Cooper JD, Muirhead DC, Taylor JE, Baker PR. Development of an assay for the simultaneous determination of sildenafil (Viagra) and its metabolite (UK-103,320) using automated sequential trace enrichment of dialysates and high-performance liquid chromatography.
JOURNAL OF CHROMATOGRAPHY. B, BIOMEDICAL SCIENCES AND APPLICATIONS 1997;
701:87-95. [PMID:
9389342 DOI:
10.1016/s0378-4347(97)00339-3]
[Citation(s) in RCA: 84] [Impact Index Per Article: 3.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 02/05/2023]
Abstract
The development of the ASTED (automated sequential trace enrichment of dialysates) system to prepare plasma samples prior to high-performance liquid chromatography (HPLC) of sildenafil and its demethylated metabolite (UK-103,320) is described. Investigations to elucidate potential pitfalls of the ASTED on-line sample preparation system prior to separation of the analytes by HPLC are presented. The procedure is shown to be selective for sildenafil and UK-103,320, and linear over the range 1.00-250 ng/ml. The intra-batch imprecision (C.V.) of the method at plasma analyte concentrations of 1.00, 5.00, 50.0 and 200 ng/ml was 11.2, 3.10, 1.50, and 1.30%, respectively, and the corresponding inter-batch imprecision was estimated to be 13.5, 7.09, 3.69, and 5.43%. At these plasma analyte concentrations the overall inaccuracy (% bias) of the procedure ranged from 3.6 to 8.4%. The method showed similar assay performance for the estimation of the metabolite, UK-103,320. The application of the assay to a pharmacokinetic investigation during a clinical study is presented.