Evaluation of the agreement of two oscillometric blood pressure devices with invasive blood pressure in anaesthetized chimpanzees (Pan troglodytes).
Vet Anaesth Analg 2021;
48:688-696. [PMID:
34275756 DOI:
10.1016/j.vaa.2021.01.010]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/12/2020] [Revised: 11/12/2020] [Accepted: 01/08/2021] [Indexed: 11/21/2022]
Abstract
OBJECTIVE
To evaluate the agreement of two noninvasive blood pressure devices: a human device with the cuff placed on the wrist (Omron R1) and a veterinary device with the cuff placed on the upper brachium (Surgivet Advisor Vital Signs Monitor) with invasive blood pressure (IBP) measurement in anaesthetized chimpanzees.
STUDY DESIGN
Prospective clinical study.
ANIMALS
A convenience sample of 11 adult chimpanzees undergoing anaesthesia for translocation and routine health checks.
METHODS
Systolic (SAP) and diastolic arterial pressures (DAP) were continuously recorded via a transducer connected to a femoral artery cannula, and at 5 minute intervals from the two oscillometric devices. Agreement was explored using Bland-Altman analysis and bias defined as the mean difference between the two measurement methods. Spearman correlation coefficients were calculated. Significance was set at p < 0.05.
RESULTS
Bias and standard deviation for the Surgivet compared with IBP were 8.6 ± 18 for SAP and 8.4 ± 9.9 for DAP, showing a significant underestimation of both variables. Limits of agreement (LOA) were from -27 to 44 for SAP and from -11 to 28 for DAP. Correlation coefficients between the Surgivet and IBP values were 0.86 for SAP and 0.85 for DAP (p < 0.0001). Bias and standard deviation for the Omron compared with the IBP were -21 ± 25 for SAP and -18 ± 15 for DAP, showing a significant overestimation of both variables. LOA were from -70 to -28 for SAP and from -47 to 11 for DAP. Spearman correlation coefficients between the Omron and IBP values were 0.64 for SAP and 0.72 for DAP (p < 0.0001).
CONCLUSIONS AND CLINICAL RELEVANCE
Although neither device met all the criteria for device validation, the Surgivet presented better agreement with IBP values than the Omron in adult anaesthetized chimpanzees.