Oncology, Centro di Riferimento Oncologico di Aviano (CRO) IRCCS, 33081 Aviano, Italy
| | - Giulio Bertola
- Surgical Oncology, Centro di Riferimento Oncologico di Aviano (CRO) IRCCS, 33081 Aviano, Italy
| | - Antonino De Paoli
- Radiation Oncology, Centro di Riferimento Oncologico di Aviano (CRO) IRCCS, 33081 Aviano, Italy
| | | | - Giovanna Sartor
- Department of Medical Physics, Centro di Riferimento Oncologico di Aviano (CRO) IRCCS, 33081 Aviano, Italy
| |
Collapse
|
8
|
Martin-Martin G, Aguilar PB, Barbés B, Guibelalde E. Assessment of ion recombination correction and polarity effects for specific ionization chambers in flattening-filter-free photon beams. Phys Med 2019; 67:176-184. [PMID: 31734555 DOI: 10.1016/j.ejmp.2019.07.018] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 04/11/2019] [Revised: 07/23/2019] [Accepted: 07/24/2019] [Indexed: 11/17/2022] Open
Abstract
PURPOSE To investigate ion recombination correction and polarity effects in four ion chamber models in flattening-filter-free (FFF) beams to (1) evaluate their suitability for reference dosimetry; (2) assess the accuracy of the two-voltage technique (TVA) against the Bruggmoser formalism; and (3) examine the influence of the accelerator type on the recombination correction. METHODS Jaffé plots were created for a variety of microchambers, small-volume and Farmer-type chambers to obtain kS, the recombination correction factor, using two different types of accelerators. These values were plotted against dose-per-pulse and Jaffé plots for opposite polarities were created to determine which chambers meet the AAPM TG-51 addendum recombination and polarity specifications. RESULTS Nearly all small-volume chambers exhibited reference-class behavior with respect to ion recombination and polarity effects. The microchambers exhibited anomalous recombination and polarity effects, precluding their use for reference dosimetry in FFF beams. For the reference-class chambers, agreement between TVA-determined kS values and Jaffé and Bruggmoser formalisms-determined kS values was within 0.1%. No significant differences were found between the kS values obtained with the two different accelerators used in this work. CONCLUSIONS This study stresses the need to characterize ion recombination correction and polarity effects for small-volume chambers and microchambers on an individual chamber basis and with the more rigorous criteria of the AAPM TG-51 addendum. Furthermore, the study demonstrated the suitability of the TVA method for chambers that exhibit reference-class behavior in FFF beams. Finally, this work has shown that the recombination correction does not depend on the type of accelerator but on its dose-per-pulse.
Collapse
Affiliation(s)
- Guadalupe Martin-Martin
- Medical Physics and Radiation Protection Service, Hospital Universitario de Fuenlabrada, C/ Camino del Molino 2, 28492 Fuenlabrada, Madrid, Spain.
| | - Pedro Borja Aguilar
- Clínica Universidad de Navarra, Department of Radiation Physics, Avenida Pío XII, 31080 Pamplona, Navarra, Spain
| | - Benigno Barbés
- Clínica Universidad de Navarra, Department of Radiation Physics, Avenida Pío XII, 31080 Pamplona, Navarra, Spain
| | - Eduardo Guibelalde
- Medical Physics Group, Department of Radiology, University Complutense of Madrid, 28040 Madrid, Spain
| |
Collapse
|
9
|
Pimpinella M, Andreoli S, De Angelis C, Della Monaca S, D'Arienzo M, Menegotti L. Output factor measurement in high dose-per-pulse IORT electron beams. Phys Med 2019; 61:94-102. [DOI: 10.1016/j.ejmp.2019.04.021] [Citation(s) in RCA: 13] [Impact Index Per Article: 2.6] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 12/14/2018] [Revised: 04/18/2019] [Accepted: 04/24/2019] [Indexed: 12/14/2022] Open
|
10
|
Güngör G, Aydın G, Mustafayev TZ, Özyar E. Output factors of ionization chambers and solid state detectors for mobile intraoperative radiotherapy (IORT) accelerator electron beams. J Appl Clin Med Phys 2019; 20:13-23. [PMID: 30632271 PMCID: PMC6370992 DOI: 10.1002/acm2.12522] [Citation(s) in RCA: 7] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/27/2018] [Revised: 12/06/2018] [Accepted: 12/09/2018] [Indexed: 11/11/2022] Open
Abstract
Purpose The electron energy characteristics of mobile intraoperative radiotherapy (IORT) accelerator LIAC® differ from commonly used linear accelerators, thus some of the frequently used detectors can give less accurate results. The aim of this study is to evaluate the output factors (OFs) of several ionization chambers (IC) and solid state detectors (SS) for electron beam energies generated by LIAC® and compare with the output factor of Monte Carlo model (MC) in order to determine the adequate detectors for LIAC®. Methods The OFs were measured for 6, 8, 10, and 12 MeV electron energies with PTW 23343 Markus, PTW 34045 Advanced Markus, PTW 34001 Roos, IBA PPC05, IBA PPC40, IBA NACP‐02, PTW 31010 Semiflex, PTW 31021 Semiflex 3D, PTW 31014 Pinpoint, PTW 60017 Diode E, PTW 60018 Diode SRS, SNC Diode EDGE, and PTW 60019 micro Diamond detectors. Ion recombination factors (ksat) of IC were measured for all applicator sizes and OFs were corrected according to ksat. The measured OFs were compared with Monte Carlo output factors (OFMC). Results The measured OFs of IBA PPC05, PTW Advanced Markus, PTW Pinpoint, PTW microDiamond, and PTW Diode E detectors are in good agreement with OFMC. The maximum deviations of IBA PPC05 OFs to OFMC are −1.6%, +1.5%, +1.5%, and +2.0%; for PTW Advanced Markus +1.0%, +1.5%, +2.0%, and +2.0%; for PTW Pinpoint +2.0%, +1.6%, +4.0%, and +2.0%; for PTW microDiamond −1.6%, +2%, +1.1%, and +1.0%; and for PTW Diode E −+1.7%, +1.7%, +1.3%, and +2.5% for 6, 8, 10, and 12 MeV, respectively. PTW Roos, PTW Markus, IBA PPC40, PTW Semiflex, PTW Semiflex 3D, SNC Diode Edge measured OFs with a maximum deviation of +5.6%, +4.5%, +5.6%, +8.1%, +4.8%, and +9.6% with respect to OFMC, while PTW Diode SRS and IBA NACP‐02 were the least accurate (with highest deviations −37.1% and −18.0%, respectively). Conclusion The OFs results of solid state detectors PTW microDiamond and PTW Diode E as well as the ICs with small electrode spacing distance such as IBA PPC05, PTW Advanced Markus and PTW Pinpoint are in excellent agreement with OFMC. The measurements of the other detectors evaluated in this study are less accurate, thus they should be used with caution. Particularly, PTW Diode SRS and IBA NACP‐02 are not suitable and their use should be avoided in relative dosimetry measurements under high dose per pulsed (DPP) electron beams.
Collapse
Affiliation(s)
- Görkem Güngör
- Department of Medical Physics, Medipol University Institute of Health Sciences, Istanbul, Turkey.,Department of Radiation Oncology, Acıbadem Mehmet Ali Aydınlar University School of Medicine, Istanbul, Turkey
| | - Gökhan Aydın
- Department of Radiation Oncology, Acıbadem Mehmet Ali Aydınlar University School of Medicine, Istanbul, Turkey
| | - Teuta Zoto Mustafayev
- Department of Radiation Oncology, Acıbadem Mehmet Ali Aydınlar University School of Medicine, Istanbul, Turkey
| | - Enis Özyar
- Department of Radiation Oncology, Acıbadem Mehmet Ali Aydınlar University School of Medicine, Istanbul, Turkey
| |
Collapse
|
11
|
Gotz M, Karsch L, Pawelke J. A new model for volume recombination in plane-parallel chambers in pulsed fields of high dose-per-pulse. ACTA ACUST UNITED AC 2017; 62:8634-8654. [DOI: 10.1088/1361-6560/aa8985] [Citation(s) in RCA: 21] [Impact Index Per Article: 3.0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
|
12
|
Petersson K, Jaccard M, Germond JF, Buchillier T, Bochud F, Bourhis J, Vozenin MC, Bailat C. High dose-per-pulse electron beam dosimetry - A model to correct for the ion recombination in the Advanced Markus ionization chamber. Med Phys 2017; 44:1157-1167. [PMID: 28094853 DOI: 10.1002/mp.12111] [Citation(s) in RCA: 117] [Impact Index Per Article: 16.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/23/2016] [Revised: 01/11/2017] [Accepted: 01/11/2017] [Indexed: 11/12/2022] Open
Abstract
PURPOSE The purpose of this work was to establish an empirical model of the ion recombination in the Advanced Markus ionization chamber for measurements in high dose rate/dose-per-pulse electron beams. In addition, we compared the observed ion recombination to calculations using the standard Boag two-voltage-analysis method, the more general theoretical Boag models, and the semiempirical general equation presented by Burns and McEwen. METHODS Two independent methods were used to investigate the ion recombination: (a) Varying the grid tension of the linear accelerator (linac) gun (controls the linac output) and measuring the relative effect the grid tension has on the chamber response at different source-to-surface distances (SSD). (b) Performing simultaneous dose measurements and comparing the dose-response, in beams with varying dose rate/dose-per-pulse, with the chamber together with dose rate/dose-per-pulse independent Gafchromic™ EBT3 film. Three individual Advanced Markus chambers were used for the measurements with both methods. All measurements were performed in electron beams with varying mean dose rate, dose rate within pulse, and dose-per-pulse (10-2 ≤ mean dose rate ≤ 103 Gy/s, 102 ≤ mean dose rate within pulse ≤ 107 Gy/s, 10-4 ≤ dose-per-pulse ≤ 101 Gy), which was achieved by independently varying the linac gun grid tension, and the SSD. RESULTS The results demonstrate how the ion collection efficiency of the chamber decreased as the dose-per-pulse increased, and that the ion recombination was dependent on the dose-per-pulse rather than the dose rate, a behavior predicted by Boag theory. The general theoretical Boag models agreed well with the data over the entire investigated dose-per-pulse range, but only for a low polarizing chamber voltage (50 V). However, the two-voltage-analysis method and the Burns & McEwen equation only agreed with the data at low dose-per-pulse values (≤ 10-2 and ≤ 10-1 Gy, respectively). An empirical model of the ion recombination in the chamber was found by fitting a logistic function to the data. CONCLUSIONS The ion collection efficiency of the Advanced Markus ionization chamber decreases for measurements in electron beams with increasingly higher dose-per-pulse. However, this chamber is still functional for dose measurements in beams with dose-per-pulse values up toward and above 10 Gy, if the ion recombination is taken into account. Our results show that existing models give a less-than-accurate description of the observed ion recombination. This motivates the use of the presented empirical model for measurements with the Advanced Markus chamber in high dose-per-pulse electron beams, as it enables accurate absorbed dose measurements (uncertainty estimation: 2.8-4.0%, k = 1). The model depends on the dose-per-pulse in the beam, and it is also influenced by the polarizing chamber voltage, with increasing ion recombination with a lowering of the voltage.
Collapse
Affiliation(s)
- Kristoffer Petersson
- CHUV, Institut de Radiophysique, Rue du Grand-Pré 1, CH-1007, Lausanne, Switzerland
| | - Maud Jaccard
- CHUV, Institut de Radiophysique, Rue du Grand-Pré 1, CH-1007, Lausanne, Switzerland
| | | | - Thierry Buchillier
- CHUV, Institut de Radiophysique, Rue du Grand-Pré 1, CH-1007, Lausanne, Switzerland
| | - François Bochud
- CHUV, Institut de Radiophysique, Rue du Grand-Pré 1, CH-1007, Lausanne, Switzerland
| | - Jean Bourhis
- CHUV, Service de Radio-Oncologie, Rue du Bugnon 46, CH - 1011, Lausanne, Switzerland
| | | | - Claude Bailat
- CHUV, Institut de Radiophysique, Rue du Grand-Pré 1, CH-1007, Lausanne, Switzerland
| |
Collapse
|
13
|
Hyun MA, Miller JR, Micka JA, DeWerd LA. Ion recombination and polarity corrections for small-volume ionization chambers in high-dose-rate, flattening-filter-free pulsed photon beams. Med Phys 2017; 44:618-627. [PMID: 28001291 DOI: 10.1002/mp.12053] [Citation(s) in RCA: 13] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/02/2016] [Revised: 10/25/2016] [Accepted: 11/16/2016] [Indexed: 11/06/2022] Open
Abstract
PURPOSE To investigate ion recombination and polarity effects in scanning and microionization chambers when used with digital electrometers and high-dose-rate linac beams such as flattening-filter-free (FFF) fields, and to compare results against conventional pulsed and continuous photon beams. METHODS Saturation curves were obtained for one Farmer-type ionization chamber and eight small-volume chamber models with volumes ranging from 0.01 to 0.13 cm3 using a Varian TrueBeam™ STx with FFF capability. Three beam modes (6 MV, 6 MV FFF, and 10 MV FFF) were investigated, with nominal dose-per-pulse values of 0.0278, 0.0648, and 0.111 cGy/pulse, respectively, at dmax . Saturation curves obtained using the Theratronics T1000 60 Co unit at the UWADCL and a conventional linear accelerator (Varian Clinac iX) were used to establish baseline behavior. Jaffé plots were fitted to obtain Pion , accounting for exponential effects such as charge multiplication. These values were compared with the two-voltage technique recommended in TG-51, and were plotted as a function of dose-per-pulse to assess the ability of small-volume chambers to meet reference-class criteria in FFF beams. RESULTS Jaffé- and two-voltage-determined Pion values measured for high-dose-rate beams agreed within 0.1% for the Farmer-type chamber and 1% for scanning and microionization chambers, with the exception of the CC01 which agreed within 2%. With respect to ion recombination and polarity effects, the Farmer-type chamber, scanning chambers and the Exradin A26 microchamber exhibited reference-class behavior in all beams investigated, with the exception of the IBA CC04 scanning chamber, which had an initial recombination correction that varied by 0.2% with polarity. All microchambers investigated, with the exception of the A26, exhibited anomalous polarity and ion recombination behaviors that make them unsuitable for reference dosimetry in conventional and high-dose-rate photon beams. CONCLUSIONS The results of this work demonstrate that recombination and polarity behaviors seen in conventional pulsed and continuous photon beams trend accordingly in high-dose-rate FFF linac beams. Several models of small-volume ionization chambers used with a digital electrometer have been shown to meet reference-class requirements with respect to ion recombination and polarity, even in the high-dose-rate environment. For such chambers, a two-voltage technique agreed well with more rigorous methods of determining Pion . However, the results emphasize the need for careful reference detector selection, and indicate that ionization chambers ought to be extensively tested in each beam of interest prior to their use for reference dosimetry.
Collapse
Affiliation(s)
- Megan A Hyun
- Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, WI, 53705, USA
| | - Jessica R Miller
- Department of Human Oncology, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, WI, 53792, USA
| | - John A Micka
- Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, WI, 53705, USA
| | - Larry A DeWerd
- Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, WI, 53705, USA
| |
Collapse
|
14
|
Abstract
Literature was reviewed to assess the physical aspects governing the present and emerging technologies used in intraoperative radiation therapy (IORT). Three major technologies were identified: treatment with electrons, treatment with external generators of kV X-rays and electronic brachytherapy. Although also used in IORT, literature on brachytherapy with radioactive sources is not systematically reviewed since an extensive own body of specialized literature and reviews exists in this field. A comparison with radioactive sources is made in the use of balloon catheters for partial breast irradiation where these are applied in almost an identical applicator technique as used with kV X-ray sources. The physical constraints of adaption of the dose distribution to the extended target in breast IORT are compared. Concerning further physical issues, the literature on radiation protection, commissioning, calibration, quality assurance (QA) and in-vivo dosimetry of the three technologies was reviewed. Several issues were found in the calibration and the use of dosimetry detectors and phantoms for low energy X-rays which require further investigation. The uncertainties in the different steps of dose determination were estimated, leading to an estimated total uncertainty of around 10-15% for IORT procedures. The dose inhomogeneity caused by the prescription of electrons at 90% and by the steep dose gradient of kV X-rays causes additional deviations from prescription dose which must be considered in the assessment of dose response in IORT.
Collapse
Affiliation(s)
- Frank W Hensley
- Department of Radiation Oncology, University Hospital of Heidelberg, Im Neuenheimer Feld 400, 69120, Heidelberg, Germany.
- , Present address: Birkenweg 35, 69221, Dossenheim, Germany.
| |
Collapse
|
15
|
Scalchi P, Ciccotelli A, Felici G, Petrucci A, Massafra R, Piazzi V, D'Avenia P, Cavagnetto F, Cattani F, Romagnoli R, Soriani A. Use of parallel-plate ionization chambers in reference dosimetry of NOVAC and LIAC®mobile electron linear accelerators for intraoperative radiotherapy: a multi-center survey. Med Phys 2017; 44:321-332. [DOI: 10.1002/mp.12020] [Citation(s) in RCA: 15] [Impact Index Per Article: 2.1] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/15/2016] [Revised: 11/06/2016] [Accepted: 11/11/2016] [Indexed: 11/12/2022] Open
Affiliation(s)
- Paolo Scalchi
- Department of Medical Physics; Azienda U.L.S.S. 8; Vicenza 36100 Italy
| | | | | | - Assunta Petrucci
- Department of Medical Physics; S. Filippo Neri Hospital; Rome 00135 Italy
| | - Raffaella Massafra
- Department of Medical Physics; I.R.C.C.S. Institute of Oncology Giovanni Paolo II; Bari 70124 Italy
| | - Valeria Piazzi
- Department of Radiotherapy; Multimedica Hospital; Castellanza 21053 Italy
| | - Paola D'Avenia
- Department of Medical Physics; ASUR MARCHE AV3; Macerata 62100 Italy
| | | | - Federica Cattani
- Department of Medical Physics; European Institute of Oncology; Milan 20141 Italy
| | | | - Antonella Soriani
- Laboratory of Medical Physics and Expert Systems; Regina Elena National Cancer Institute; Rome 00144 Italy
| |
Collapse
|
16
|
Determination of the ion recombination correction factor for intraoperative electron beams. Z Med Phys 2016; 26:35-44. [DOI: 10.1016/j.zemedi.2015.06.011] [Citation(s) in RCA: 10] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/11/2015] [Revised: 05/29/2015] [Accepted: 06/15/2015] [Indexed: 11/15/2022]
|
17
|
Characterization of a microDiamond detector in high-dose-per-pulse electron beams for intra operative radiation therapy. Phys Med 2015; 31:897-902. [DOI: 10.1016/j.ejmp.2015.06.008] [Citation(s) in RCA: 23] [Impact Index Per Article: 2.6] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 02/05/2015] [Revised: 06/11/2015] [Accepted: 06/11/2015] [Indexed: 01/06/2023] Open
|
18
|
Baghani HR, Aghamiri SMR, Mahdavi SR, Akbari ME, Mirzaei HR. Comparing the dosimetric characteristics of the electron beam from dedicated intraoperative and conventional radiotherapy accelerators. J Appl Clin Med Phys 2015; 16:5017. [PMID: 26103175 PMCID: PMC5690101 DOI: 10.1120/jacmp.v16i2.5017] [Citation(s) in RCA: 28] [Impact Index Per Article: 3.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/08/2014] [Revised: 11/11/2014] [Accepted: 10/31/2014] [Indexed: 11/28/2022] Open
Abstract
The specific design of the mobile dedicated intraoperative radiotherapy (IORT) accelerators and different electron beam collimation system can change the dosimetric characteristics of electron beam with respect to the conventional accelerators. The aim of this study is to measure and compare the dosimetric characteristics of electron beam produced by intraoperative and conventional radiotherapy accelerators. To this end, percentage depth dose along clinical axis (PDD), transverse dose profile (TDP), and output factor of LIAC IORT and Varian 2100C/D conventional radiotherapy accelerators were measured and compared. TDPs were recorded at depth of maximum dose. The results of this work showed that depths of maximum dose, R90,R50, and RP for LIAC beam are lower than those of Varian beam. Furthermore, for all energies, surface doses related to the LIAC beam are substantially higher than those of Varian beam. The symmetry and flatness of LIAC beam profiles are more desirable compared to the Varian ones. Contrary to Varian accelerator, output factor of LIAC beam substantially increases with a decrease in the size of the applicator. Dosimetric characteristics of beveled IORT applicators along clinical axis were different from those of the flat ones. From these results, it can be concluded that dosimetric characteristics of intraoperative electron beam are substantially different from those of conventional clinical electron beam. The dosimetric characteristics of the LIAC electron beam make it a useful tool for intraoperative radiotherapy purposes. PACS number: 87.56.‐v, 87.56.bd
Collapse
|
19
|
Kry SF, Popple R, Molineu A, Followill DS. Ion recombination correction factors (P(ion)) for Varian TrueBeam high-dose-rate therapy beams. J Appl Clin Med Phys 2012; 13:3803. [PMID: 23149774 PMCID: PMC5718527 DOI: 10.1120/jacmp.v13i6.3803] [Citation(s) in RCA: 30] [Impact Index Per Article: 2.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/01/2011] [Revised: 07/12/2012] [Accepted: 07/22/2012] [Indexed: 11/23/2022] Open
Abstract
Ion recombination is approximately corrected for in the Task Group 51 protocol by Pion, which is calculated by a two‐voltage measurement. This measurement approach may be a poor estimate of the true recombination, particularly if Pion is large (greater than 1.05). Concern exists that Pion in high‐dose‐per‐pulse beams, such as flattening filter free (FFF) beams, may be unacceptably high, rendering the two‐voltage measurement technique inappropriate. Therefore, Pion was measured for flattened beams of 6, 10, 15, and 18 MV and for FFF beams of 6 and 10 MV. The values for the FFF beams were verified with 1/V versus 1/Q curves (Jaffé plots). Pion was also measured for electron beams of 6, 12, 16, 18, and 20 MeV on a traditional accelerator, as well as on the high‐dose‐rate Varian TrueBeam accelerator. The measurements were made at a range of depths and with PTW, NEL, and Exradin Farmer‐type chambers. Consistent with the increased dose per pulse, Pion was higher for FFF beams than for flattening filter beams. However, for all beams, measurement locations, and chambers examined, Pion never exceeded 1.018. Additionally, Pion was always within 0.3% of the recombination calculated from the Jaffé plots. We conclude that ion recombination can be adequately accounted for in high‐dose‐rate FFF beams using Pion determined with the standard two‐voltage technique. PACS numbers: 87.56.‐v, 87.56.Da
Collapse
Affiliation(s)
- Stephen F Kry
- Department of Radiation Physics, Radiological Physics Center, The University of Texas M. D. Anderson Cancer Center, Houston, TX 77030, USA.
| | | | | | | |
Collapse
|
20
|
Iaccarino G, Strigari L, D'Andrea M, Bellesi L, Felici G, Ciccotelli A, Benassi M, Soriani A. Monte Carlo simulation of electron beams generated by a 12 MeV dedicated mobile IORT accelerator. Phys Med Biol 2011; 56:4579-96. [DOI: 10.1088/0031-9155/56/14/022] [Citation(s) in RCA: 35] [Impact Index Per Article: 2.7] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
|
21
|
Karsch L, Richter C, Pawelke J. Experimentelle Untersuchung der Sättigungskorrektion einer PTW Roos-Ionisationskammer in gepulsten Strahlungsfeldern mit hoher Pulsdosis bei verschiedenen Pulsdauern. Z Med Phys 2011; 21:4-10. [DOI: 10.1016/j.zemedi.2010.10.008] [Citation(s) in RCA: 12] [Impact Index Per Article: 0.9] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/26/2009] [Revised: 10/22/2010] [Accepted: 10/25/2010] [Indexed: 10/18/2022]
|
22
|
Cella L, Liuzzi R, Salvatore M. The Italian affair: The employment of parallel-plate ionization chambers for dose measurements in high dose-per-pulse IORT electron beams. Med Phys 2010; 37:2918-24. [DOI: 10.1118/1.3432601] [Citation(s) in RCA: 24] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
|
23
|
Palmans H, Thomas RAS, Duane S, Sterpin E, Vynckier S. Ion recombination for ionization chamber dosimetry in a helical tomotherapy unit. Med Phys 2010; 37:2876-89. [DOI: 10.1118/1.3427411] [Citation(s) in RCA: 39] [Impact Index Per Article: 2.8] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
|
24
|
Fiandra C, Ragona R, Ricardi U, Anglesio S, Giglioli FR. Absolute and relative dose measurements with Gafchromic™ EBT film for high energy electron beams with different doses per pulse. Med Phys 2008; 35:5463-70. [DOI: 10.1118/1.3005975] [Citation(s) in RCA: 12] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
|
25
|
Castro P, García-Vicente F, Mínguez C, Floriano A, Sevillano D, Pérez L, Torres JJ. Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration. J Appl Clin Med Phys 2008; 9:70-86. [PMID: 18449162 PMCID: PMC5721533 DOI: 10.1120/jacmp.v9i1.2676] [Citation(s) in RCA: 19] [Impact Index Per Article: 1.2] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/23/2007] [Revised: 06/24/2007] [Accepted: 08/21/2007] [Indexed: 11/23/2022] Open
Abstract
To achieve a good clinical outcome in radiotherapy treatment, a certain accuracy in the dose delivered to the patient is required. Therefore, it is necessary to keep the uncertainty in each of the steps of the process inside some acceptable values, which implies as low a global uncertainty as possible. The work reported here focused on the uncertainty evaluation of absorbed dose to water in the routine calibration for clinical beams in the range of energies used in external‐beam radiotherapy. With this aim, we considered various uncertainty components (corrected electrometer reading, calibration factor, beam quality correction factor, and reference conditions) associated with beam calibration. Results show a typical uncertainty in the determination of absorbed dose to water during beam calibration of approximately 1.3% for photon beams and 1.5% for electron beams (k=1 in both cases) when the ND,w formalism is used and kQ,Q0 is calculated theoretically. These values may vary depending on the uncertainty provided by the standards laboratory for calibration factor, which is shown in the work. For primary standards based on clinical linear accelerator beam energies, the uncertainty in this step of the process could be placed close to 1.0%. We also discuss the possibility of an uncertainty reduction with the adoption of the absorbed dose to water formalism as compared with the air kerma formalism. PACS numbers: 87.53.Dq, 87.53.Hv
Collapse
Affiliation(s)
- Pablo Castro
- Servicio de Oncología Radioterápica, Departamento de Radiofísica, Hospital Universitario "La Princesa", Madrid, Spain
| | - Feliciano García-Vicente
- Servicio de Oncología Radioterápica, Departamento de Radiofísica, Hospital Universitario "La Princesa", Madrid, Spain
| | - Cristina Mínguez
- Servicio de Oncología Radioterápica, Departamento de Radiofísica, Hospital Universitario "La Princesa", Madrid, Spain
| | - Alejandro Floriano
- Servicio de Oncología Radioterápica, Departamento de Radiofísica, Hospital Universitario "La Princesa", Madrid, Spain
| | - David Sevillano
- Servicio de Oncología Radioterápica, Departamento de Radiofísica, Hospital Universitario "La Princesa", Madrid, Spain
| | - Leopoldo Pérez
- Servicio de Oncología Radioterápica, Departamento de Radiofísica, Hospital Universitario "La Princesa", Madrid, Spain
| | - Juan J Torres
- Servicio de Oncología Radioterápica, Departamento de Radiofísica, Hospital Universitario "La Princesa", Madrid, Spain
| |
Collapse
|
26
|
Soriani A, Landoni V, Marzi S, Iaccarino G, Saracino B, Arcangeli G, Benassi M. Setup verification and in vivo dosimetry during intraoperative radiation therapy (IORT) for prostate cancer. Med Phys 2007; 34:3205-10. [PMID: 17879783 DOI: 10.1118/1.2750965] [Citation(s) in RCA: 24] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
The purpose of this study was to check the setup and dose delivered to the patients during intraoperative electron beam radiation therapy (IORT) for prostate cancer. Twenty eight patients underwent IORT after radical prostatectomy for prostate cancer by means of a dedicated mobile accelerator, Novac7 (by Hitesys, SpA, Italy). A 9 MeV electron beam at high dose per pulse was used. Eighteen patients received IORT at escalating doses of 16, 18, and 20 Gy at 85% isodose, six patients for each dose level. Further, ten patients received 20 Gy at 85% isodose. The electron applicator position was checked in all cases by means of two orthogonal images obtained with brilliance intensifier. Target and organ at risk doses were measured in vivo by a MOSFETs dosimetry system. MOSFETs and microMOSFET dosimeters were inserted into sterile catheters and directly positioned into the rectal lumen, for ten patients, and into the bladder to urethra anastomosis, in the last 14 cases. Verification at 0 degree led to very few adjustments of setup while verifications at 90 degrees often suggested to bring the applicator closer to the target. In vivo dosimetry showed an absorbed dose into the rectum wall < or =1% of the total dose. The average dose value inside the anastomosis, for the 12 patients analyzed, was 23.7 Gy with a standard deviation of +/-7.6%, when the prescription was 20 Gy at 85% isodose. Using a C-arm mobile image intensifier, it is possible to assess if the positioning is correct and safe. Radio-opaque clips and liquid were necessary to obtain good visible images. In vivo MOSFETs dosimetry is feasible and reliable. A satisfactory agreement between measured and expected doses was found.
Collapse
Affiliation(s)
- Antonella Soriani
- Laboratory of Medical Physics, Istituto Regina Elena, via Elio Chianesi 53, 00144, Roma, Italy.
| | | | | | | | | | | | | |
Collapse
|
27
|
Sureka CS, Sunny CS, Subbaiah KV, Aruna P, Ganesan S. Dose distribution for endovascular brachytherapy using Ir-192 sources: comparison of Monte Carlo calculations with radiochromic film measurements. Phys Med Biol 2006; 52:525-37. [PMID: 17202631 DOI: 10.1088/0031-9155/52/2/015] [Citation(s) in RCA: 9] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
An analysis of Ir-192 source distribution using the Monte Carlo method and radiochromic film experiments for endovascular brachytherapy is presented. Three different source possibilities, namely, mHDR Ir-192 sources with 5 mm and 2.5 mm step sizes and Ir-192 seed sources with 1 mm air gap are investigated to obtain uniform radial dose distribution throughout the treatment area. From this study, it is inferred that mHDR Ir-192 sources with 2.5 mm step size are effective for getting dose uniformity. Hence, different restenosis geometries, namely, linear, dumb bell and hairpin, are simulated with 2.5 mm step size, 15 mHDR Ir-192 sources using the Monte Carlo technique and the results are compared experimentally by using radiochromic films. The results from both methods agreed to within 7%. Further, it is also inferred that for the dosimetry of endovascular brachytherapy, the film dosimetry may be considered adequate, even if the film calibration is time consuming and requires adequate dosimetric procedures.
Collapse
Affiliation(s)
- C S Sureka
- Division of Medical Physics and Lasers, Department of Physics, Anna University, Chennai 600 025, India
| | | | | | | | | |
Collapse
|
28
|
Bruggmoser G, Saum R, Schmachtenberg A, Schmid F, Schüle E. Determination of the recombination correction factorkSfor some specific plane-parallel and cylindrical ionization chambers in pulsed photon and electron beams. Phys Med Biol 2006; 52:N35-50. [PMID: 17202615 DOI: 10.1088/0031-9155/52/2/n01] [Citation(s) in RCA: 50] [Impact Index Per Article: 2.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
Abstract
It has been shown from an evaluation of the inverse reading of the dosemeter (1/M) against the inverse of the polarizing voltage (1/V), obtained with a number of commercially available ionization chambers, using dose per pulse values between 0.16 and 5 mGy, that a linear relationship between the recombination correction factor kS and dose per pulse (DPP) can be found. At dose per pulse values above 1 mGy the method of a general equation with coefficients dependent on the chamber type gives more accurate results than the Boag method. This method was already proposed by Burns and McEwen (1998, Phys. Med. Biol. 43 2033) and avoids comprehensive and time-consuming measurements of Jaffé plots which are a prerequisite for the application of the multi-voltage analysis (MVA) or the two-voltage analysis (TVA). We evaluated and verified the response of ionization chambers on the recombination effect in pulsed accelerator beams for both photons and electrons. Our main conclusions are: (1) The correction factor k(S) depends only on the DPP and the chamber type. There is no influence of radiation type and energy. (2) For all the chambers investigated there is a linear relationship between kS and DPP up to 5 mGy/pulse, and for two chambers we could show linearity up to 40 mGy/pulse. (3) A general formalism, such as that of Boag, characterizes chambers exclusively by the distance of the electrodes and gives a trend for the correction factor, and therefore (4) a general formalism has to reflect the influence of the chamber construction on the recombination by the introduction of chamber-type dependent coefficients.
Collapse
Affiliation(s)
- G Bruggmoser
- Univ.-Klinikum Freiburg, Klinik f Strahlenheilkunde, FB Med. Physik, Freiburg, Germany.
| | | | | | | | | |
Collapse
|
29
|
Laitano RF, Guerra AS, Pimpinella M, Caporali C, Petrucci A. Charge collection efficiency in ionization chambers exposed to electron beams with high dose per pulse. Phys Med Biol 2006; 51:6419-36. [PMID: 17148826 DOI: 10.1088/0031-9155/51/24/009] [Citation(s) in RCA: 60] [Impact Index Per Article: 3.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
Abstract
The correction for charge recombination was determined for different plane-parallel ionization chambers exposed to clinical electron beams with low and high dose per pulse, respectively. The electron energy was nearly the same (about 7 and 9 MeV) for any of the beams used. Boag's two-voltage analysis (TVA) was used to determine the correction for ion losses, k(s), relevant to each chamber considered. The presence of free electrons in the air of the chamber cavity was accounted for in determining k(s) by TVA. The determination of k(s) was made on the basis of the models for ion recombination proposed in past years by Boag, Hochhäuser and Balk to account for the presence of free electrons. The absorbed dose measurements in both low-dose-per-pulse (less than 0.3 mGy per pulse) and high-dose-per-pulse (20-120 mGy per pulse range) electron beams were compared with ferrous sulphate chemical dosimetry, a method independent of the dose per pulse. The results of the comparison support the conclusion that one of the models is more adequate to correct for ion recombination, even in high-dose-per-pulse conditions, provided that the fraction of free electrons is properly assessed. In this respect the drift velocity and the time constant for attachment of electrons in the air of the chamber cavity are rather critical parameters because of their dependence on chamber dimensions and operational conditions. Finally, a determination of the factor k(s) was also made by zero extrapolation of the 1/Q versus 1/V saturation curves, leading to the conclusion that this method does not provide consistent results in high-dose-per-pulse beams.
Collapse
Affiliation(s)
- R F Laitano
- Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti, ENEA Centro Ricerche Casaccia, c.p. 2400 Roma, Italy
| | | | | | | | | |
Collapse
|
30
|
Beddar AS, Biggs PJ, Chang S, Ezzell GA, Faddegon BA, Hensley FW, Mills MD. Intraoperative radiation therapy using mobile electron linear accelerators: report of AAPM Radiation Therapy Committee Task Group No. 72. Med Phys 2006; 33:1476-89. [PMID: 16752582 DOI: 10.1118/1.2194447] [Citation(s) in RCA: 80] [Impact Index Per Article: 4.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
Intraoperative radiation therapy (IORT) has been customarily performed either in a shielded operating suite located in the operating room (OR) or in a shielded treatment room located within the Department of Radiation Oncology. In both cases, this cancer treatment modality uses stationary linear accelerators. With the development of new technology, mobile linear accelerators have recently become available for IORT. Mobility offers flexibility in treatment location and is leading to a renewed interest in IORT. These mobile accelerator units, which can be transported any day of use to almost any location within a hospital setting, are assembled in a nondedicated environment and used to deliver IORT. Numerous aspects of the design of these new units differ from that of conventional linear accelerators. The scope of this Task Group (TG-72) will focus on items that particularly apply to mobile IORT electron systems. More specifically, the charges to this Task Group are to (i) identify the key differences between stationary and mobile electron linear accelerators used for IORT, (ii) describe and recommend the implementation of an IORT program within the OR environment, (iii) present and discuss radiation protection issues and consequences of working within a nondedicated radiotherapy environment, (iv) describe and recommend the acceptance and machine commissioning of items that are specific to mobile electron linear accelerators, and (v) design and recommend an efficient quality assurance program for mobile systems.
Collapse
Affiliation(s)
- A Sam Beddar
- Department of Radiation Physics, Division of Radiation Oncology, Unit 94, The University of Texas M.D. Anderson Cancer Center, Houston, Texas 77030, USA.
| | | | | | | | | | | | | |
Collapse
|
31
|
Calvo FA, Meirino RM, Orecchia R. Intraoperative radiation therapy first part: rationale and techniques. Crit Rev Oncol Hematol 2006; 59:106-15. [PMID: 16844383 DOI: 10.1016/j.critrevonc.2005.11.004] [Citation(s) in RCA: 64] [Impact Index Per Article: 3.6] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/30/2005] [Revised: 11/17/2005] [Accepted: 11/17/2005] [Indexed: 11/15/2022] Open
Abstract
Intraoperative radiotherapy (IORT) is a technique where a high, single-fraction radiation dose is delivered during a surgical procedure to macroscopic tumours or tumour beds with minimal exposure of surroundings tissues which are displaced and shielded during the procedure. In this paper, the rationale for and use of IORT, both with electron beams (IOERT) and high-dose-rate brachytherapy (HDR-IORT) are discussed. For most tumours, the likelihood of obtaining local control (LC) improves when increasing doses can be administered. In many clinical situations, however, the dose that can be delivered safely to the tumour target is limited by the risk of damaging normal tissues. Special consideration is therefore given on this paper to the relationship between dose, LC and possible complications. Criteria for patient's selection and evaluation and information on sequencing and techniques are presented as well as some considerations on the need for a proper programme on quality assurance and periodical reporting of data.
Collapse
Affiliation(s)
- Felipe A Calvo
- Hospital General Universitario Gregorio Marañon, Madrid, Spain.
| | | | | |
Collapse
|
32
|
Ciocca M, Piazzi V, Lazzari R, Vavassori A, Luini A, Veronesi P, Galimberti V, Intra M, Guido A, Tosi G, Veronesi U, Orecchia R. Real-time in vivo dosimetry using micro-MOSFET detectors during intraoperative electron beam radiation therapy in early-stage breast cancer. Radiother Oncol 2006; 78:213-6. [PMID: 16359743 DOI: 10.1016/j.radonc.2005.11.011] [Citation(s) in RCA: 40] [Impact Index Per Article: 2.2] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/16/2005] [Revised: 10/17/2005] [Accepted: 11/23/2005] [Indexed: 11/24/2022]
Abstract
PURPOSE In a previous paper we reported the results of off-line in vivo measurements using radiochromic films in IOERT. In the present study, a further step was made, aiming at the improvement of the effectiveness of in vivo dosimetry, based on a real-time check of the dose. MATERIALS AND METHODS Entrance dose was determined using micro-MOSFET detectors placed inside a thin, sterile, transparent catheter. The epoxy side of the detector was faced towards the beam to minimize the anisotropy. Each detector was plugged into a bias supply (standard sensitivity) and calibrated at 5 Gy using 6 MeV electrons produced by a conventional linac. Detectors were characterized in terms of linearity, precision and dose per pulse dependence. No energy and temperature dependence was found. The sensitivity change of detectors was about 1% per 20 Gy accumulated dose. Correction factors to convert surface to entrance dose were determined for each combination of energy and applicator. From November 2004 to May 2005, in vivo dosimetry was performed on 45 patients affected by early-stage breast cancer, who underwent IOERT to the tumour bed. IOERT was delivered using electrons (4-10 MeV) at high dose per pulse, produced by either a Novac7 or a Liac mobile linac. RESULTS The mean ratio between measured and expected dose was 1.006+/-0.035 (1 SD), in the range 0.92-1.1. The procedure uncertainty was 3.6%. Micro-MOSFETs appeared suitable for in vivo dosimetry in IOERT, although some unfavourable aspects, like the limited lifetime and the anisotropy with no build-up, were found. Prospectively, a real-time action level (+/-6%) on dose discrepancy was defined. CONCLUSIONS Excellent agreement between measured and expected doses was found. Real-time in vivo dosimetry appeared feasible, reliable and more effective than the method previously published.
Collapse
Affiliation(s)
- Mario Ciocca
- Department of Medical Physics, European Institute of Oncology, Milano, Italy.
| | | | | | | | | | | | | | | | | | | | | | | |
Collapse
|
33
|
De Angelis C, Soriani A, Benassi M, Onori S. On measuring the output of an IORT mobile dedicated accelerator. RADIATION PROTECTION DOSIMETRY 2006; 120:221-5. [PMID: 16644963 DOI: 10.1093/rpd/nci621] [Citation(s) in RCA: 9] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/08/2023]
Abstract
In the present work, dose-to-water values derived from Fricke (operated by the Italian Primary Standard Dosimetry Laboratory) and alanine (operated by the Istituto Superiore di Sanità, Italy) measurements in IORT electron beams characterised by a high dose per pulse were compared to show the extent of equivalence of the two dosimetry systems. This study demonstrates agreement (within 2%) of the two dosimetry systems for plane-base IORT applicators, but in the case of small diameter (40 mm) and bevelled (22.5 degrees ) applicators, Fricke dosemeters underestimated doses by 2.4%. For base bevelled IORT applicators (22.5 degrees or more) with small diameter ( approximately 40 mm or less) reduced dimensions of the dosemeter are needed. Under these measurement conditions, the alanine dosemeter gives better accuracy in beam output determination compared with the Fricke dosemeter used.
Collapse
Affiliation(s)
- C De Angelis
- Department of Technology and Health, Istituto Superiore di Sanità, Viale Regina Elena, 299, 00161 Rome, Italy.
| | | | | | | |
Collapse
|
34
|
Consorti R, Petrucci A, Fortunato F, Soriani A, Marzi S, Iaccarino G, Landoni V, Benassi M. In vivo dosimetry with MOSFETs: dosimetric characterization and first clinical results in intraoperative radiotherapy. Int J Radiat Oncol Biol Phys 2005; 63:952-60. [PMID: 16199324 DOI: 10.1016/j.ijrobp.2005.02.049] [Citation(s) in RCA: 42] [Impact Index Per Article: 2.2] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/06/2004] [Revised: 02/11/2005] [Accepted: 02/21/2005] [Indexed: 11/22/2022]
Abstract
PURPOSE To investigate the use of metal oxide silicon field effect transistors (MOSFETs) as in vivo dosimetry detectors during electron beams at high dose-per-pulse intraoperative radiotherapy. METHODS AND MATERIALS The MOSFET system response in terms of reproducibility, energy, dose rate and temperature dependence, dose-linearity from 1 to 25 Gy, angular response, and dose perturbation was analyzed in the 6-9-MeV electron beam energy range produced by an intraoperative radiotherapy-dedicated mobile accelerator. We compared these with the 6- and 9-MeV electron beams produced by a conventional accelerator. MOSFETs were also used in clinical dosimetry. RESULTS In experimental conditions, the overall uncertainty of the MOSFET response was within 3.5% (+/-SD). The investigated electron energies and the dose rate did not significantly influence the MOSFET calibration factors. The dose perturbation was negligible. In vivo dosimetry results were in accordance with the predicted values within +/-5%. A discordance occurred either for an incorrect position of the dosimeter on the patient or when a great difference existed between the clinical and calibration setup, particularly when performing exit dose measurements. CONCLUSION Metal oxide silicon field effect transistors are suitable for in vivo dosimetry during intraoperative radiotherapy because their overall uncertainty is comparable to the accuracy required in target dose delivery.
Collapse
Affiliation(s)
- Rita Consorti
- Unità Operativa di Fisica Sanitaria, H.S. Filippo Neri, Roma, Italy.
| | | | | | | | | | | | | | | |
Collapse
|
35
|
Orecchia R, Ciocca M, Tosi G, Franzetti S, Luini A, Gatti G, Veronesi U. Intraoperative electron beam radiotherapy (ELIOT) to the breast: A need for a quality assurance programme. Breast 2005; 14:541-6. [PMID: 16242331 DOI: 10.1016/j.breast.2005.08.038] [Citation(s) in RCA: 31] [Impact Index Per Article: 1.6] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/24/2022] Open
Abstract
Intraoperative radiotherapy (IORT) is a technique in which a high, single-fraction radiation dose is delivered directly to the tumour bed during a surgical intervention, after the removal of a neoplastic mass. IORT has been recently used in early stage cancer as an exclusive radiation modality, rather than as a boost, especially for breast tumours, in particular at the European Institute of Oncology in Milan, where the technique has been called electron intraoperative therapy (ELIOT). Our studies on more than 1000 patients have demonstrated the feasibility of the technique and it is expected that its application will become more widespread in the immediate future. It is important to emphasise that ELIOT relies not only on new technological developments, but also on a multidisciplinary team with clear roles and responsibilities, the establishment of a programme of quality assurance with appropriate guidelines and a comprehensive staff development programme.
Collapse
Affiliation(s)
- R Orecchia
- Radiation Oncology Department, European Institute of Oncology, University of Milan, Via Ripamonti 435 20141 Milan, Italy.
| | | | | | | | | | | | | |
Collapse
|
36
|
Di Martino F, Giannelli M, Traino AC, Lazzeri M. Ion recombination correction for very high dose-per-pulse high-energy electron beams. Med Phys 2005; 32:2204-2210. [PMID: 16121574 DOI: 10.1118/1.1940167] [Citation(s) in RCA: 65] [Impact Index Per Article: 3.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/09/2004] [Accepted: 04/30/2005] [Indexed: 11/07/2022] Open
Abstract
The parallel-plate ionization chamber is the recommended tool for the absorbed dose measurement in pulsed high-energy electron beams. Typically, the electron beams used in radiotherapy have a dose-per-pulse value less then 0.1 cGy/pulse. In this range the factor to correct the response of an ionization chamber for the lack of complete charge collection due to ion recombination (ksat) can be properly evaluated with the standard "two voltage" method proposed by the international dosimetric reports. Very high dose-per-pulse electron beams are employed in some special Linac dedicated to the Intra-Operatory-Radiation-Therapy (IORT). The high dose-per-pulse values (3-13 cGy/pulse) characterizing the IORT electron beams allow to deliver the therapeutic dose (10-20 Gy) in less than a minute. This considerably reduces the IORT procedure time, but some dosimetric problems arise because the standard method to evaluate ksat overestimates its value by 20%. Moreover, if the dose-per-pulse value >1 cGy/pulse, the dependence of ksat on the dose-per-pulse value cannot be neglected for relative dosimetry. In this work the dependence of ksat on the dose-per-pulse value is derived, based on the general equation that describes the ion recombination in the Boag theory. A new equation for ksat, depending on known or measurable quantities, is presented. The new ksat equation is experimentally tested by comparing the absorbed doses to water measured with parallel-plate ionization chambers (Roos and Markus) to that measured using dose-per-pulse independent dosimeters, such as radiochromic films and chemical Fricke dosimeters. These measurements are performed in the high dose-per-pulse (3-13 cGy/pulse) electron beams of the IORT dedicated Linac Hitesys Novac7 (Aprilia-Latina, Italy). The dose measurements made using the parallel-plate chambers and those made using the dose-per-pulse independent dosimeters are in good agreement (<3%). This demonstrates the possibility of using the parallel-plate ionization chambers also for the very high dose-per-pulse (> 1 cGy/pulse) electron-beam dosimetry.
Collapse
Affiliation(s)
- F Di Martino
- UO Fisica Sanitaria, Sezione di Fisica Medica, Azienda Ospedaliera Universitaria Pisana, via Roma 67, 56126 Pisa, Italy.
| | | | | | | |
Collapse
|
37
|
Berg M, Noerrevang O. Recombination factors for the cylindrical FC65-G ionization chamber in pulsed photon beams and the plane-parallel Roos ionization chamber in pulsed electron beams. Phys Med Biol 2004; 49:5309-18. [PMID: 15656279 DOI: 10.1088/0031-9155/49/23/008] [Citation(s) in RCA: 14] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
The use of ionization chambers in linac radiotherapy dosimetry requires various corrections to the measured charges, one of these being the recombination correction. The recombination correction factor (k(s)) is generally estimated from the two-voltage analysis (TVA) for each beam quality. However, it is possible that the ionization chamber above some threshold polarizing voltage does not follow the accepted Boag theory very well. Secondly the TVA is time-consuming as the chamber needs to stabilize after each polarizing voltage change and since it must be performed for each beam quality. Another approach consists in using the fact that k(s) is predicted to depend linearly on dose per pulse by Boag theory: determining this relationship once and for all using a multi-voltage analysis (MVA), one also checks the range validity of the Boag theory for the chamber considered. This work presents a thorough analysis of k(s) dependence on dose per pulse of FC65-G (cylindrical) and Roos (plane-parallel) ionization chambers in pulsed photon and electron beams, respectively. Within the uncertainties, the recombination factors are found to be independent of beam quality, and no deviation from the Boag theory is observed within the tested range of polarizing voltages. Before adapting the equations given using the MVA other users should check that their ionization chambers show the same dose per pulse dependence using the TVA for a few beam qualities.
Collapse
Affiliation(s)
- Martin Berg
- Department of Medical Physics, Vejle Hospital, Kabbeltoft 25, DK-7100 Vejle, Denmark
| | | |
Collapse
|
38
|
Ciocca M, Orecchia R, Garibaldi C, Rondi E, Luini A, Gatti G, Intra M, Veronesi P, Lazzari R, Tosi G, Veronesi U. In vivo dosimetry using radiochromic films during intraoperative electron beam radiation therapy in early-stage breast cancer. Radiother Oncol 2003; 69:285-9. [PMID: 14644488 DOI: 10.1016/j.radonc.2003.09.001] [Citation(s) in RCA: 45] [Impact Index Per Article: 2.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/16/2022]
Abstract
BACKGROUND AND PURPOSE To check the dose delivered to patients during intraoperative electron beam radiation therapy (IOERT) for early breast cancer and also to define appropriate action levels. PATIENTS AND METHODS Between December 2000 and June 2001, 54 patients affected by early-stage breast cancer underwent exclusive IOERT to the tumour bed using a Novac7 mobile linac, after quadrantectomy. Electron beams (5, 7, 9 MeV) at high dose per pulse values (0.02-0.09 Gy/pulse) were used. The prescribed single dose was 21 Gy at the depth of 90% isodose (14-22 mm). In 35 cases, in vivo dosimetry was performed. The entrance dose was derived from the surface dose measured with thin and calibrated MD-55-2 radiochromic films, wrapped in sterile envelopes. Films were analysed 24-72 h after the irradiation using a charge-coupled-device imaging system. Field disturbance caused by the film envelope was negligible. RESULTS The mean deviation between measured and expected doses was 1.8%, with one SD equal to 4.7%. Deviations larger than 7% were found in 23% of cases, never consecutively, not correlated with beam energy or field size and with no evidence of linac daily output variation or serious malfunctioning or human mistake. The estimated overall uncertainty of dose measurement was about 4%. In vivo dosimetry appeared both reliable and feasible. Two action levels, for unexplained observed deviations larger than 7 and 10%, were preliminary defined. CONCLUSIONS Satisfactory agreement between measured and expected doses was found. The implementation of in vivo dosimetry in IOERT is suggested, particularly for patients enrolled in a clinical trial.
Collapse
Affiliation(s)
- Mario Ciocca
- Department of Medical Physics, European Institute of Oncology, via Ripamonti 435, 20141 Milano, Italy
| | | | | | | | | | | | | | | | | | | | | |
Collapse
|
39
|
Biggs PJ, Noyes RD, Willett CG. Clinical physics, applicator choice, technique, and equipment for electron intraoperative radiation therapy. Surg Oncol Clin N Am 2003; 12:899-924. [PMID: 14989123 DOI: 10.1016/s1055-3207(03)00101-7] [Citation(s) in RCA: 9] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/01/2022]
Abstract
IORT has been a widely used modality since the 1980s. The initial euphoria experienced at the beginning, however, has subsided, with the result that most centers still practicing IORT are academic institutions. The reason for the reduction in IORT performed at community hospitals is partly related to the method of treatment--namely, transporting the patient from the OR to the radiation therapy department. The advent of mobile linear accelerators, which require little or no shielding and can therefore be used in most OR rooms, is likely to reiginite interest in this modality. There are currently six new centers in the United States that practice IORT with a mobile linear accelerator and more than that in Europe.
Collapse
Affiliation(s)
- Peter J Biggs
- Department of Radiation Oncology, Massachusetts General Hospital, Fruit Street, Boston, MA 02114, USA.
| | | | | |
Collapse
|
40
|
Piermattei A, Fidanzio A, Perrone F, Azario L, Grimaldi L, Viola P, Capote R. Experimental dosimetry of a 32P catheter-based endovascular brachytherapy source. Phys Med Biol 2003; 48:2283-96. [PMID: 12953898 DOI: 10.1088/0031-9155/48/15/303] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
Abstract
The experimental dosimetry in a water phantom of a 32P linear source, 20 mm in length, used for the brachytherapy of coronary vessels is reported. The source content activity, A, was determined by means of a calibrated well ion-chamber and the value was compared with the contained activity reported in the manufacturer's certification. In this field of brachytherapy dosimetry, radiochromic film supplies a high enough spatial resolution. A highly sensitive radiochromic film, that presents only one active layer, was used in this work for the source dosimetry in a water phantom. The radiochromic film was characterized by electron beams produced by a clinical linac. A Monte Carlo calculation of beta spectra in water at different distances along the source transverse bisector axis allowed to take into account the low dependence of film response from the electron beam energy. The adopted experimental set-up, with the source in its catheter positioned on the film plane inside the water phantom, supplies accurate dosimetric information. The measured dose rate to water per unit of source activity at reference distance, D(r0, theta0)/A, in units of cGy s(-1) GBq(-1), was in agreement with the value reported in the manufacturer's certification within the experimental uncertainty. The radial dose function, g(r), is in good agreement with the literature data. The anisotropy function F(r, theta) is also reported. The analysis of the dose profile obtained at 2 mm from the source longitudinal axis shows that the uniformity is within 10% along 75% of the 20 mm treatment length. The adopted experimental set-up seems to be adequate for the quality control procedure of the dose homogeneity distribution in the water medium.
Collapse
Affiliation(s)
- A Piermattei
- Istituto di Fisica, Università Cattolica S Cuore, Roma, Italy
| | | | | | | | | | | | | |
Collapse
|
41
|
Piermattei A, Fidanzio A, Azario L, Russo A, Perrone F, Capote R, Toni MP. A standard dosimetry procedure for 192Ir sources used for endovascular brachytherapy. Phys Med Biol 2002; 47:4205-21. [PMID: 12502044 DOI: 10.1088/0031-9155/47/23/307] [Citation(s) in RCA: 12] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
Abstract
The experimental dosimetry of a high dose rate (HDR) 192Ir source used for the brachytherapy of peripheral vessels is reported. The direct determination of the reference air kerma rate Kr agrees, within the experimental uncertainty, with the results obtained by a well ionization chamber calibrated at the NIST and the manufacturer's certification. A highly sensitive (HS) radiochromic film (RCF), that presents only one active layer, was used for the source dosimetry in a water phantom. The adopted experimental set-up, with the source in its catheter positioned on the RCF plane, seems to have given better accuracy of the RCF optical density measurements. The agreement between the measurement of the dose rate constant DKr (10 mm, pi/2) and the literature data confirmed the coherence of the HS RCF calibration obtained by the kerma in air measurements. The RCF measurements supplied dosimetric information about the dose to water per reference air kerma rate D(r, theta)/Kr along the source transverse bisector axis, the radial dose function g(r) and the anisotropy function F(r, theta). The value D(2 mm, pi/2)/Kr = 22.4 +/- 1.2 cGy h(-1)/(microGy h(-1)) is supplied with a dose uncertainty that is essentially due to the indeterminacy of the source position in the catheter. The data of the radial and anisotropy functions have been compared with Monte Carlo determinations reported in the literature.
Collapse
Affiliation(s)
- A Piermattei
- Istituto di Fisica, Università Cattolica, S Cuore, Roma, Italy.
| | | | | | | | | | | | | |
Collapse
|
42
|
Clift MA, Johnston PN, Webb DV. A temporal method of avoiding the Cerenkov radiation generated in organic scintillator dosimeters by pulsed mega-voltage electron and photon beams. Phys Med Biol 2002; 47:1421-33. [PMID: 12030564 DOI: 10.1088/0031-9155/47/8/313] [Citation(s) in RCA: 76] [Impact Index Per Article: 3.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
The output signal of an organic scintillator probe consists of a scintillation signal and Cerenkov and fluorescence radiation (CFR) signal when the probe is exposed to a mega-voltage photon or electron beam. The CFR signal is usually unwanted because it comes from both the scintillator and light guide and so it is not proportional to the absorbed dose in the scintillator. A new organic scintillator detector system has been constructed for absorbed dose measurement in pulsed mega-voltage electron and photon beams that are commonly used in radiotherapy treatment, eliminating most of the CFR signal. The new detector system uses a long decay constant BC-444G (Bicron, Newbury, OH, USA) scintillator which gives a signal that can be time resolved from the prompt CFR signal so that the measured contribution of prompt signal is negligible. The response of the new scintillator detector system was compared with the measurements from a plastic scintillator detector that were corrected for the signal contribution from the CFR, and to appropriately corrected ion chamber measurements showing agreement in the 16 MeV electron beam used.
Collapse
Affiliation(s)
- M A Clift
- Department of Applied Physics, Royal Melbourne Institute of Technology, Victoria, Australia.
| | | | | |
Collapse
|