Rickard CM, Couchman BA, Schmidt SJ, Dank A, Purdie DM. A discard volume of twice the deadspace ensures clinically accurate arterial blood gases and electrolytes and prevents unnecessary blood loss.
Crit Care Med 2003;
31:1654-8. [PMID:
12794400 DOI:
10.1097/01.ccm.0000063448.98777.ef]
[Citation(s) in RCA: 27] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/26/2022]
Abstract
OBJECTIVE
To determine the blood discard volume, as a multiple of deadspace, that is required for accurate arterial blood gas and electrolyte testing from arterial catheters.
DESIGN
Prospective, controlled, crossover trial.
SETTING
Eighteen-bed intensive care unit of a metropolitan teaching hospital.
PATIENTS
A total of 84 critically ill patients with a 20-gauge, radial arterial cannulae, pressure monitoring transducer set, and stable oxygenation.
INTERVENTIONS
System deadspace (priming volume from sampling port to catheter tip) was established. Patients had six 0.5-mL arterial blood samples taken sequentially in random order using discard volumes of 1, 1.5, 2, 2.3, and 3.6 times the deadspace (experimental values) and 5.5 times the deadspace (control).
MEASUREMENTS AND MAIN RESULTS
Samples were analyzed for Pao(2), Sao(2), pH, Paco(2), HCO(3)-, Na+, and K+. We performed repeated-measures analysis of variance with post hoc linear contrasts and compared mean experimental and control values. The smallest discard volumes that provided measurements that were statistically equal to control were twice the deadspace (Pao(2), p =.563; Sao(2), p =.371) and 3.6 times the deadspace (pH, p =.107; Paco(2), p=.519; HCO(3)-, p =.10). All discard volumes tested provided results that were statistically different from control for Na+ (p <.003) and K+ (p <.001).
CONCLUSIONS
Many results were statistically different from control, although the actual discrepancies were very small. At clinically relevant levels of measurement, there was minimal variation between values obtained after a discard volume of twice the deadspace and control values. The level of error was clinically acceptable and within or close to the precision limits of the blood gas analyzer. Slight fluctuation in patient variables during sampling could also have contributed to the error. A blood discard volume of twice the deadspace is recommended for all variables. This will provide clinically accurate results and avoid the deleterious effects of unnecessary blood loss.