Vittoria G, Pantaleo G, Blasi A, Spagnuolo G, Iandolo A, Amato M. Thermafil: A New Clinical Approach Due to New Dimensional Evaluations.
Open Dent J 2018. [PMID:
29541263 PMCID:
PMC5838644 DOI:
10.2174/1874210601812010173]
[Citation(s) in RCA: 1] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/22/2022] Open
Abstract
Background
There are a lot of techniques to obturate the root canals, but lateral condensation of gutta-percha is the most used one. An important aspect of thermafil is the error margin tolerated by the manufacturer in the production of plastic carriers. In literature, there is no evidence about discrepancy percentage between different carriers. It is demonstrated that the error margin of gutta-percha is 0.5% and is 0.2% for metal files (ISO standards).
Objective
The aim of this study was to evaluate the real dimensions of thermafil plastic carriers observed by the stereo microscope measuring the dimensional discrepancy between them.
Methods
For this study, 80 new thermafil (Dentsply Maillefer) have been selected. 40 thermafil 0.25 and 40 thermafil 0.30. Through 60X stereo microscope, the dimensions of the plastic carrier tips have been measured. The dimensions of the plastic carrier were also measured after a heating cycle. ZL GAL 11TUSM (Zetaline stereo evolution) microscope was used to observe the samples. Measurements were made through a dedicated software (Image Focus). All samples were analysed at 60X.
Results
A non-parametric paired test (Wilcoxon test) was used to compare baseline and after heating values; p-values ≤ 0.05 were assumed to be statistically significant.
Conclusion
The samples we measured showed a mean value of the diameters in Thermafil 25 that was 0.27 mm, for Thermafil 30 the mean value was 0.33 mm.We have measured a dimensional variable of 8% in the 25 group while in group 30 the maximum possible variation found was 4%, that's why we propose a new protocol of obturation with thermafil. We can also conclude that a single heating process does not affect clinically the plastic carrier dimensions.
Collapse