1
|
Sharma DS, Shaju P, Sawant MB, Kaushik S. Benchmarking a New Circular Cone-based Radiosurgery System against Clinically Tested Radiosurgery System on the same Novel Digital Linear Accelerator Platform. J Med Phys 2023; 48:111-119. [PMID: 37576095 PMCID: PMC10419751 DOI: 10.4103/jmp.jmp_93_22] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/01/2022] [Revised: 03/06/2023] [Accepted: 03/20/2023] [Indexed: 08/15/2023] Open
Abstract
Objective To examine the dosimetric characteristics of circular cones, the accuracy of dose modeling and overall treatment delivery of two radiosurgery systems integrated on a linear accelerator (Linac). Materials and Methods The dosimetric characteristics of circular cones (4-17.5 mm) from Varian (VC) and BrainLAB (BLC) were measured for 6 MV flattening filter free beam from Edge linac using stereotactic field diode and 0.65 cc ionization chamber following established protocols. The Eclipse and iPlan modeled dose distribution for VCs and BLCs were validated with EBT3-film measurement. End-to-end tests were performed using stereotactic phantom having PTW 60008 diode connected to a Dose-1 electrometer. Results The depth at dose maximum, TRP2010 and dose at 10cm depth of the same size VC and BLC agree within ± 0.7 mm, ± 0.71% and ± 0.81% respectively. Full width at half maximum (FWHM) of any cone beyond 15 mm depth increases at 1% of nominal cone size per 10 mm depth. The penumbra of 4mm and 17.5mm VC at 15 mm depth was 1.1 mm and 1.50 mm. At 300 mm depth, penumbra increased by around 0.4 mm for 4 mm cone and up to 1 mm for cone size ≥12.5 mm. The VCs penumbra values were within ±1mm of the corresponding BLCs. Scatter factors for VCs varies from 0.609 to 0.841 and were within ± 1.0% of corresponding values of BLCs. Agreement between the Eclipse and iPlan computed dose fluence and the EBT3-film measured dose fluence was >98% (γ: 1%@1 mm), and the absolute dose difference was ≤ 2.2%, except for the 4 mm cone in which it was >96% and ≤4.83%. Target localization using cone-beam computed tomography was accurate within ± 0.8 mm and ± 0.3° in translation and rotation. The end-to-end dose delivery accuracy for both radiosurgery systems was within ± 3.62%. Conclusion The dosimetric characteristics of Varian and BLC cones of same diameter was comparable. Both Eclipse and iPlan cone planning system modeled dose fluences agree well with the EBT3 film measurement. The end-to-end tests revealed an excellent target localization accuracy of Edge linac with satisfactory and comparable absolute dose agreement between Varian and BLC radiosurgery systems and hence these can be interchanged on edge linac.
Collapse
Affiliation(s)
| | - P Shaju
- Department of Radiation Oncology, Kokilaben Dhirubhai Ambani Hospital and Medical Research Institute, Mumbai, Maharashtra, India
| | - Mayur B Sawant
- Department of Medical Physics, Apollo Proton Cancer Centre, Chennai, Tamil Nadu, India
| | - Suryakant Kaushik
- Department of Medical Physics, Apollo Proton Cancer Centre, Chennai, Tamil Nadu, India
| |
Collapse
|
2
|
Brown TAD, Fagerstrom JM, Beck C, Holloway C, Burton K, Kaurin DGL, Mahendra S, Luckstead M, Kielar K, Kerns J. Determination of commissioning criteria for multileaf-collimator, stereotactic radiosurgery treatments on Varian TrueBeam and Edge machines using a novel anthropomorphic phantom. J Appl Clin Med Phys 2022; 23:e13581. [PMID: 35290710 PMCID: PMC9195028 DOI: 10.1002/acm2.13581] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/09/2021] [Revised: 10/22/2021] [Accepted: 02/21/2022] [Indexed: 11/11/2022] Open
Abstract
An anthropomorphic phantom has been developed by Varian Medical Systems for commissioning multileaf‐collimator (MLC), stereotactic radiosurgery (SRS) treatments on Varian TrueBeam and Edge linear accelerators. Northwest Medical Physics Center (NMPC) has collected end‐to‐end data on these machines, at six independent clinical sites, to establish baseline dosimetric and geometric commissioning criteria for SRS measurements with this phantom. The Varian phantom is designed to accommodate four interchangeable target cassettes, each designed for a specific quality assurance function. End‐to‐end measurements utilized the phantom to verify the coincidence of treatment isocenter with a hidden target in a Winston‐Lutz cassette after localization using cone‐beam computed tomography (CBCT). Dose delivery to single target (2 cm) and single‐isocenter, multitarget (2 and 1 cm) geometries was verified using ionization chamber and EBT3 film cassettes. A nominal dose of 16 Gy was prescribed for each plan using a site's standard beam geometry for SRS cases. Measurements were performed with three Millennium and three high‐definition MLC machines at beam energies of 6‐MV and 10‐MV flattening‐filter‐free energies. Each clinical site followed a standardized procedure for phantom simulation, treatment planning, quality assurance, and treatment delivery. All treatment planning and delivery was performed using ARIA oncology information system and Eclipse treatment planning software. The isocenter measurements and irradiated film were analyzed using DoseLab quality assurance software; gamma criteria of 3%/1 mm, 3%/0.5 mm, and 2%/1 mm were applied for film analysis. Based on the data acquired in this work, the recommended commissioning criteria for end‐to‐end SRS measurements with the Varian phantom are as follows: coincidence of treatment isocenter and CBCT‐aligned hidden target < 1 mm, agreement of measured chamber dose with calculated dose ≤ 5%, and film gamma passing > 90% for gamma criteria of 3%/1 mm after DoseLab auto‐registration shifts ≤ 1 mm in any direction.
Collapse
Affiliation(s)
| | | | - Caleb Beck
- Northwest Medical Physics Center, Lynnwood, Washington, USA
| | | | - Krista Burton
- Northwest Medical Physics Center, Lynnwood, Washington, USA
| | | | | | | | - Kayla Kielar
- Varian Medical Systems, Palo Alto, California, USA
| | - James Kerns
- Varian Medical Systems, Palo Alto, California, USA
| |
Collapse
|
3
|
Hermida-López M, Sánchez-Artuñedo D, Rodríguez M, Brualla L. Monte Carlo simulation of conical collimators for stereotactic radiosurgery with a 6 MV flattening-filter-free photon beam. Med Phys 2021; 48:3160-3171. [PMID: 33715167 DOI: 10.1002/mp.14837] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/15/2020] [Revised: 02/19/2021] [Accepted: 03/08/2021] [Indexed: 11/07/2022] Open
Abstract
PURPOSE Conical collimators, or cones, are tertiary collimators that attach to a radiotherapy linac and are suited for the stereotactic radiosurgery treatment of small brain lesions. The small diameter of the most used cones makes difficult the acquisition of the dosimetry data needed for the commissioning of treatment planning systems. Although many publications report dosimetric data of conical collimators for stereotactic radiosurgery, most of the works use different setups, which complicates comparisons. In other cases, the cone output factors reported do not take into account the effect of the small cone diameter on the detector response. Finally, few data exist on the dosimetry of cones with flattening-filter-free (FFF) beams from modern linac models. This work aims at obtaining a dosimetric characterization of the conical collimators manufactured by Brainlab AG (Munich, Germany) in a 6 MV FFF beam from a TrueBeam STx linac (Varian Medical Systems). METHODS Percentage depth dose curves, lateral dose profiles and cone output factors were obtained using Monte Carlo simulations for the cones with diameters of 4, 5, 6, 7.5, 8, 10, 12.5, 15, 17.5, 20, 25, and 30 mm. The simulation of the linac head was carried out with the PRIMO Monte Carlo software, and the simulations of the cones and the water phantom were run with the general-purpose Monte Carlo code PENELOPE. The Monte Carlo model was validated by comparing the simulation results with measurements performed for the cones of 4, 5, and 7.5 mm of diameter using a stereotactic field diode, a microDiamond detector and EBT3 radiochromic film. In addition, for those cones, simulations and measurements were done for comparison purposes, by reproducing the experimental setups from the available publications. RESULTS The experimental data acquired for the cones of 4, 5, and 7.5 mm validated the developed Monte Carlo model. The simulations accurately reproduced the experimental depths of maximum dose and the dose ratio at 20- and 10-cm depth (PDD20/10 ). A good agreement was obtained between simulated and experimental lateral dose profiles: The differences in the full-width at half-maximum were smaller than 0.2 mm, and the differences in the penumbra 80%-20% were smaller than 0.25 mm. The difference between the simulated and the average of the experimental output factors for the cones of 4, 5, and 7.5 mm of diameter was 0.0%, 0.0%, and 3.0%, respectively, well within the statistical uncertainty of the simulations (4.4% with coverage factor k = 2). It was also found that the simulated cone output factors agreed within 2% with the average of output factors reported in the literature for a variety of setup conditions, detectors, beam qualities, and cone manufacturers. CONCLUSION A Monte Carlo model of cones for stereotactic radiosurgery has been developed and validated. The cone dosimetry dataset obtained in this work, consisting of percentage depth doses, lateral dose profiles and output factors, is useful to benchmark data acquired for the commissioning of cone-based radiosurgery treatment planning systems.
Collapse
Affiliation(s)
- Marcelino Hermida-López
- Servei de Física i Protecció Radiològica, Hospital Universitari Vall d'Hebron, Vall d'Hebron Barcelona Hospital Campus, Passeig Vall d'Hebron 119-129, Barcelona, 08035, Spain
| | - David Sánchez-Artuñedo
- Servei de Física i Protecció Radiològica, Hospital Universitari Vall d'Hebron, Vall d'Hebron Barcelona Hospital Campus, Passeig Vall d'Hebron 119-129, Barcelona, 08035, Spain
| | - Miguel Rodríguez
- Centro Médico Paitilla, Calle 53 y ave. Balboa, Panama City, Panama.,Instituto de Investigaciones Científicas y de Alta Tecnología, INDICASAT-AIP, City of Knowledge, Building 219, Panama City, Panama
| | - Lorenzo Brualla
- West German Proton Therapy Centre Essen (WPE), Hufelandstr. 55, Essen, 45147, Germany.,West German Cancer Centre (WTZ), Hufelandstr. 55, Essen, 45147, Germany.,Faculty of Medicine, University of Duisburg-Essen, Hufelandstr. 55, Essen, 45147, Germany
| |
Collapse
|
4
|
Niroomand‐Rad A, Chiu‐Tsao S, Grams MP, Lewis DF, Soares CG, Van Battum LJ, Das IJ, Trichter S, Kissick MW, Massillon‐JL G, Alvarez PE, Chan MF. Report of AAPM Task Group 235 Radiochromic Film Dosimetry: An Update to TG‐55. Med Phys 2020; 47:5986-6025. [DOI: 10.1002/mp.14497] [Citation(s) in RCA: 64] [Impact Index Per Article: 16.0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/09/2020] [Revised: 09/15/2020] [Accepted: 09/17/2020] [Indexed: 12/12/2022] Open
Affiliation(s)
| | | | | | | | | | | | - Indra J. Das
- Radiation Oncology Northwestern University Memorial Hospital Chicago IL USA
| | - Samuel Trichter
- New York‐Presbyterian HospitalWeill Cornell Medical Center New York NY USA
| | | | - Guerda Massillon‐JL
- Instituto de Fisica Universidad Nacional Autonoma de Mexico Mexico City Mexico
| | - Paola E. Alvarez
- Imaging and Radiation Oncology Core MD Anderson Cancer Center Houston TX USA
| | - Maria F. Chan
- Memorial Sloan Kettering Cancer Center Basking Ridge NJ USA
| |
Collapse
|
5
|
Rose MS, Tirpak L, Van Casteren K, Zack J, Simon T, Schoenfeld A, Simon W. Multi‐institution validation of a new high spatial resolution diode array for SRS and SBRT plan pretreatment quality assurance. Med Phys 2020; 47:3153-3164. [DOI: 10.1002/mp.14153] [Citation(s) in RCA: 22] [Impact Index Per Article: 5.5] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/18/2019] [Revised: 02/20/2020] [Accepted: 03/12/2020] [Indexed: 12/31/2022] Open
Affiliation(s)
- Mark S. Rose
- Sun Nuclear Corporation 3275 Suntree Blvd Melbourne Florida 32940 USA
| | - Lena Tirpak
- Sun Nuclear Corporation 3275 Suntree Blvd Melbourne Florida 32940 USA
| | | | - Jeff Zack
- Sun Nuclear Corporation 3275 Suntree Blvd Melbourne Florida 32940 USA
| | - Tom Simon
- Sun Nuclear Corporation 3275 Suntree Blvd Melbourne Florida 32940 USA
| | | | - William Simon
- Sun Nuclear Corporation 3275 Suntree Blvd Melbourne Florida 32940 USA
| |
Collapse
|
6
|
Three-dimensional dose comparison of flattening filter (FF) and flattening filter-free (FFF) radiation therapy by using NIPAM gel dosimetry. PLoS One 2019; 14:e0212546. [PMID: 30789968 PMCID: PMC6383886 DOI: 10.1371/journal.pone.0212546] [Citation(s) in RCA: 7] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/27/2018] [Accepted: 02/05/2019] [Indexed: 12/21/2022] Open
Abstract
Intensity-modulated radiotherapy and volumetric modulated arc therapy are modern radiation therapy technologies that can create the desired dose distribution by multileaf collimator movement and dose-rate control. However, the homogeneous dose delivery of small-field irradiation techniques shows disagreement with that of treatment planning system. The removal of the flattening filter by flattening filter free (FFF) beam irradiation increases dose conformity and further reduces treatment delivery time in radiosurgery. This study aims to investigate the dose distribution of FFF and flattened beams for small-field irradiation by using the 3D gel dosimeter. The N-isopropylacrylamide (NIPAM) polymer gel dosimeter was employed to record the 3D dose distribution. In addition, flattened and FFF beams were compared using the gamma evaluation technique. The use of an FFF accelerator resulted in excellent radiation treatments with short delivery times and low doses to normal tissues and organs. Results also showed that the passing rate increased with the decrease of field size (30 × 30, 20 × 20, and 10 × 10 mm2) at post-irradiation times of 24, 48, 72, and 96 h. The passing rates for each field size were retained at the same level when gamma criteria, namely, distance-to-agreement (DTA) = 3 mm/dose difference (DD) = 3%, were used. Nevertheless, the passing rates for each field size decreased slowly after 48 h when DTA = 2 mm/DD = 2%. The Wilcoxon signed-rank test was used to determine statistical difference with a significant level of p < 0.05. The passing rates of flattened and FFF beams showed no significant difference. The edge enhancement effect in the flattened beam was more evident than in the FFF beam. The 3D NIPAM gel dosimeter can be used for dose verification of small field for radiation therapy with high dose rate.
Collapse
|
7
|
Stevens MTR, Lobb EC, Yenice KM. Validation of MLC-based linac radiosurgery for trigeminal neuralgia. J Appl Clin Med Phys 2018; 19:214-221. [PMID: 29901278 PMCID: PMC6036389 DOI: 10.1002/acm2.12381] [Citation(s) in RCA: 6] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/26/2017] [Revised: 04/04/2018] [Accepted: 05/16/2018] [Indexed: 11/23/2022] Open
Abstract
This study details a validation process for linear accelerator‐based treatment of trigeminal neuralgia using HD‐MLC field collimation. Nine trigeminal neuralgia treatment plans utilizing HD‐MLC were selected for absolute dose measurement at isocenter using a commercial scintillating detector in an anthropomorphic phantom. Four plans were chosen for film dosimetry measurements in each of the three principal planes to assess spatial dose distribution agreement with the treatment planning system. Additionally, trajectory log analysis for each treatment field in the nine cases was performed to assess mechanical positioning accuracy of the MLC system during delivery. Scintillator and film measurements both revealed mean dose agreement at isocenter of better than 3% while FWHM of the 2D dose distribution in each plane showed agreement between plan and measurement within 0.2 mm. Analysis of log files revealed a maximum MLC leaf positioning error of 0.04 mm across 178 treatment fields. In conjunction with a quality‐controlled treatment delivery methodology, an appropriately commissioned treatment planning system can be used for accurate and clinically appropriate design of trigeminal neuralgia treatment plans utilizing HD‐MLC.
Collapse
Affiliation(s)
- M Tynan R Stevens
- Department of Radiation and Cellular Oncology, University of Chicago Medicine, Chicago, IL, USA
| | - Eric C Lobb
- Department of Radiation and Cellular Oncology, University of Chicago Medicine, Chicago, IL, USA
| | - Kamil M Yenice
- Department of Radiation and Cellular Oncology, University of Chicago Medicine, Chicago, IL, USA
| |
Collapse
|
8
|
Borzov E, Nevelsky A, Bar-Deroma R, Orion I. Dosimetric characterization of Elekta stereotactic cones. J Appl Clin Med Phys 2017; 19:194-203. [PMID: 29266744 PMCID: PMC5768017 DOI: 10.1002/acm2.12242] [Citation(s) in RCA: 13] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/15/2017] [Revised: 11/18/2017] [Accepted: 11/20/2017] [Indexed: 11/14/2022] Open
Abstract
Purpose Dosimetry of small fields defined by stereotactic cones remains a challenging task. In this work, we report the results of commissioning measurements for the new Elekta stereotactic conical collimator system attached to the Elekta VersaHD linac and present the comparison between the measured and Monte Carlo (MC) calculated data for the 6 MV FFF beam. In addition, relative output factor (ROF) dependence on the stereotactic cone aperture variation was studied and penumbra comparison for small MLC‐based and cone‐based fields was performed. Methods Cones with nominal diameters of 15 mm, 12.5 mm, 10 mm, 7.5 mm, and 5 mm were employed in our study. Percentage depth dose (PDD), off‐axis ratios (OAR), and ROF were measured using a stereotactic field diode (SFD). BEAMnrc code was used for MC simulations. Results MC calculated and measured PDDs for all cones agreed within 1%/0.5 mm, and OAR profiles agreed within 1%/0.5 mm. ROF obtained from the measurements and MC calculations agreed within 2% for all cone sizes. Small‐field correction factors for the SFD detector Kfield,3 × 3(SFD) were derived using MC calculations as a baseline and were found to be 0.982, 0.992, 0.997, 1.015, and 1.017 for the 5, 7.5, 10, 12.5, and 15‐mm cones respectively. The difference in ROF was about 10%, 6%, 3.5%, 3%, 2.5%, and 2% for ±0.3 mm variations in 5, 7.5, 10, 12.5, and 15‐mm cone aperture respectively. In case of single static field, cone‐based collimation produced a sharper penumbra compared to the MLC‐based. Conclusions Accurate MC simulation can be an effective tool for verification of dosimetric measurements of small fields. Due to the very high sensitivity of output factors on the cone diameter, manufacture‐related variations in cone size may lead to considerable variations in dosimetric characteristics of stereotactic cones.
Collapse
Affiliation(s)
- Egor Borzov
- Department of Radiotherapy, Division of Oncology, Rambam Health Care Campus, Haifa, Israel
| | - Alexander Nevelsky
- Department of Radiotherapy, Division of Oncology, Rambam Health Care Campus, Haifa, Israel
| | - Raquel Bar-Deroma
- Department of Radiotherapy, Division of Oncology, Rambam Health Care Campus, Haifa, Israel
| | - Itzhak Orion
- Department of Nuclear Engineering, Ben-Gurion University of the Negev, Beer Sheva, Israel
| |
Collapse
|
9
|
Chaudhari SH, Dobhal R, Kinhikar RA, Kadam SS, Deshpande DD. Measurement of Total Scatter Factor for Stereotactic Cones with Plastic Scintillation Detector. J Med Phys 2017; 42:9-13. [PMID: 28405102 PMCID: PMC5370342 DOI: 10.4103/jmp.jmp_114_16] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/13/2022] Open
Abstract
Advanced radiotherapy modalities such as stereotactic radiosurgery (SRS) and image-guided radiotherapy may employ very small beam apertures for accurate localized high dose to target. Accurate measurement of small radiation fields is a well-known challenge for many dosimeters. The purpose of this study was to measure total scatter factors for stereotactic cones with plastic scintillation detector and its comparison against diode detector and theoretical estimates. Measurements were performed on Novalis Tx™ linear accelerator for 6MV SRS beam with stereotactic cones of diameter 6 mm, 7.5 mm, 10 mm, 12.5 mm, and 15 mm. The advantage of plastic scintillator detector is in its energy dependence. The total scatter factor was measured in water at the depth of dose maximum. Total scatter factor with plastic scintillation detector was determined by normalizing the readings to field size of 10 cm × 10 cm. To overcome energy dependence of diode detector for the determination of scatter factor with diode detector, daisy chaining method was used. The plastic scintillator detector was calibrated against the ionization chamber, and the reproducibility in the measured doses was found to be within ± 1%. Total scatter factor measured with plastic scintillation detector was 0.728 ± 0.3, 0.783 ± 0.05, 0.866 ± 0.55, 0.885 ± 0.5, and 0.910 ± 0.06 for cone sizes of 6 mm, 7.5 mm, 10 mm, 12.5 mm, and 15 mm, respectively. Total scatter factor measured with diode detector was 0.733 ± 0.03, 0.782 ± 0.02, 0.834 ± 0.07, 0.854 ± 0.02, and 0.872 ± 0.02 for cone sizes of 6 mm, 7.5 mm, 10 mm, 12.5 mm, and 15 mm, respectively. The variation in the measurement of total scatter factor with published Monte Carlo data was found to be −1.3%, 1.9%, −0.4%, and 0.4% for cone sizes of 7.5 mm, 10 mm, 12.5 mm, and 15 mm, respectively. We conclude that total scatter factor measurements for stereotactic cones can be adequately carried out with a plastic scintillation detector. Our results show a high level of consistency within our data and compared well with published data.
Collapse
Affiliation(s)
- Suresh H Chaudhari
- Department of Radiation Oncology, Apollo Hospitals, Navi Mumbai, Maharashtra, India
| | - Rishabh Dobhal
- Department of Radiation Oncology, Batra Hospital and Medical Research Centre, New Delhi, India
| | - Rajesh A Kinhikar
- Department of Medical Physics, Tata Memorial Hospital, Mumbai, Maharashtra, India
| | - Sudarshan S Kadam
- Department of Medical Physics, Tata Memorial Hospital, Mumbai, Maharashtra, India
| | - Deepak D Deshpande
- Department of Medical Physics, Tata Memorial Hospital, Mumbai, Maharashtra, India
| |
Collapse
|
10
|
Dechambre D, Baart V, Cucchiaro S, Ernst C, Jansen N, Berkovic P, Mievis C, Coucke P, Gulyban A. Commissioning Monte Carlo algorithm for robotic radiosurgery using cylindrical 3D-array with variable density inserts. Phys Med 2017; 33:152-158. [DOI: 10.1016/j.ejmp.2017.01.005] [Citation(s) in RCA: 13] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 07/06/2016] [Revised: 11/18/2016] [Accepted: 01/07/2017] [Indexed: 10/20/2022] Open
|
11
|
Jacqmin DJ, Bredfeldt JS, Frigo SP, Smilowitz JB. Implementation of the validation testing in MPPG 5.a "Commissioning and QA of treatment planning dose calculations-megavoltage photon and electron beams". J Appl Clin Med Phys 2016; 18:115-127. [PMID: 28291929 PMCID: PMC5689890 DOI: 10.1002/acm2.12015] [Citation(s) in RCA: 11] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Key Words] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/01/2016] [Accepted: 10/17/2016] [Indexed: 11/17/2022] Open
Abstract
The AAPM Medical Physics Practice Guideline (MPPG) 5.a provides concise guidance on the commissioning and QA of beam modeling and dose calculation in radiotherapy treatment planning systems. This work discusses the implementation of the validation testing recommended in MPPG 5.a at two institutions. The two institutions worked collaboratively to create a common set of treatment fields and analysis tools to deliver and analyze the validation tests. This included the development of a novel, open‐source software tool to compare scanning water tank measurements to 3D DICOM‐RT Dose distributions. Dose calculation algorithms in both Pinnacle and Eclipse were tested with MPPG 5.a to validate the modeling of Varian TrueBeam linear accelerators. The validation process resulted in more than 200 water tank scans and more than 50 point measurements per institution, each of which was compared to a dose calculation from the institution's treatment planning system (TPS). Overall, the validation testing recommended in MPPG 5.a took approximately 79 person‐hours for a machine with four photon and five electron energies for a single TPS. Of the 79 person‐hours, 26 person‐hours required time on the machine, and the remainder involved preparation and analysis. The basic photon, electron, and heterogeneity correction tests were evaluated with the tolerances in MPPG 5.a, and the tolerances were met for all tests. The MPPG 5.a evaluation criteria were used to assess the small field and IMRT/VMAT validation tests. Both institutions found the use of MPPG 5.a to be a valuable resource during the commissioning process. The validation testing in MPPG 5.a showed the strengths and limitations of the TPS models. In addition, the data collected during the validation testing is useful for routine QA of the TPS, validation of software upgrades, and commissioning of new algorithms.
Collapse
Affiliation(s)
- Dustin J Jacqmin
- Department of Radiation Oncology, Medical University of South Carolina, Charleston, SC, USA.,Department of Human Oncology, University of Wisconsin-Madison, Madison, WI, USA
| | - Jeremy S Bredfeldt
- Department of Human Oncology, University of Wisconsin-Madison, Madison, WI, USA
| | - Sean P Frigo
- Department of Human Oncology, University of Wisconsin-Madison, Madison, WI, USA
| | | |
Collapse
|
12
|
Denton TR, Shields LBE, Howe JN, Spalding AC. Quantifying isocenter measurements to establish clinically meaningful thresholds. J Appl Clin Med Phys 2015; 16:5183. [PMID: 26103187 PMCID: PMC5690087 DOI: 10.1120/jacmp.v16i2.5183] [Citation(s) in RCA: 16] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/01/2014] [Revised: 11/23/2014] [Accepted: 11/23/2014] [Indexed: 11/23/2022] Open
Abstract
A dataset range of isocenter congruency verification tests have been examined from a statistical perspective for the purpose of establishing tolerance levels that are meaningful, based on the fundamental limitation of linear accelerator isocentricity and the demands of a high-precision stereotactic radiosurgery program. Using a laser-defined isocenter, a total of 149 individual isocenter congruency tests were examined with recorded values for ideal spatial corrections to the isocenter test tool. These spatial corrections were determined from radiation exposures recorded on an electronic portal imaging device (EPID) at various gantry, collimator, and treatment couch combinations. The limitations of establishing an ideal isocenter were quantified from each variable which contributed to uncertainty in isocenter definition. Individual contributors to uncertainty, specifically, daily positioning setup errors, gantry sag, multileaf collimator (MLC) offset, and couch walkout, were isolated from isocenter congruency measurements to determine a clinically meaningful isocenter measurement. Variations in positioning of the test tool constituted, on average, 0.38 mm magnitude of correction. Gantry sag and MLC offset contributed 0.4 and 0.16 mm, respectively. Couch walkout had an average degrading effect to isocenter of 0.72 mm. Considering the magnitude of uncertainty contributed by each uncertainty variable and the nature of their combination, an appropriate schedule action and immediate action level were determined for use in analyzing daily isocenter congruency test results in a stereotactic radiosurgery (SRS) program. The recommendations of this study for this linear accelerator include a schedule action level of 1.25 mm and an immediate action level of 1.50mm, requiring prompt correction response from clinical medical physicists before SRS or stereotactic body radiosurgery (SBRT) is administered. These absolute values were derived from considering relative data from a specific linear accelerator and, therefore, represent a means by which a numerical quantity can be used as a test threshold with relative specificity to a particular linear accelerator.
Collapse
Affiliation(s)
- Travis R Denton
- The Norton Cancer Institute Radiation Center, Louisville, KY and Associates in Medical Physics, LLC, Greenbelt, MD.
| | | | | | | |
Collapse
|
13
|
Rout B, Muralidhar K, Ali M, Shekar M, Kumar A. Dosimetric study of RapidArc plans with flattened beam (FB) and flattening filter-free (FFF) beam for localized prostate cancer based on physical indices. INTERNATIONAL JOURNAL OF CANCER THERAPY AND ONCOLOGY 2014. [DOI: 10.14319/ijcto.0204.6] [Citation(s) in RCA: 9] [Impact Index Per Article: 0.9] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/05/2022] Open
|