Firstenberg MS, Morehead AJ, Thomas JD, Smedira NG, Cosgrove DM, Marchand MA. Short-term hemodynamic performance of the mitral Carpentier-Edwards PERIMOUNT pericardial valve. Carpentier-Edwards PERIMOUNT Investigators.
Ann Thorac Surg 2001;
71:S285-8. [PMID:
11388206 DOI:
10.1016/s0003-4975(01)02514-0]
[Citation(s) in RCA: 13] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/15/2022]
Abstract
BACKGROUND
Although long-term durability data exist, little data are available concerning the hemodynamic performance of the Carpentier-Edwards PERIMOUNT pericardial valve in the mitral position.
METHODS
Sixty-nine patients who were implanted with mitral PERIMOUNT valves at seven international centers between January 1996 and February 1997 consented to participate in a short-term echocardiography follow-up. Echocardiographs were collected at a mean of 600+/-133 days after implantation (range, 110 to 889 days); all underwent blinded core lab analysis.
RESULTS
At follow-up, peak gradients were 9.09+/-3.43 mm Hg (mean, 4.36+/-1.79 mm Hg) and varied inversely with valve size (p < 0.05). The effective orifice areas were 2.5+/-0.6 cm2 and tended to increase with valve size (p = 0.08). Trace mitral regurgitation (MR) was common (n = 48), 9 patients had mild MR, 1 had moderate MR, none had severe MR. All MR was central (n = 55) or indeterminate (n = 3). No paravalvular leaks were observed. Mitral regurgitation flow areas were 3.4+/-2.8 cm2 and were without significant volumes.
CONCLUSIONS
In this multicenter study, these mitral valves are associated with trace, although physiologically insignificant, central MR. Despite known echocardiographic limitations, the PERIMOUNT mitral valves exhibit similar hemodynamics to other prosthetic valves.