Wu Z, Kumon RE, Laughner JI, Efimov IR, Deng CX. Electrophysiological changes correlated with temperature increases induced by high-intensity focused ultrasound ablation.
ULTRASOUND IN MEDICINE & BIOLOGY 2015;
41:432-448. [PMID:
25516446 PMCID:
PMC4297512 DOI:
10.1016/j.ultrasmedbio.2014.09.009]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Received: 11/25/2013] [Revised: 08/30/2014] [Accepted: 09/04/2014] [Indexed: 06/04/2023]
Abstract
To gain better understanding of the detailed mechanisms of high-intensity focused ultrasound (HIFU) ablation for cardiac arrhythmias, we investigated how the cellular electrophysiological (EP) changes were correlated with temperature increases and thermal dose (cumulative equivalent minutes [CEM43]) during HIFU application using Langendorff-perfused rabbit hearts. Employing voltage-sensitive dye di-4-ANEPPS, we measured the EP and temperature during HIFU using simultaneous optical mapping and infrared imaging. Both action potential amplitude (APA) and action potential duration at 50% repolarization (APD50) decreased with temperature increases, and APD50 was more thermally sensitive than APA. EP and tissue changes were irreversible when HIFU-induced temperature increased above 52.3 ± 1.4°C and log10(CEM43) above 2.16 ± 0.51 (n = 5), but were reversible when temperature was below 50.1 ± 0.8°C and log10(CEM43) below -0.9 ± 0.3 (n = 9). EP and temperature/thermal dose changes were spatially correlated with HIFU-induced tissue necrosis surrounded by a transition zone.
Collapse