Rapp ES, Pawar SR, Longoria RG. Hybrid Mock Circulatory Loop Simulation of Extreme Cardiac Events.
IEEE Trans Biomed Eng 2022;
69:2883-2892. [PMID:
35254970 PMCID:
PMC9466991 DOI:
10.1109/tbme.2022.3156963]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/08/2022]
Abstract
OBJECTIVE
This paper presents preliminary methods of incorporating the pathological conditions of cardiac arrhythmias and valvular stenosis in hybrid mock circulation loop (hMCL) operation for the enhanced verification and validation of mechanical circulatory support devices such as VADs.
METHODS
The MGH/MF Waveform datasets from PhysioNet database (including both nominal and clinically diagnosed arrhythmic ECG measurements) as well as cardiovascular system model updates are used to recreate arrhythmic events and valvular stenosis in vitro.
RESULTS
Preliminary results show the hMCL can recreate each tested cardiac event within 2% and 4% mean error for reference pressure tracking in the aortic and left ventricular pressure chambers, respectively. Further, frequency spectrum analysis comparisons using the magnitude-squared coherence analysis shows close alignment between measured arrhythmic and hMCL realized pressure frequency content.
CONCLUSION
The generation of cardiac arrhythmias and valvular stenosis around a VAD via both model and acute measurement based methods was achieved.
SIGNIFICANCE
Pathological conditions such as cardiac arrhythmias and valvular stenosis are limited in documentation despite the large percentage of patients who experience these events. This paper provides a means to begin incorporating these events into hardware-in-the-loop mock circulatory systems for next generation VAD validation and verification.
Collapse