Monteiro T, Pinto C, Franqueira N, Faria-Correia F, Mendes J, Alfonso Sánchez J, Vaz F. Efficacy and Safety After Toric Posterior Chamber Implantable Collamer Lens and Toric Iris-Fixated Foldable Phakic Intraocular Lens for Myopic Astigmatism.
J Refract Surg 2022;
38:339-347. [PMID:
35686710 DOI:
10.3928/1081597x-20220406-01]
[Citation(s) in RCA: 4] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/20/2022]
Abstract
PURPOSE
To compare visual, refractive, and safety outcomes of toric posterior chamber Implantable Collamer Lens (T-ICL) (STAAR Surgical) and toric iris-fixated foldable phakic intraocular lens (IOL) (T-Artiflex; Ophtec BV) implantation for the correction of myopic astigmatism.
METHODS
This retrospective cohort study included 312 eyes of 312 patients who had phakic IOL implantation for myopic astigmatism. Two groups were defined: 205 eyes that underwent T-ICL implantation and 107 eyes that underwent T-Artiflex implantation. Safety, efficacy, and predictability outcomes were evaluated preoperatively and at 12 months postoperatively. Refractive and corneal astigmatic vector analysis were performed using the Alpins method.
RESULTS
One year postoperatively, uncorrected distance visual acuity was 0.05 ± 0.18 (T-ICL) and 0.10 ± 0.16 (T-Artiflex) logMAR, with efficacy indexes of 1.16 ± 0.27 and 1.05 ± 0.31, respectively (P < .001). Safety indexes were 1.28 ± 0.30 and 1.21 ± 0.31, respectively (P = .04). Spherical equivalent was within ±0.50 diopters (D) of emmetropia in 165 (80.5%) and 88 (82.2%) eyes, respectively. Refractive astigmatic analysis showed an index of success of 0.28 ± 0.33 (T-ICL) and 0.31 ± 0.26 (T-Artiflex) (P = .07). Surgically induced corneal astigmatism was 0.48 ± 0.74 and 0.81 ± 0.61 D, respectively (P < .001). Mean endothelial loss was 1.11% and 2.05%, respectively (P = .42). Six (2.9%) eyes in the T-ICL group and 1 (0.9%) eye in the T-Artiflex group had phakic IOL repositioning due to significant misalignment. No vision-threatening complications occurred.
CONCLUSIONS
Both the T-ICL and T-Artiflex groups showed high visual and refractive efficacy with a good safety profile for the correction of myopic astigmatism. T-ICL implantation demonstrated significantly better efficacy and safety indexes after 12 months. Vector analysis showed similar refractive astigmatic correction in both groups, but T-Artiflex implantation revealed higher surgically induced corneal astigmatism. [J Refract Surg. 2022;38(6):339-347.].
Collapse